Optimization of the imaging protocol of an X-ray CT scanner for evaluation of normoxic polymer gel dosimeters

被引:9
|
作者
Subramanian, Brindha [1 ]
Ravindran, Paul B. [1 ]
Baldock, Clive [2 ]
机构
[1] Christian Med Coll & Hosp, Dept Radiotherapy, Vellore 632004, Tamil Nadu, India
[2] Univ Sydney, Inst Med Phys, Sch Phys, Sydney, NSW, Australia
关键词
Normoxic polymer gels; X-ray CT; imaging protocol;
D O I
10.4103/0971-6203.26692
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
X-ray computer tomography (CT) has previously been reported as an evaluation tool for polymer gel (PAG) dosimeters. In this study, the imaging protocol of a Siemens Emotion X-ray CT scanner was optimized to evaluate PAGAT normoxic gel dosimeters. The scan parameters were optimized as 130 kV and 150 mA with a slice thickness of 3 mm for smaller fields and 5 mm for larger fields of irradiation. The number of images to be averaged to reduce noise to an acceptable level was concluded to be 25. It was also concluded that the total monomer concentration required is 7% with 10 mM THP to obtain a maximum change in CT number at dose levels up to 15 Gy for evaluation with X-ray CT. Optimal scan parameters may vary with X-ray CT scanner. Hence the imaging protocol of each scanner to be used for evaluating polymer gels requires individual optimization for the purposes of gel dosimetry evaluation.
引用
收藏
页码:72 / 77
页数:6
相关论文
共 50 条
  • [41] Evaluation of epitaxial silicon diodes as dosimeters in X-ray mammography
    Goncalves, J. A. C.
    Pereira, L. N.
    Potiens, M. P. A.
    Vivolo, V.
    Bueno, C. C.
    [J]. RADIATION MEASUREMENTS, 2014, 71 : 384 - 388
  • [42] Evaluation of radiation dosimeters for kilovoltage x-ray beam dosimetry
    Hill, R.
    Me, Z.
    Haque, M.
    Baldock, C.
    [J]. MEDICAL PHYSICS, 2007, 34 (06) : 2336 - 2337
  • [43] Optimization of a secondary VOI protocol for lung imaging in a clinical CT scanner
    Larsen, Thomas C.
    Gopalakrishnan, Vissagan
    Yao, Jianhua
    Nguyen, Catherine P.
    Chen, Marcus Y.
    Moss, Joel
    Wen, Han
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2018, 19 (04): : 271 - 280
  • [44] An x-ray CT polymer gel dosimetry prototype: II. Gel characterization and clinical application
    Johnston, H.
    Hilts, M.
    Carrick, J.
    Jirasek, A.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (10): : 3155 - 3175
  • [45] X-RAY CT SCANNER FOR HEAVY AND VALUABLE COMPONENTS.
    Tanimoto, Yoshinori
    Matsuura, Shigeo
    Fujii, Masashi
    Maniwa, Takao
    [J]. 1600,
  • [46] Characterization of failure in soils using X-ray CT scanner
    Otani, J
    Obara, Y
    Mukunoki, T
    [J]. DEFORMATION AND PROGRESSIVE FAILURE IN GEOMECHANICS - IS-NAGOYA'97, 1997, : 311 - 314
  • [47] Improvement of sensitivity of X-ray CT reading method for polymer gel in radiation therapy
    Javaheri, Negin
    Yarahmadi, Mehran
    Refaei, Abdollah
    Aghamohammadi, Ali
    [J]. REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2020, 25 (01) : 100 - 103
  • [48] Analytical and rheological studies of modified gel dosimeters exposed to X-ray beams
    Romero, M.
    Mattea, F.
    Vedelago, J.
    Chacon, D.
    Valente, M.
    Alvarez Igarzabal, C.
    Strumia, M.
    [J]. MICROCHEMICAL JOURNAL, 2016, 127 : 231 - 236
  • [49] Evaluation of three-dimensional polymer gel dosimetry using X-ray CT and R2 MRI
    Kawamura, H.
    Sakae, T.
    Terunuma, T.
    Ishida, M.
    Shibata, Y.
    Matsumura, A.
    [J]. APPLIED RADIATION AND ISOTOPES, 2013, 77 : 94 - 102
  • [50] Evaluation of x-ray scatter properties in a dedicated cone-beam breast CT scanner
    Kwan, ALC
    Boone, JM
    Shah, N
    [J]. MEDICAL PHYSICS, 2005, 32 (09) : 2967 - 2975