Lead isotopic analysis of Antarctic snow by quadrupole ICP-MS using a total-consumption sample introduction system

被引:4
|
作者
Ardini, Francisco [1 ]
Bazzano, Andrea [1 ]
Grotti, Marco [1 ]
机构
[1] Univ Genoa, Dept Chem & Ind Chem, Via Dodecaneso 31, I-16146 Genoa, Italy
关键词
PLASMA-MASS SPECTROMETRY; DYNAMIC REACTION CELL; DOME-C; SHEATHING GAS; LAW DOME; DUST; METALS; EPICA; CORE; CLIMATE;
D O I
10.1039/c8ja00296g
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Measurement of lead isotope ratios in snow samples using inductively coupled plasma-dynamic reaction cell-mass spectrometry (ICP-DRC-MS) at a low sample consumption rate was thoroughly studied. Sample volumes were reduced from 20.0 to 0.200 mL by freeze-drying, and the pre-concentrates were efficiently introduced into the plasma source at 20 mu L min(-1) by using the heated torch-integrated sample introduction system (hTISIS). In addition, the DRC was pressurised with neon to improve the precision by collisional damping. Sensitivity was maximised by optimising the operating parameters of the sample introduction system (nebulizer/sheathing gas flow rate ratio), the reaction cell (rod offset and damping gas flow rate) and the signal measurement conditions. The developed method was then characterised in terms of the analytical working range, accuracy and uncertainty estimation. Under the optimal conditions, isotope ratios did not differ in the 0.1-20 mu g L-1 range. Internal precision, expressed as relative standard deviation, varied from 0.12 to 0.18% (c = 10 mu g L-1; n = 12), while the external precision was <0.1%. The potential influence of the pre-concentration procedure on the analytical data was carefully investigated in terms of blank contribution, recovery and isotopic fractionation. The lowest Pb concentration needed in the samples to limit the influence of the blanks within an experimental precision of 0.5% was 4.4 pg g(-1), and the recovery of five standard reference solutions with a Pb concentration of 100 ng L-1 was 90 +/- 5%, with no significant isotopic fractionation. The developed ICP-DRC-MS method was finally applied to a number of representative Antarctic snow samples previously characterised by multi-collector ICP-MS, obtaining a satisfactory agreement between the data provided by the two techniques and analytical results consistent with literature data on the glaciochemistry of Antarctic ice cores.
引用
收藏
页码:2124 / 2132
页数:9
相关论文
共 50 条
  • [41] Comparison of interface cones for analysis of sodium-rich samples using quadrupole ICP-MS
    Brenner, Isaac
    Pacheco, Jaime
    Valiente, Manuel
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2009, 24 (11) : 1558 - 1563
  • [42] Improving sensitivity and detection limits in ICP-MS with a novel high-efficiency sample introduction system
    Debrah, E
    Légère, G
    ATOMIC SPECTROSCOPY, 1999, 20 (02) : 73 - 77
  • [43] Effect of the sample introduction system in ICP-MS on the formation of both oxygenated polyatomic and doubly charged ions
    Tubaro, F
    Barbangelo, F
    Toniolo, R
    Di Narda, F
    Bontempelli, G
    ANNALI DI CHIMICA, 1999, 89 (11-12) : 863 - 872
  • [44] Spectroscopy tutorial - A beginner's guide to ICP-MS - Part II: The sample-introduction system
    Thomas, R
    SPECTROSCOPY, 2001, 16 (05) : 56 - +
  • [45] Single cell ICP-MS using on line sample introduction systems: Current developments and remaining challenges
    Corte-Rodriguez, M.
    Alvarez-Fernandez, R.
    Garcia-Cancela, P.
    Montes-Bayon, M.
    Bettmer, J.
    TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 2020, 132
  • [46] Analysis of wasteform leachates by ICP-MS using isotopic composition tables for interference correction.
    Warren, AR
    Cummings, DG
    Johnson, SG
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2000, 219 : U88 - U88
  • [47] Determination of trace elements in fuel ethanol by ICP-MS using direct sample introduction by a microconcentric nebulizer
    Tormen, Luciano
    Chaves, Eduardo S.
    Saint'Pierre, Tatiana D.
    Frescura, Vera L. A.
    Curtius, Adilson J.
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2008, 23 (09) : 1300 - 1304
  • [48] Design, Optimization, and Application of a 3D-Printed Polymer Sample Introduction System for the ICP-MS Analysis of Nanoparticles and Cells
    Kajner, Gyula
    Belteki, Adam
    Cseh, Martin
    Geretovszky, Zsolt
    Ajtai, Tibor
    Barna, Lilla
    Deli, Maria A.
    Pap, Bernadett
    Maroti, Gergely
    Galbacs, Gabor
    NANOMATERIALS, 2023, 13 (23)
  • [49] Determination of plutonium and its isotopic ratio in marine sediment samples using quadrupole ICP-MS with the shield torch system under normal plasma conditions
    Jian Zheng
    Masatoshi Yamada
    Zhongliang Wang
    Tatsuo Aono
    Masashi Kusakabe
    Analytical and Bioanalytical Chemistry, 2004, 379 : 532 - 539
  • [50] Trace elemental analysis of glass and paint samples of forensic interest by ICP-MS using Laser Ablation solid sample introduction
    Almirall, JR
    Trejos, T
    Hobbs, A
    Furton, K
    SENSORS, AND COMMAND, CONTROL, COMMUNICATIONS, AND INTELLIGENCE (C3I) TECHNOLOGIES FOR HOMELAND DEFENSE AND LAW ENFORCEMENT II, 2003, 5071 : 193 - 204