Effect of Tube Voltage and Radiation Dose on Image Quality in Pediatric Abdominal CT Using Deep Learning Reconstruction: A Phantom Study

被引:1
|
作者
Kim, Daehong [1 ]
Jeon, Pil-Hyun [2 ]
Lee, Chang-Lae [3 ]
Chung, Myung-Ae [4 ]
机构
[1] Eulji Univ, Dept Radiol Sci, Seongnam 13135, South Korea
[2] Wonju Severance Christian Hosp, Dept Radiol, Wonju 26426, South Korea
[3] Samsung Elect, Hlth & Med Equipment Business Unit, Suwon 16677, South Korea
[4] Eulji Univ, Dept Bigdata Med Convergence, Seongnam 13135, South Korea
来源
SYMMETRY-BASEL | 2023年 / 15卷 / 02期
关键词
tube voltage; radiation dose; image quality; pediatric computed tomography (CT); deep learning reconstruction (DLR); ALGORITHM; REDUCTION;
D O I
10.3390/sym15020501
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Background: Children have a potential risk from radiation exposure because they are more sensitive to radiation than adults. Objective: The purpose of this work is to estimate image quality according to tube voltage (kV) and radiation dose in pediatric computed tomography (CT) using deep learning reconstruction (DLR). Methods: Phantom images of children and adults were obtained for kV, radiation dose, and image reconstruction methods. The CT emits a fan beam to the opposite detector, and the geometry of the detector was symmetrical. Phantom images of children and adults were acquired at a volume CT dose index (CTDIvol) from 0.5 to 10.0 mGy for tube voltages at 80, 100, and 120 kV. A DLR was used to reconstruct the phantom image, and filtered back projection (FBP) and iterative reconstruction (IR) were also performed for comparison with the DLR. Image quality was evaluated by measuring the contrast-to-noise ratio (CNR) and noise. Results: Under the same imaging conditions, the DLR images of pediatric and adult phantoms generally provided improved CNR and noise compared with the FBP and IR images. At a similar CNR and noise, the FBP, IR, and DLR of the pediatric images showed a dose reduction compared with the FBP, IR, and DLR of the adult images, respectively. In terms of the effect of tube voltage, the CNR of the 100 kV DLR images was higher than that of the 120 kV DLR images. Conclusion: According to the results, since pediatric CT images maintain the same image quality at lower doses compared with adult CT images, DLR can improve image quality while reducing the radiation dose in children's abdominal CT scans.
引用
收藏
页数:11
相关论文
共 50 条
  • [41] Contrast agent and radiation dose reduction in abdominal CT by a combination of low tube voltage and advanced image reconstruction algorithms
    Buls, Nico
    Van Gompel, Gert
    Van Cauteren, Toon
    Nieboer, Koenraad
    Willekens, Inneke
    Verfaillie, Guy
    Evans, Paul
    Macholl, Sven
    Newton, Ben
    de Mey, Johan
    [J]. EUROPEAN RADIOLOGY, 2015, 25 (04) : 1023 - 1031
  • [42] Contrast agent and radiation dose reduction in abdominal CT by a combination of low tube voltage and advanced image reconstruction algorithms
    Nico Buls
    Gert Van Gompel
    Toon Van Cauteren
    Koenraad Nieboer
    Inneke Willekens
    Guy Verfaillie
    Paul Evans
    Sven Macholl
    Ben Newton
    Johan de Mey
    [J]. European Radiology, 2015, 25 : 1023 - 1031
  • [43] The impact of iterative reconstruction on image quality and radiation dose in thoracic and abdominal CT
    Kalmar, Peter I.
    Quehenberger, Franz
    Steiner, Juergen
    Lutfi, Andre
    Bohlsen, Dennis
    Talakic, Emina
    Hassler, Eva Maria
    Schoellnast, Helmut
    [J]. EUROPEAN JOURNAL OF RADIOLOGY, 2014, 83 (08) : 1416 - 1420
  • [44] Dose reduction and image quality in CT examinations using an iterative reconstruction algorithm: a phantom study
    Guariglia, S.
    Meliado, G.
    Zivelonghi, E.
    Pinali, L.
    Montemezzi, S.
    Cavedon, C.
    [J]. BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2015, 1 (04):
  • [45] Evaluation of the effect of low tube voltage on radiation dose and image quality
    Din, Nik Norhasrina Nik
    Zainon, Rafidah
    Rahman, A. T. Abdul
    [J]. 10TH INTERNATIONAL SEMINAR ON MEDICAL PHYSICS & 1ST AMDI INTERNATIONAL ONCOLOGY SYMPOSIUM, 2017, 851
  • [46] Radiation dose reduction and image quality in pediatric abdominal CT with kVp and mAs modulation and an iterative reconstruction technique
    Kim, Jun-Hwee
    Kim, Myung-Joon
    Kim, Ha Yan
    Lee, Mi-Jung
    [J]. CLINICAL IMAGING, 2014, 38 (05) : 710 - 714
  • [47] Evaluation of an Iterative model-based reconstruction of pediatric abdominal CT with regard to image quality and radiation dose
    Aurumskjold, Marie-Louise
    Soderberg, Marcus
    Stalhammar, Fredrik
    von Steyern, Kristina Vult
    Tingberg, Anders
    Ydstrom, Kristina
    [J]. ACTA RADIOLOGICA, 2018, 59 (06) : 740 - 747
  • [48] Variation in tube voltage for pediatric neck 64VCT: Effect on radiation dose and image quality
    Chen, Li-Guo
    Wu, Ping-An
    Tu, Hsing-Yang
    Sheu, Ming-Huei
    Huang, Li-Chuan
    [J]. PLOS ONE, 2021, 16 (11):
  • [49] Assessment of Image Quality on Effects of Varying Tube Voltage and Automatic Tube Current Modulation With Hybrid and Pure Iterative Reconstruction Techniques in Abdominal/Pelvic CT A Phantom Study
    Vardhanabhuti, Varut
    Loader, Robert
    Roobottom, Carl A.
    [J]. INVESTIGATIVE RADIOLOGY, 2013, 48 (03) : 167 - 174
  • [50] Evaluation of radiation dose and image quality of CT scan for whole-body pediatric PET/CT: A phantom study
    Yang, Ching-Ching
    Liu, Shu-Hsin
    Mok, Greta S. P.
    Wu, Tung-Hsin
    [J]. MEDICAL PHYSICS, 2014, 41 (09)