Lead isotopic analysis of Antarctic snow by quadrupole ICP-MS using a total-consumption sample introduction system

被引:4
|
作者
Ardini, Francisco [1 ]
Bazzano, Andrea [1 ]
Grotti, Marco [1 ]
机构
[1] Univ Genoa, Dept Chem & Ind Chem, Via Dodecaneso 31, I-16146 Genoa, Italy
关键词
PLASMA-MASS SPECTROMETRY; DYNAMIC REACTION CELL; DOME-C; SHEATHING GAS; LAW DOME; DUST; METALS; EPICA; CORE; CLIMATE;
D O I
10.1039/c8ja00296g
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Measurement of lead isotope ratios in snow samples using inductively coupled plasma-dynamic reaction cell-mass spectrometry (ICP-DRC-MS) at a low sample consumption rate was thoroughly studied. Sample volumes were reduced from 20.0 to 0.200 mL by freeze-drying, and the pre-concentrates were efficiently introduced into the plasma source at 20 mu L min(-1) by using the heated torch-integrated sample introduction system (hTISIS). In addition, the DRC was pressurised with neon to improve the precision by collisional damping. Sensitivity was maximised by optimising the operating parameters of the sample introduction system (nebulizer/sheathing gas flow rate ratio), the reaction cell (rod offset and damping gas flow rate) and the signal measurement conditions. The developed method was then characterised in terms of the analytical working range, accuracy and uncertainty estimation. Under the optimal conditions, isotope ratios did not differ in the 0.1-20 mu g L-1 range. Internal precision, expressed as relative standard deviation, varied from 0.12 to 0.18% (c = 10 mu g L-1; n = 12), while the external precision was <0.1%. The potential influence of the pre-concentration procedure on the analytical data was carefully investigated in terms of blank contribution, recovery and isotopic fractionation. The lowest Pb concentration needed in the samples to limit the influence of the blanks within an experimental precision of 0.5% was 4.4 pg g(-1), and the recovery of five standard reference solutions with a Pb concentration of 100 ng L-1 was 90 +/- 5%, with no significant isotopic fractionation. The developed ICP-DRC-MS method was finally applied to a number of representative Antarctic snow samples previously characterised by multi-collector ICP-MS, obtaining a satisfactory agreement between the data provided by the two techniques and analytical results consistent with literature data on the glaciochemistry of Antarctic ice cores.
引用
收藏
页码:2124 / 2132
页数:9
相关论文
共 50 条
  • [1] Determination of major elements in Antarctic snow by inductively coupled plasma optical emission spectrometry using a total-consumption sample introduction system
    Bertinetti, Stefano
    Ardini, Francisco
    Caiazzo, Laura
    Grotti, Marco
    Spectrochimica Acta - Part B Atomic Spectroscopy, 2021, 181
  • [2] Determination of major elements in Antarctic snow by inductively coupled plasma optical emission spectrometry using a total-consumption sample introduction system
    Bertinetti, Stefano
    Ardini, Francisco
    Caiazzo, Laura
    Grotti, Marco
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2021, 181
  • [3] Ultratrace analysis of Antarctic snow samples by reaction cell inductively coupled plasma mass spectrometry using a total-consumption micro- sample-introduction system
    Grotti, Marco
    Soggia, Francesco
    Luis Todoli, Jose
    ANALYST, 2008, 133 (10) : 1388 - 1394
  • [4] Analysis of whole blood by ICP-MS equipped with a high temperature total sample consumption system
    Canabate, Agueda
    Garcia-Ruiz, Esperanza
    Resano, Martin
    Todoli, Jose-Luis
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2017, 32 (01) : 78 - 87
  • [5] High transport efficiency of nanoparticles through a total-consumption sample introduction system and its beneficial application for particle size evaluation in single-particle ICP-MS
    Shin-ichi Miyashita
    Hiroaki Mitsuhashi
    Shin-ichiro Fujii
    Akiko Takatsu
    Kazumi Inagaki
    Toshiyuki Fujimoto
    Analytical and Bioanalytical Chemistry, 2017, 409 : 1531 - 1545
  • [6] High transport efficiency of nanoparticles through a total-consumption sample introduction system and its beneficial application for particle size evaluation in single-particle ICP-MS
    Miyashita, Shin-ichi
    Mitsuhashi, Hiroaki
    Fujii, Shin-ichiro
    Takatsu, Akiko
    Inagaki, Kazumi
    Fujimoto, Toshiyuki
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2017, 409 (06) : 1531 - 1545
  • [7] A convenient sample introduction system for ICP, ICP-MS, and AA
    McLaughlin, Roger
    Cheese, Paula
    Ding, Min
    Ptolemy, Adam
    Conn, April
    Brindle, Ian
    AMERICAN LABORATORY, 2006, 38 (18) : 13 - +
  • [9] Lead isotope ratio analysis of bullet samples by using quadrupole ICP-MS
    Tamura, Shu-ichi
    Hokura, Akiko
    Oishi, Masahiro
    Nakai, Izumi
    BUNSEKI KAGAKU, 2006, 55 (11) : 827 - 834
  • [10] Determination of lead isotope ratios in environmental matrices by quadrupole ICP-MS working at low sample consumption rates
    Bazzano, Andrea
    Grotti, Marco
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2014, 29 (05) : 926 - 933