Multifaceted bench comparative evaluation of latest intensive care unit ventilators

被引:26
|
作者
Garnier, M. [1 ,5 ]
Quesnel, C. [1 ,5 ]
Fulgencio, J. -P. [1 ]
Degrain, M. [4 ]
Carteaux, G. [8 ]
Bonnet, F. [1 ,5 ]
Similowski, T. [2 ,3 ,5 ,6 ]
Demoule, A. [2 ,3 ,5 ,7 ]
机构
[1] Hop Tenon, Anaesthesiol & Intens Care Dept, F-75020 Paris, France
[2] Grp Hosp Pitie Salpetriere, Med Intens Care Unit, F-75634 Paris, France
[3] Grp Hosp Pitie Salpetriere, Div Resp, F-75634 Paris, France
[4] AP HP, Agence Gen Equipements & Prod Sante, Paris, France
[5] Univ Paris 06, Fac Med Pierre & Marie Curie, Paris, France
[6] Univ Paris 06, ER10, Paris, France
[7] Univ Paris 06, INSERM, U974, Paris, France
[8] Ctr Hosp Albert Chenevier Henri Mondor, AP HP, Med Intens Care Unit, Creteil, France
关键词
bench study; mechanical ventilation; non-invasive ventilation; quality improvement; ventilator performance; RESPIRATORY-DISTRESS-SYNDROME; PRESSURE SUPPORT VENTILATION; END-EXPIRATORY PRESSURE; ACUTE LUNG INJURY; NONINVASIVE VENTILATION; PERFORMANCE-CHARACTERISTICS; ICU VENTILATORS; TIDAL VOLUME; ANESTHESIA VENTILATORS; MECHANICAL VENTILATORS;
D O I
10.1093/bja/aev028
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Background: Independent bench studies using specific ventilation scenarios allow testing of the performance of ventilators in conditions similar to clinical settings. The aims of this study were to determine the accuracy of the latest generation ventilators to deliver chosen parameters in various typical conditions and to provide clinicians with a comprehensive report on their performance. Methods: Thirteen modern intensive care unit ventilators were evaluated on the ASL5000 test lung with and without leakage for: (i) accuracy to deliver exact tidal volume (VT) and PEEP in assist-control ventilation (ACV); (ii) performance of trigger and pressurization in pressure support ventilation (PSV); and (iii) quality of non-invasive ventilation algorithms. Results: In ACV, only six ventilators delivered an accurate VT and nine an accurate PEEP. Eleven devices failed to compensate VT and four the PEEP in leakage conditions. Inspiratory delays differed significantly among ventilators in invasive PSV (range 75-149 ms, P=0.03) and non-invasive PSV (range 78-165 ms, P<0.001). The percentage of the ideal curve (concomitantly evaluating the pressurization speed and the levels of pressure reached) also differed significantly (range 57-86% for invasive PSV, P=0.04; and 60-90% for non-invasive PSV, P<0.001). Non-invasive ventilation algorithms efficiently prevented the decrease in pressurization capacities and PEEP levels induced by leaks in, respectively, 10 and 12 out of the 13 ventilators. Conclusions: We observed real heterogeneity of performance amongst the latest generation of intensive care unit ventilators. Although non-invasive ventilation algorithms appear to maintain adequate pressurization efficiently in the case of leakage, basic functions, such as delivered VT in ACV and pressurization in PSV, are often less reliable than the values displayed by the device suggest.
引用
收藏
页码:89 / 98
页数:10
相关论文
共 50 条
  • [1] Multifaceted Bench Comparative Evaluation Of Latest Intensive Care Unit Ventilators Performance
    Garnier, M.
    Quesnel, C.
    Fulgencio, J. -P.
    Degrain, M.
    Carteaux, G.
    Bonnet, F.
    Similowski, T.
    Demoule, A.
    [J]. AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2014, 189
  • [2] Evaluation of transport ventilators used by a paediatric intensive care transport services: a bench study
    Manning, G.
    Ray, S.
    Chigaru, L.
    [J]. ANAESTHESIA, 2018, 73 : 31 - 31
  • [3] Bacterial Contamination of Ventilators in the Intensive Care Unit
    Motamedi, Mohammad Hosein Kalantar
    Jarineshin, Hashem
    Mehrvarz, Shaban
    Danial, Zahra
    Rasouli, Hamid Reza
    Rashidi, Ali
    [J]. TRAUMA MONTHLY, 2018, 23 (02)
  • [4] Telerobotic Operation of Intensive Care Unit Ventilators
    Vagvolgyi, Balazs P.
    Khrenov, Mikhail
    Cope, Jonathan
    Deguet, Anton
    Kazanzides, Peter
    Manzoor, Sajid
    Taylor, Russell H.
    Krieger, Axel
    [J]. FRONTIERS IN ROBOTICS AND AI, 2021, 8
  • [5] A Bench Evaluation of Eight Home-Care Ventilators
    Martinez Castro, Sara
    Belda Nacher, Francisco Javier
    Delgado Navarro, Carlos
    Puig Bernabeu, Jaume
    [J]. RESPIRATORY CARE, 2021, 66 (10) : 1531 - 1541
  • [6] Bench Evaluation of 7 Home-Care Ventilators
    Blakeman, Thomas C.
    Rodriquez, Dario
    Hanseman, Dennis
    Branson, Richard D.
    [J]. RESPIRATORY CARE, 2011, 56 (11) : 1791 - 1798
  • [7] Performance characteristics of five new anesthesia ventilators and four intensive care ventilators in pressure-support mode - A comparative bench study
    Jaber, Samir
    Tassaux, Didier
    Sebbane, Mustapha
    Pouzeratte, Yvan
    Battisti, Anne
    Capdevila, Xavier
    Eledjam, Jean-Jacques
    Jolliet, Philippe
    [J]. ANESTHESIOLOGY, 2006, 105 (05) : 944 - 952
  • [8] A multifaceted approach to intensive care unit capacity
    Verdonk, Franck
    Zacharowski, Kai
    Ahmed, Aamer
    Orliaguet, Gilles
    Pottecher, Julien
    [J]. LANCET PUBLIC HEALTH, 2021, 6 (07): : E448 - E448
  • [9] A bench study of intensive-care-unit ventilators: new versus old and turbine-based versus compressed gas-based ventilators
    Arnaud W. Thille
    Aissam Lyazidi
    Jean-Christophe M. Richard
    Fabrice Galia
    Laurent Brochard
    [J]. Intensive Care Medicine, 2009, 35 : 1368 - 1376
  • [10] A bench study of intensive-care-unit ventilators: new versus old and turbine-based versus compressed gas-based ventilators
    Thille, Arnaud W.
    Lyazidi, Aissam
    Richard, Jean-Christophe M.
    Galia, Fabrice
    Brochard, Laurent
    [J]. INTENSIVE CARE MEDICINE, 2009, 35 (08) : 1368 - 1376