Variation in tube voltage for pediatric neck 64VCT: Effect on radiation dose and image quality

被引:1
|
作者
Chen, Li-Guo [1 ]
Wu, Ping-An [1 ]
Tu, Hsing-Yang [1 ]
Sheu, Ming-Huei [1 ]
Huang, Li-Chuan [1 ,2 ]
机构
[1] Buddhist Tzu Chi Med Fdn, Hualien Tzu Chi Hosp, Dept Med Imaging, Hualien, Taiwan
[2] Tzu Chi Univ Sci & Technol, Dept Med Imaging & Radiol Sci, Hualien, Taiwan
来源
PLOS ONE | 2021年 / 16卷 / 11期
关键词
CT; PHANTOM; RECONSTRUCTION; REDUCTION; EXPOSURE;
D O I
10.1371/journal.pone.0259772
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Exposure to ionizing radiation can cause cancer, especially in children. In computed tomography (CT), a trade-off exists between the radiation dose and image quality. Few studies have investigated the effect of dose reduction on image quality in pediatric neck CT. We aimed to assess the effect of peak kilovoltage on the radiation dose and image quality in pediatric neck multidetector-row CT. Measurements were made using three phantoms representative of children aged 1, 5, and 10 years, with tube voltages of 80, 100, and 120 kilovoltage peak (kVp); tube current of 10, 40, 80, 120, 150, 200, and 250 mA; and exposure time = 0.5 s (pitch, 0.984:1). Radiation dose estimates were derived from the dose-length product with a 64-multidetector-row CT scanner. Images obtained from the control protocol (120 kVp) were compared with the 80- and 100-kVp protocols. The effective dose (ED) was determined for each protocol and compared with the 120-kVp protocol. Quantitative analysis entailed noise measurements by recording the standard deviation of attenuation for a circular 1-cm(2) region of interest placed on homogeneous soft tissue structures in the phantom. The mean noise of the various kVp protocols was compared using the unpaired Student t-test. Reduction of ED was 37.58% and 68.58% for neck CT with 100 kVp and 80 kVp, respectively. The image noise level increased with the decrease in peak kilovoltage. Noise values were higher at 80 kVp at all neck levels, but did not increase at 100 kVp, compared to 120 kVp in the three phantoms. The measured noise difference was the greatest at 80 kVp (absolute increases<2.5 HU). The subjective image quality did not differ among the protocols. Thus, reducing voltage from 120 to 80 kVp for neck CT may achieve ED reduction of 68.58%, without compromising image quality.
引用
收藏
页数:12
相关论文
共 50 条
  • [1] Variation in Tube Voltage for Adult Neck MDCT: Effect on Radiation Dose and Image Quality
    Hoang, Jenny K.
    Yoshizumi, Terry T.
    Giao Nguyen
    Toncheva, Greta
    Choudhury, Kingshuk Roy
    Gafton, Andreia R.
    Eastwood, James D.
    Lowry, Carolyn
    Hurwitz, Lynne M.
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2012, 198 (03) : 621 - 627
  • [2] Evaluation of the effect of low tube voltage on radiation dose and image quality
    Din, Nik Norhasrina Nik
    Zainon, Rafidah
    Rahman, A. T. Abdul
    [J]. 10TH INTERNATIONAL SEMINAR ON MEDICAL PHYSICS & 1ST AMDI INTERNATIONAL ONCOLOGY SYMPOSIUM, 2017, 851
  • [3] Radiation Dose and Image Quality in Pediatric Neck CT
    Tipnis, S. V.
    Rieter, W. J.
    Patel, D.
    Stalcup, S. T.
    Matheus, M. G.
    Spampinato, M. V.
    [J]. AMERICAN JOURNAL OF NEURORADIOLOGY, 2019, 40 (06) : 1067 - 1073
  • [4] Effect of Tube Voltage and Radiation Dose on Image Quality in Pediatric Abdominal CT Using Deep Learning Reconstruction: A Phantom Study
    Kim, Daehong
    Jeon, Pil-Hyun
    Lee, Chang-Lae
    Chung, Myung-Ae
    [J]. SYMMETRY-BASEL, 2023, 15 (02):
  • [5] Automatic Tube Current Modulation and Tube Voltage Selection in Pediatric Computed Tomography A Phantom Study on Radiation Dose and Image Quality
    Papadalds, Antonios E.
    Damilakis, John
    [J]. INVESTIGATIVE RADIOLOGY, 2019, 54 (05) : 265 - 272
  • [6] Impact of Tube Voltage on Radiation Dose (CTDI) and Image Quality at Chest CT Examination
    El Mansouri, M.
    Choukri, A.
    Talbi, M.
    Hakam, O. K.
    [J]. ATOM INDONESIA, 2021, 47 (02) : 105 - +
  • [7] Automated tube voltage adaptation in head and neck computed tomography between 120 and 100 kV: effects on image quality and radiation dose
    Matthias S. May
    Manuel R. Kramer
    Achim Eller
    Wolfgang Wuest
    Michael Scharf
    Michael Brand
    Marc Saake
    Bernhard Schmidt
    Michael Uder
    Michael M. Lell
    [J]. Neuroradiology, 2014, 56 : 797 - 803
  • [8] Automated tube voltage adaptation in head and neck computed tomography between 120 and 100 kV: effects on image quality and radiation dose
    May, Matthias S.
    Kramer, Manuel R.
    Eller, Achim
    Wuest, Wolfgang
    Scharf, Michael
    Brand, Michael
    Saake, Marc
    Schmidt, Bernhard
    Uder, Michael
    Lell, Michael M.
    [J]. NEURORADIOLOGY, 2014, 56 (09) : 797 - 803
  • [9] Evaluation of image quality and radiation dose in low tube voltage coronary computed tomography angiography
    Mousavi-Gazafroudi, Seyedeh Shokoofeh
    Sajjadieh-Khajouei, Amirreza
    Moradi, Maryam
    Mousavi-Gazafroudi, Seyedeh Shabnam
    Yadegarfar, Ghasem
    Tavakoli, Mohammad Bagher
    [J]. ARYA ATHEROSCLEROSIS, 2019, 15 (05) : 205 - 210
  • [10] Evaluation of image quality and radiation dose in computed tomography urography following tube voltage optimisation
    Saidu, A. M.
    Garba, I.
    Abba, M.
    Yahuza, M. A.
    Yusuf, L.
    Tahir, N. M.
    Garko, S. S.
    [J]. RADIOGRAPHY, 2024, 30 (01) : 301 - 307