Effect of automatic tube voltage selection on image quality and radiation dose in abdominal CT angiography of various body sizes: A phantom study

被引:37
|
作者
Schindera, S. T. [1 ,2 ]
Winklehner, A. [3 ]
Alkadhi, H. [3 ]
Goetti, R. [3 ]
Fischer, M. [3 ]
Gnannt, R. [3 ]
Szucs-Farkas, Z. [1 ]
机构
[1] Univ Bern, Univ Hosp Berne, Inst Diagnost Intervent & Pediat Radiol, Bern, Switzerland
[2] Univ Basel Hosp, Dept Radiol, CH-4031 Basel, Switzerland
[3] Univ Zurich Hosp, Dept Diagnost & Intervent Radiol, CH-8091 Zurich, Switzerland
关键词
PATIENT SIZE; NOISE; ENHANCEMENT; ATTENUATION; REDUCTION; SETTINGS; 16-MDCT; ENERGY; TUMORS; PHASE;
D O I
10.1016/j.crad.2012.10.007
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
AIM: To assess the effect of an automatic tube voltage selection technique on image quality and radiation dose in abdominal computed tomography (CT) angiography of various body sizes. MATERIALS AND METHODS: An abdominal aortic phantom was filled with iodinated contrast medium and placed into three different cylindrical water containers, which simulated a small, intermediate-sized, and large patient. The phantom was scanned with a standard 120 kVp abdominal CT angiography protocol and with an optimized tube voltage protocol that was modulated by an automatic tube voltage technique. The attenuation of the aorta, background, and image noise was measured, and the contrast-to-noise ratio (CNR) was calculated. Three independent readers assessed the overall image quality. RESULTS: The automatic tube voltage technique selected 70 kVp as the optimal tube voltage for the small phantom, 80 kVp for the intermediate phantom, and 100 kVp for the large phantom. Compared to the standard 120 kVp protocol, the automatic tube voltage selection yielded significantly increased CNR values in the small phantom (15.8 versus 19.4, p < 0.001), intermediate phantom (8.4 versus 8.7, p < 0.05), and large phantom (4.3 versus 4.6, p < 0.01). The automatic tube voltage selection resulted in a 55%, 49%, and 39% reduction in the volume CT dose index (CTDIvol) in the small, intermediate, and large phantoms, respectively. The subjective overall image quality of the three phantom sizes at different tube voltages ranged between poor and good. CONCLUSION: Compared to a standard 120 kVp abdominal CT angiography protocol, the automatic tube voltage selection substantially reduced the radiation dose without compromising image quality in various simulated patient sizes. (C) 2012 The Royal College of Radiologists. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:E79 / E86
页数:8
相关论文
共 50 条
  • [1] Effect of Automatic Tube Current Modulation on Radiation Dose and Image Quality for Low Tube Voltage Multidetector Row CT Angiography: Phantom Study
    Schindera, Sebastian T.
    Nelson, Rendon C.
    Yoshizumi, Terry
    Toncheva, Greta
    Nguyen, Giao
    DeLong, David M.
    Szucs-Farkas, Zsolt
    [J]. ACADEMIC RADIOLOGY, 2009, 16 (08) : 997 - 1002
  • [2] Effect of Tube Voltage and Radiation Dose on Image Quality in Pediatric Abdominal CT Using Deep Learning Reconstruction: A Phantom Study
    Kim, Daehong
    Jeon, Pil-Hyun
    Lee, Chang-Lae
    Chung, Myung-Ae
    [J]. SYMMETRY-BASEL, 2023, 15 (02):
  • [3] Automatic Tube Current Modulation and Tube Voltage Selection in Pediatric Computed Tomography A Phantom Study on Radiation Dose and Image Quality
    Papadalds, Antonios E.
    Damilakis, John
    [J]. INVESTIGATIVE RADIOLOGY, 2019, 54 (05) : 265 - 272
  • [4] Effect of Low Tube Voltage on Image Quality, Radiation Dose, and Low-Contrast Detectability at Abdominal Multidetector CT: Phantom Study
    Tang, Kun
    Wang, Ling
    Li, Rui
    Lin, Jie
    Zheng, Xiangwu
    Cao, Guoquan
    [J]. JOURNAL OF BIOMEDICINE AND BIOTECHNOLOGY, 2012,
  • [5] The application of automatic tube current modulation (ATCM) on image quality and radiation dose at abdominal computed tomography (CT): A phantom study
    Wang, Qian
    Zhao, Xinming
    Song, Junfeng
    Guo, Ning
    Zhu, Ying
    Liu, Jianxin
    Qi, Weiwei
    Wu, Jing
    Liang, Yuan
    Feng, Shichao
    Hu, Mancang
    Zhou, Chunwu
    Wang, Xiaoying
    Hong, Nan
    [J]. JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2013, 21 (04) : 453 - 464
  • [6] Abdominal multislice CT for obese patients: Effect on image quality and radiation dose in a phantom study
    Schindera, Sebastian T.
    Nelson, Rendon C.
    Lee, Ellie R.
    DeLong, David M.
    Ngyen, Giao
    Toncheva, Greta
    Yoshizumi, Terry T.
    [J]. ACADEMIC RADIOLOGY, 2007, 14 (04) : 486 - 494
  • [7] Performance of Iterative Reconstruction and Automated Tube Voltage Selection on the Image Quality and Radiation Dose in Abdominal CT Scans
    Desai, Gaurav S.
    Orrego, Jorge M. Fuentes
    Kambadakone, Avinash R.
    Sahani, Dushyant V.
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2013, 37 (06) : 897 - 903
  • [8] Effects of automatic tube potential selection on radiation dose index, image quality, and lesion detectability in pediatric abdominopelvic CT and CTA: a phantom study
    Brinkley, Michael F.
    Ramirez-Giraldo, Juan C.
    Samei, Ehsan
    Frush, Daniel J.
    Choudhury, Kingshuk Roy
    Wilson, Joshua M.
    Christianson, Olav I.
    Frush, Donald P.
    [J]. EUROPEAN RADIOLOGY, 2016, 26 (01) : 157 - 166
  • [9] Effects of automatic tube potential selection on radiation dose index, image quality, and lesion detectability in pediatric abdominopelvic CT and CTA: a phantom study
    Michael F. Brinkley
    Juan C. Ramirez-Giraldo
    Ehsan Samei
    Daniel J. Frush
    Kingshuk Roy Choudhury
    Joshua M. Wilson
    Olav I. Christianson
    Donald P. Frush
    [J]. European Radiology, 2016, 26 : 157 - 166
  • [10] Effect of low tube voltage and low iodine concentration abdominal CT on image quality and radiation dose in children: preliminary study
    You, Sun Kyoung
    Choi, Young Hun
    Cheon, Jung-Eun
    Kim, Woo Sun
    Kim, In-One
    Lee, So Mi
    Cho, Hyun-Hae
    [J]. ABDOMINAL RADIOLOGY, 2019, 44 (05) : 1928 - 1935