The dose response of normoxic polymer gel dosimeters measured using X-ray CT

被引:55
|
作者
Hill, B
Venning, A
Baldock, C
机构
[1] Canberra Hosp, Biomed Engn Serv, Med Phys Sect, Woden, ACT 2606, Australia
[2] Queensland Univ Technol, Sch Phys & Chem Sci, Brisbane, Qld, Australia
[3] Univ Sydney, Sch Phys, Inst Med Phys, Sydney, NSW 2006, Australia
来源
BRITISH JOURNAL OF RADIOLOGY | 2005年 / 78卷 / 931期
关键词
D O I
10.1259/bjr/46029447
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
X-ray CT was used to determine the dose response of normoxic polymer gel dosimeters. Normoxic polymer gel dosimeters were manufactured and irradiated up to 150 Gy. Up to 50 CT images were acquired oil a Toshiba Aquilion Multislice CT scanner using protocols for 80 kV and 135 kV to determine dose response. HU-dose sensitivity, the linear regression of data for the HU versus dose for the linear part of the curve LIP to 60 Gy was 0.38 +/- 0.07 HU Gy(-1) for 135 kV and 0.37 +/- 0.01 H U Gy(-1) for 80 kV. Dose resolution was found to be < 1.3 Gy for all absorbed dose range up to 70 Gy for 135 kV, similar to that measured previously for polyacrylamide gel (PAG). Although the HU-dose sensitivity was lower than that previously measured for PAG gel dosimeters it had a greater range of absorbed dose indicating that normoxic polymer gel dosimeters have potential in CT gel dosimetry.
引用
收藏
页码:623 / 630
页数:8
相关论文
共 50 条
  • [21] Evaluation of an x-ray CT polymer gel dosimetry system in the measurement of deformed dose
    Maynard, E.
    Heath, E.
    Hilts, M.
    Jirasek, A.
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2020, 6 (03):
  • [22] Considerations for x-ray CT polymer gel dosimetry
    Jirasek, A.
    7TH INTERNATIONAL CONFERENCE ON 3D RADIATION DOSIMETRY (IC3DDOSE), 2013, 444
  • [23] Depth dose profiles measured in the MAGAS normoxic polymer gel dosimeter
    Venning, A
    Nitschke, K
    Healy, B
    Baldock, C
    MEDICAL PHYSICS, 2004, 31 (06) : 1877 - 1877
  • [24] DYNAMIC MODELLING OF THE DOSE RESPONSE OF POLYMER GEL DOSIMETERS
    Ceberg, S.
    Lepage, M.
    Back, S.
    Gustafsson, H.
    Ceberg, C.
    RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S408 - S409
  • [25] X-ray energy dependence of the dose response of SIRAD radiation dosimeters
    Cheung, Tsang
    Butson, Martin J.
    Yu, Peter K. N.
    APPLIED RADIATION AND ISOTOPES, 2007, 65 (07) : 814 - 817
  • [26] Normoxic polyacrylamide gel doped with iodine: Response versus X-ray energy
    Gastaldo, Jerome
    Boudou, Caroline
    Lamalle, Laurent
    Tropres, Irene
    Corde, Stephanie
    Sollier, Alberic
    Rucka, Guenther
    Elleaume, Helene
    EUROPEAN JOURNAL OF RADIOLOGY, 2008, 68 (03) : S118 - S120
  • [27] A feasibility study of multislice X-ray CT imaging of gel dosimeters using the "zero scan" method
    Kakakhel, Muhammad B.
    Kairn, Tanya
    Charles, Paul
    Trapp, Jamie V.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2014, 15 (04): : 367 - 371
  • [28] First characterization of X-ray dose in murine animals using the Inveon CT platform and nanoDot dosimeters
    Osborne, Dustin
    Stuckey, Alan
    Richey, Tina
    Kennel, Steve
    Pryer, Lindy
    Wall, Jonathan
    JOURNAL OF NUCLEAR MEDICINE, 2012, 53
  • [29] Severe dose inaccuracies caused by an oxygen-antioxidant imbalance in normoxic polymer gel dosimeters
    Sedaghat, Mahbod
    Bujold, Rachel
    Lepage, Martin
    PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (03): : 601 - 625
  • [30] The influence of cooling rate on the accuracy of normoxic polymer gel dosimeters
    De Deene, Y.
    Pittomvils, G.
    Visalatchi, S.
    PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (10): : 2719 - 2728