Low Dose, Low Noise and High Resolution Volume of Interest (VOI) Imaging in C-Arm Flat-Detector CT

被引:0
|
作者
Kolditz, Daniel [1 ]
Kyriakou, Yiannis [1 ]
Kalender, Willi A. [1 ]
机构
[1] Univ Erlangen Nurnberg, IMP, D-91052 Erlangen, Germany
关键词
Volume of interest (VOI) imaging; Flat-detector CT (FDCT); Cone-beam CT; C-arm CT; Dose; Resolution; Noise; Truncation; BEAM BREAST CT; COMPUTED-TOMOGRAPHY; RECONSTRUCTION; FEASIBILITY; SCATTER; REGION;
D O I
10.1117/12.844129
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
The high flexibility of C-arm flat-detector computed tomography (FDCT) is used in a volume of interest (VOI) imaging method to handle the challenges of increasing spatial resolution, reducing noise and saving dose. A low-dose overview scan of the object and a high-dose scan of an arbitrary VOI are combined. The first scan is adequate for orientation to select the VOI and the second scan assures high image quality in the VOI. The combination is based on a forward projection of the reconstructed overview volume and the measured VOI data in the raw data domain. Differences in the projection values are matched before a standard Feldkamp-type reconstruction is performed. In simulations, spatial resolution, noise and contrast detectability were evaluated. Measurements of an anthropomorphic phantom were used to validate the proposed method for realistic application. In Monte Carlo dose simulations the dose reduction potential was investigated. By combination of the two scans an image is generated which covers the whole object and provides the actual VOI at high image quality. Spatial resolution was increased whereas noise was decreased from outside to inside the VOI, e. g. for the simulations from 0.8 lp/mm to 3.0 lp/mm and from 39 HU to 18 HU, respectively. Simultaneously, the cumulative dose for this two-scan procedure was significantly reduced in comparison to a conventional high dose scan, e. g. for the performed simulations and measurements by about 95 %. The proposed VOI approach offers significant benefits with respect to high-resolution and low-contrast imaging of a VOI at reduced dose.
引用
收藏
页数:12
相关论文
共 50 条
  • [1] Volume-of-interest, (VOI) imaging in C-arm flat-detector CT for high image quality at reduced dose
    Kolditz, Daniel
    Kyriakou, Yiannis
    Kalender, Willi A.
    [J]. MEDICAL PHYSICS, 2010, 37 (06) : 2719 - 2730
  • [2] Low-Dose Volume-of-Interest C-Arm CT Imaging of Intracranial Stents and Flow Diverters
    Yang, P.
    Ahmed, A.
    Schafer, S.
    Niemann, D.
    Aagaard-Kienitz, B.
    Royalty, K.
    Strother, C.
    [J]. AMERICAN JOURNAL OF NEURORADIOLOGY, 2016, 37 (04) : 648 - 654
  • [3] Patient-bounded extrapolation using low-dose priors for volume-of-interest imaging in C-arm CT
    Xia, Y.
    Bauer, S.
    Maier, A.
    Berger, M.
    Hornegger, J.
    [J]. MEDICAL PHYSICS, 2015, 42 (04) : 1787 - 1796
  • [4] The comparison of radiation dose between C-arm flat-detector CT (DynaCT) and multi-slice CT (MSCT): A phantom study
    Bai, Mei
    Liu, Bin
    Mu, Hongqiu
    Liu, Xianghua
    Jiang, Yuan
    [J]. EUROPEAN JOURNAL OF RADIOLOGY, 2012, 81 (11) : 3577 - 3580
  • [5] Brain imaging with a flat detector C-arm
    Soederman, M.
    Babic, D.
    Holmin, S.
    Andersson, T.
    [J]. NEURORADIOLOGY, 2008, 50 (10) : 863 - 868
  • [6] Preliminary performance of image quality for a low-dose C-arm CT system with a flat-panel detector
    Cha, Bo Kyung
    Seo, Chang-Woo
    Yang, Keedong
    Jeon, Seongchae
    Huh, Young
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2015, 784 : 565 - 569
  • [7] Neuroradiologic Applications with Routine C-arm Flat Panel Detector CT: Evaluation of Patient Dose Measurements
    Kyriakou, Y.
    Richter, G.
    Doerfler, A.
    Kalender, W. A.
    [J]. AMERICAN JOURNAL OF NEURORADIOLOGY, 2008, 29 (10) : 1930 - 1936
  • [8] Comparison of extended field-of-view reconstructions in C-arm flat-detector CT using patient size, shape or attenuation information
    Kolditz, Daniel
    Meyer, Michael
    Kyriakou, Yiannis
    Kalender, Willi A.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (01): : 39 - 56
  • [9] Artifact Reduction of Different Metallic Implants in Flat Detector C-Arm CT
    Hung, S. -C.
    Wu, C. -C.
    Lin, C. -J.
    Guo, W. -Y.
    Luo, C. -B.
    Chang, F. -C.
    Chang, C. -Y.
    [J]. AMERICAN JOURNAL OF NEURORADIOLOGY, 2014, 35 (07) : 1288 - 1292
  • [10] Developing low-dose C-arm CT imaging for temporomandibular joint (TMJ) disorder in interventional radiology
    Zhu, Xiaowei
    Felice, Marc
    Johnson, Laura
    Sarmiento, Marily
    Cahill, Anne Marie
    [J]. PEDIATRIC RADIOLOGY, 2011, 41 (04) : 476 - 482