Effect of Tube Voltage and Radiation Dose on Image Quality in Pediatric Abdominal CT Using Deep Learning Reconstruction: A Phantom Study

被引:1
|
作者
Kim, Daehong [1 ]
Jeon, Pil-Hyun [2 ]
Lee, Chang-Lae [3 ]
Chung, Myung-Ae [4 ]
机构
[1] Eulji Univ, Dept Radiol Sci, Seongnam 13135, South Korea
[2] Wonju Severance Christian Hosp, Dept Radiol, Wonju 26426, South Korea
[3] Samsung Elect, Hlth & Med Equipment Business Unit, Suwon 16677, South Korea
[4] Eulji Univ, Dept Bigdata Med Convergence, Seongnam 13135, South Korea
来源
SYMMETRY-BASEL | 2023年 / 15卷 / 02期
关键词
tube voltage; radiation dose; image quality; pediatric computed tomography (CT); deep learning reconstruction (DLR); ALGORITHM; REDUCTION;
D O I
10.3390/sym15020501
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Background: Children have a potential risk from radiation exposure because they are more sensitive to radiation than adults. Objective: The purpose of this work is to estimate image quality according to tube voltage (kV) and radiation dose in pediatric computed tomography (CT) using deep learning reconstruction (DLR). Methods: Phantom images of children and adults were obtained for kV, radiation dose, and image reconstruction methods. The CT emits a fan beam to the opposite detector, and the geometry of the detector was symmetrical. Phantom images of children and adults were acquired at a volume CT dose index (CTDIvol) from 0.5 to 10.0 mGy for tube voltages at 80, 100, and 120 kV. A DLR was used to reconstruct the phantom image, and filtered back projection (FBP) and iterative reconstruction (IR) were also performed for comparison with the DLR. Image quality was evaluated by measuring the contrast-to-noise ratio (CNR) and noise. Results: Under the same imaging conditions, the DLR images of pediatric and adult phantoms generally provided improved CNR and noise compared with the FBP and IR images. At a similar CNR and noise, the FBP, IR, and DLR of the pediatric images showed a dose reduction compared with the FBP, IR, and DLR of the adult images, respectively. In terms of the effect of tube voltage, the CNR of the 100 kV DLR images was higher than that of the 120 kV DLR images. Conclusion: According to the results, since pediatric CT images maintain the same image quality at lower doses compared with adult CT images, DLR can improve image quality while reducing the radiation dose in children's abdominal CT scans.
引用
收藏
页数:11
相关论文
共 50 条
  • [1] Deep learning image reconstruction algorithm for abdominal multidetector CT at different tube voltages: assessment of image quality and radiation dose in a phantom study
    Hye Joo Park
    Seo-Youn Choi
    Ji Eun Lee
    Sanghyeok Lim
    Min Hee Lee
    Boem Ha Yi
    Jang Gyu Cha
    Ji Hye Min
    Bora Lee
    Yunsub Jung
    [J]. European Radiology, 2022, 32 : 3974 - 3984
  • [2] Deep learning image reconstruction algorithm for abdominal multidetector CT at different tube voltages: assessment of image quality and radiation dose in a phantom study
    Park, Hye Joo
    Choi, Seo-Youn
    Lee, Ji Eun
    Lim, Sanghyeok
    Lee, Min Hee
    Yi, Boem Ha
    Cha, Jang Gyu
    Min, Ji Hye
    Lee, Bora
    Jung, Yunsub
    [J]. EUROPEAN RADIOLOGY, 2022, 32 (06) : 3974 - 3984
  • [3] Estimates of the image quality in accordance with radiation dose for pediatric imaging using deep learning CT: A phantom study
    Jeon, Pil-Hyun
    Kim, Daehong
    Chung, Myung-Ae
    [J]. 2022 IEEE INTERNATIONAL CONFERENCE ON BIG DATA AND SMART COMPUTING (IEEE BIGCOMP 2022), 2022, : 352 - 356
  • [4] Improving Image Quality and Reducing Radiation Dose for Pediatric CT by Using Deep Learning Reconstruction
    Brady, Samuel L.
    Trout, Andrew T.
    Somasundaram, Elanchezhian
    Anton, Christopher G.
    Li, Yinan
    Dillman, Jonathan R.
    [J]. RADIOLOGY, 2021, 298 (01) : 180 - 188
  • [5] Effect of Low Tube Voltage on Image Quality, Radiation Dose, and Low-Contrast Detectability at Abdominal Multidetector CT: Phantom Study
    Tang, Kun
    Wang, Ling
    Li, Rui
    Lin, Jie
    Zheng, Xiangwu
    Cao, Guoquan
    [J]. JOURNAL OF BIOMEDICINE AND BIOTECHNOLOGY, 2012,
  • [6] Effect of automatic tube voltage selection on image quality and radiation dose in abdominal CT angiography of various body sizes: A phantom study
    Schindera, S. T.
    Winklehner, A.
    Alkadhi, H.
    Goetti, R.
    Fischer, M.
    Gnannt, R.
    Szucs-Farkas, Z.
    [J]. CLINICAL RADIOLOGY, 2013, 68 (02) : E79 - E86
  • [7] Improved image quality and dose reduction in abdominal CT with deep-learning reconstruction algorithm: a phantom study
    Joël Greffier
    Quentin Durand
    Julien Frandon
    Salim Si-Mohamed
    Maeliss Loisy
    Fabien de Oliveira
    Jean-Paul Beregi
    Djamel Dabli
    [J]. European Radiology, 2023, 33 : 699 - 710
  • [8] Improved image quality and dose reduction in abdominal CT with deep-learning reconstruction algorithm: a phantom study
    Greffier, Joel
    Durand, Quentin
    Frandon, Julien
    Si-Mohamed, Salim
    Loisy, Maeliss
    de Oliveira, Fabien
    Beregi, Jean-Paul
    Dabli, Djamel
    [J]. EUROPEAN RADIOLOGY, 2023, 33 (01) : 699 - 710
  • [9] Image quality and dose reduction opportunity of deep learning image reconstruction algorithm for CT: a phantom study
    Greffier, Joel
    Hamard, Aymeric
    Pereira, Fabricio
    Barrau, Corinne
    Pasquier, Hugo
    Beregi, Jean Paul
    Frandon, Julien
    [J]. EUROPEAN RADIOLOGY, 2020, 30 (07) : 3951 - 3959
  • [10] Image quality and dose reduction opportunity of deep learning image reconstruction algorithm for CT: a phantom study
    Joël Greffier
    Aymeric Hamard
    Fabricio Pereira
    Corinne Barrau
    Hugo Pasquier
    Jean Paul Beregi
    Julien Frandon
    [J]. European Radiology, 2020, 30 : 3951 - 3959