Variation in tube voltage for pediatric neck 64VCT: Effect on radiation dose and image quality

被引:1
|
作者
Chen, Li-Guo [1 ]
Wu, Ping-An [1 ]
Tu, Hsing-Yang [1 ]
Sheu, Ming-Huei [1 ]
Huang, Li-Chuan [1 ,2 ]
机构
[1] Buddhist Tzu Chi Med Fdn, Hualien Tzu Chi Hosp, Dept Med Imaging, Hualien, Taiwan
[2] Tzu Chi Univ Sci & Technol, Dept Med Imaging & Radiol Sci, Hualien, Taiwan
来源
PLOS ONE | 2021年 / 16卷 / 11期
关键词
CT; PHANTOM; RECONSTRUCTION; REDUCTION; EXPOSURE;
D O I
10.1371/journal.pone.0259772
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Exposure to ionizing radiation can cause cancer, especially in children. In computed tomography (CT), a trade-off exists between the radiation dose and image quality. Few studies have investigated the effect of dose reduction on image quality in pediatric neck CT. We aimed to assess the effect of peak kilovoltage on the radiation dose and image quality in pediatric neck multidetector-row CT. Measurements were made using three phantoms representative of children aged 1, 5, and 10 years, with tube voltages of 80, 100, and 120 kilovoltage peak (kVp); tube current of 10, 40, 80, 120, 150, 200, and 250 mA; and exposure time = 0.5 s (pitch, 0.984:1). Radiation dose estimates were derived from the dose-length product with a 64-multidetector-row CT scanner. Images obtained from the control protocol (120 kVp) were compared with the 80- and 100-kVp protocols. The effective dose (ED) was determined for each protocol and compared with the 120-kVp protocol. Quantitative analysis entailed noise measurements by recording the standard deviation of attenuation for a circular 1-cm(2) region of interest placed on homogeneous soft tissue structures in the phantom. The mean noise of the various kVp protocols was compared using the unpaired Student t-test. Reduction of ED was 37.58% and 68.58% for neck CT with 100 kVp and 80 kVp, respectively. The image noise level increased with the decrease in peak kilovoltage. Noise values were higher at 80 kVp at all neck levels, but did not increase at 100 kVp, compared to 120 kVp in the three phantoms. The measured noise difference was the greatest at 80 kVp (absolute increases<2.5 HU). The subjective image quality did not differ among the protocols. Thus, reducing voltage from 120 to 80 kVp for neck CT may achieve ED reduction of 68.58%, without compromising image quality.
引用
收藏
页数:12
相关论文
共 50 条
  • [41] Effect of additional filtration on radiation dose and image quality
    Gai, Xinghui
    Yang, Dewu
    Li, Xunru
    Sui, Yan
    Liu, Kang
    Wang, Junying
    Zhao, Mengjuan
    [J]. 2022 IEEE 6TH ADVANCED INFORMATION TECHNOLOGY, ELECTRONIC AND AUTOMATION CONTROL CONFERENCE (IAEAC), 2022, : 1807 - 1811
  • [42] Reaching for better image quality and lower radiation dose in head and neck CT: advanced modeled and sinogram-affirmed iterative reconstruction in combination with tube voltage adaptation
    Schmid, Andrea I.
    Uder, Michael
    Lell, Michael M.
    [J]. DENTOMAXILLOFACIAL RADIOLOGY, 2016, 46 (01)
  • [43] EVOLUTION OF RADIATION DOSE AND IMAGE QUALITY IN A PEDIATRIC INTERVENTIONAL CARDIOLOGY SYSTEM
    Ubeda, Carlos
    Miranda, Patricia
    Dalmazzo, Dandaro
    [J]. INTERCIENCIA, 2014, 39 (07) : 518 - 523
  • [44] Balancing radiation dose and image quality: Clinical applications of neck volume CT
    Russell, M. T.
    Fink, J. R.
    Rebeles, F.
    Kanal, K.
    Ramos, M.
    Anzai, Y.
    [J]. AMERICAN JOURNAL OF NEURORADIOLOGY, 2008, 29 (04) : 727 - 731
  • [45] Effect of X-ray Tube Parameters and Iodine Concentration on Image Quality and Radiation Dose in Cerebral Pediatric and Adult CT Angiography A Phantom Study
    Papadakis, Antonios E.
    Perisinakis, Kostas
    Raissaki, Maria
    Damilakis, John
    [J]. INVESTIGATIVE RADIOLOGY, 2013, 48 (04) : 192 - 199
  • [46] Low tube voltage prospectively ECG-triggered coronary CT angiography: a systematic review of image quality and radiation dose
    Tan, Sock Keow
    Yeong, Chai Hong
    Aman, Raja Rizal Azman Raja
    Ng, Kwan Hoong
    Aziz, Yang Faridah Abdul
    Chee, Kok Han
    Sun, Zhonghua
    [J]. BRITISH JOURNAL OF RADIOLOGY, 2018, 91 (1088):
  • [47] Image quality and radiation dose of low-tube-voltage CT with reduced contrast media for right adrenal vein imaging
    Takahashi, Yuki
    Ota, Hideki
    Omura, Kensuke
    Dendo, Yutaka
    Otani, Katharina
    Matsuura, Tomonori
    Kitami, Masahiro
    Seiji, Kazumasa
    Tezuka, Yuta
    Nezu, Masahiro
    Ono, Yoshikiyo
    Morimoto, Ryo
    Satoh, Fumitoshi
    Takase, Kei
    [J]. EUROPEAN JOURNAL OF RADIOLOGY, 2018, 98 : 150 - 157
  • [48] Radiation dose reduction and image quality in pediatric paranasal sinus CT: with automatic tube current modulation and iterative reconstruction technique
    Chen, Xiao-Fang
    Zhang, Yi-Chi
    Ding, Ning
    Liu, Li
    Ji, Yi-Ding
    Zhang, Chao
    Chi, Jing
    [J]. RADIATION PROTECTION DOSIMETRY, 2024, 200 (15) : 1470 - 1476
  • [49] Multidetector CT of bronchiectasis: Effect of radiation dose on image quality
    Yi, CA
    Lee, KS
    Kim, TS
    Han, DH
    Sung, YM
    Kim, S
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2003, 181 (02) : 501 - 505
  • [50] A population of pediatric phantoms to investigate the tradeoff between radiation dose and image quality
    O'Reilly, Shannon
    Li, Wenxiao
    Frey, Eric
    Treves, S. Ted
    Sgouros, George
    Fahey, Frederic
    Bolch, Wesley
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2014, 55