The comparison of radiation dose between C-arm flat-detector CT (DynaCT) and multi-slice CT (MSCT): A phantom study

被引:63
|
作者
Bai, Mei [1 ]
Liu, Bin [1 ]
Mu, Hongqiu [2 ]
Liu, Xianghua [1 ]
Jiang, Yuan [2 ]
机构
[1] Capital Med Univ, Dept Biomed Engn, Xuanwu Hosp, Beijing 100053, Peoples R China
[2] Siemens Ltd, Pudong New Area, Shanghai 201318, Peoples R China
关键词
C-arm flat panel CT; Multi-slice CT; Radiation dose; SOFT-TISSUE VISUALIZATION; COMPUTED-TOMOGRAPHY; SYSTEM;
D O I
10.1016/j.ejrad.2011.09.006
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To assess and compare the radiation dose of DynaCT in phantoms to that of MSCT. Methods and materials: A male Anderson Radiation Therapy 200 phantom with embedded thermoluminescence dosimeters (TLDs) was scanned with DynaCT (Axiom Artis dTA, Siemens Healthcare, Forchheim, Germany) and MSCTs (MSCT1: Lightspeed VCT, GE, Milwaukee, USA; MSCT2: Sensation Cardiac 64, Siemens Healthcare, Erlangen, Germany), respectively. For DynaCT, radiation exposure data with automatic exposure control were obtained from scanning of head, chest and abdomen. For MSCT, scanning protocols for head, chest, and abdomen were examined under conventional exposure conditions. Effective doses were calculated from LiF-TLD measurements according to ICRP103. A Catphan phantom (Phantom Lab, USA) was used to evaluate the spatial resolution and low contrast detectability of DynaCT and MSCT. Results: The effective doses for DynaCT from head, chest and abdomen scanning were 1.18, 7.32, 7.48 mSv (20 s scan) and 0.85, 6.01, 7.04 mSv (8 s scan); the effective doses for MSCT from head, chest and abdomen scanning were 3.33, 7.62 and 8.42 mSv (MSCT1), and 1.89, 7.52, 8.23 mSv (MSCT2). Significant difference between the organ doses from DynaCT and from MSCT (p<.05) was shown. The spatial resolution of 12 lp/cm was achieved and it was able to recognize a 3 mm low contrast object at 0.5% contrast level in DynaCT, which was on the same level as in the MSCT images. Conclusion: The phantom study indicates that DynaCT applies significantly less dose to patient and achieves similar spatial resolution and low contrast detectability to standard diagnostic MSCT. (C) 2011 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:3577 / 3580
页数:4
相关论文
共 50 条
  • [1] Designing a phantom for dose evaluation in multi-slice CT
    Abboud, Samir
    Badal, Andreu
    Stern, Stanley H.
    Kyprianou, Iacovos S.
    [J]. MEDICAL IMAGING 2010: PHYSICS OF MEDICAL IMAGING, 2010, 7622
  • [2] Low Dose, Low Noise and High Resolution Volume of Interest (VOI) Imaging in C-Arm Flat-Detector CT
    Kolditz, Daniel
    Kyriakou, Yiannis
    Kalender, Willi A.
    [J]. MEDICAL IMAGING 2010: PHYSICS OF MEDICAL IMAGING, 2010, 7622
  • [3] Volume-of-interest, (VOI) imaging in C-arm flat-detector CT for high image quality at reduced dose
    Kolditz, Daniel
    Kyriakou, Yiannis
    Kalender, Willi A.
    [J]. MEDICAL PHYSICS, 2010, 37 (06) : 2719 - 2730
  • [4] Radiation exposure of the head, midface and pelvis in multi-slice CT (MSCT):: Comparison with single-slice CT (SSCT)
    Wedegärtner, U
    Thurmann, H
    Schmidt, R
    Adam, G
    [J]. ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2003, 175 (02): : 234 - 238
  • [5] Radiation dose in CT of the heart for coronary heart disease and CT of the lung for pulmonary embolism: Comparisons between single-slice detector CT, multi-slice detector CT and EBT
    Hidajat, N
    Wolf, M
    Rademaker, J
    Knollmann, FD
    Oestmann, JW
    Felix, R
    [J]. RADIOLOGY, 2000, 217 : 439 - 439
  • [6] Patient radiation dose in multi-slice CT: Dependence on detector configuration and system manufacturer
    McCollough, CH
    Zink, FE
    [J]. RADIOLOGY, 2000, 217 : 506 - 506
  • [7] CTA of the coronary arteries: Comparison of radiation exposure with EBCT and multi-slice detector CT
    Knollmann, FD
    Hidajat, N
    Felix, R
    [J]. RADIOLOGY, 2000, 217 : 374 - 374
  • [8] Comparison of extended field-of-view reconstructions in C-arm flat-detector CT using patient size, shape or attenuation information
    Kolditz, Daniel
    Meyer, Michael
    Kyriakou, Yiannis
    Kalender, Willi A.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (01): : 39 - 56
  • [9] Radiation dose from helical CT in children: Comparison of multi-slice and single-slice protocols
    Frush, DP
    Yoshizumi, TT
    Paulson, EK
    Johnson, GA
    [J]. RADIOLOGY, 2001, 221 : 246 - 246
  • [10] Imaging of cochlear implant electrode array with flat-detector CT and conventional multislice CT: comparison of image quality and radiation dose
    Struffert, Tobias
    Hertel, Victoria
    Kyriakou, Yannis
    Krause, Jens
    Engelhorn, Tobias
    Schick, Bernhard
    Iro, Heinrich
    Hornung, Joachim
    Doerfler, Arnd
    [J]. ACTA OTO-LARYNGOLOGICA, 2010, 130 (04) : 443 - 452