Optimization of the imaging protocol of an X-ray CT scanner for evaluation of normoxic polymer gel dosimeters

被引:9
|
作者
Subramanian, Brindha [1 ]
Ravindran, Paul B. [1 ]
Baldock, Clive [2 ]
机构
[1] Christian Med Coll & Hosp, Dept Radiotherapy, Vellore 632004, Tamil Nadu, India
[2] Univ Sydney, Inst Med Phys, Sch Phys, Sydney, NSW, Australia
关键词
Normoxic polymer gels; X-ray CT; imaging protocol;
D O I
10.4103/0971-6203.26692
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
X-ray computer tomography (CT) has previously been reported as an evaluation tool for polymer gel (PAG) dosimeters. In this study, the imaging protocol of a Siemens Emotion X-ray CT scanner was optimized to evaluate PAGAT normoxic gel dosimeters. The scan parameters were optimized as 130 kV and 150 mA with a slice thickness of 3 mm for smaller fields and 5 mm for larger fields of irradiation. The number of images to be averaged to reduce noise to an acceptable level was concluded to be 25. It was also concluded that the total monomer concentration required is 7% with 10 mM THP to obtain a maximum change in CT number at dose levels up to 15 Gy for evaluation with X-ray CT. Optimal scan parameters may vary with X-ray CT scanner. Hence the imaging protocol of each scanner to be used for evaluating polymer gels requires individual optimization for the purposes of gel dosimetry evaluation.
引用
收藏
页码:72 / 77
页数:6
相关论文
共 50 条
  • [1] The effect of x-ray CT dose on normoxic polymer gel dosimeters
    Baxter, P.
    Jirasek, A.
    Hilts, M.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S56 - S56
  • [2] The dose response of normoxic polymer gel dosimeters measured using X-ray CT
    Hill, B
    Venning, A
    Baldock, C
    [J]. BRITISH JOURNAL OF RADIOLOGY, 2005, 78 (931): : 623 - 630
  • [3] A preliminary study of the novel application of normoxic polymer gel dosimeters for the measurement of CTDI on diagnostic x-ray CT scanners
    Hill, B
    Venning, AJ
    Baldock, C
    [J]. MEDICAL PHYSICS, 2005, 32 (06) : 1589 - 1597
  • [4] Polymer gel dosimeters with enhanced sensitivity for use in x-ray CT polymer gel dosimetry
    Jirasek, A.
    Hilts, M.
    McAuley, K. B.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (18): : 5269 - 5281
  • [5] Improved image quality for x-ray CT imaging of gel dosimeters
    Kakakhel, M. B.
    Kairn, T.
    Kenny, J.
    Trapp, J. V.
    [J]. MEDICAL PHYSICS, 2011, 38 (09) : 5130 - 5135
  • [6] Commissioning a Multislice CT Scanner for X-Ray CT Polymer Gel Dosimetry
    Johnston, H.
    Hilts, M.
    Jirasek, A.
    [J]. MEDICAL PHYSICS, 2014, 41 (06) : 238 - 238
  • [7] Characterization of the essential dosimetric properties of cosolvent-free polymer gel dosimeters: Recent progress in x-ray CT based normoxic polymer gel dosimetry
    Johnston, H.
    Hilts, M.
    Carrick, J.
    Jirasek, A.
    [J]. 7TH INTERNATIONAL CONFERENCE ON 3D RADIATION DOSIMETRY (IC3DDOSE), 2013, 444
  • [8] Characterization of a commercial optical CT scanner for imaging BANG® polymer gel dosimeters
    Islam, K
    Dempsey, J
    Maryanski, M
    Ranade, M
    Sohn, J
    Low, D
    [J]. MEDICAL PHYSICS, 2002, 29 (06) : 1371 - 1371
  • [9] X-ray CT dose in normoxic polyacrylamide gel dosimetry
    Baxter, P.
    Jirasek, A.
    Hilts, M.
    [J]. MEDICAL PHYSICS, 2007, 34 (06) : 1934 - 1943
  • [10] Dose rate properties of NIPAM-based x-ray CT polymer gel dosimeters
    Jirasek, A.
    Johnston, H.
    Hilts, M.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (11): : 4399 - 4411