Bench Evaluation of 7 Home-Care Ventilators

被引:21
|
作者
Blakeman, Thomas C. [1 ]
Rodriquez, Dario [2 ]
Hanseman, Dennis [1 ]
Branson, Richard D. [1 ]
机构
[1] Univ Cincinnati, Med Ctr, Dept Surg, Div Trauma Crit Care, Cincinnati, OH 45267 USA
[2] Univ Cincinnati, Med Ctr, USAF, Ctr Sustainment Trauma & Readiness Skills, Cincinnati, OH 45267 USA
关键词
mechanical ventilation; portable ventilator; home care; triggering; battery duration; MECHANICAL VENTILATORS; PERFORMANCE;
D O I
10.4187/respcare.01176
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
BACKGROUND: Portable ventilators continue to decrease in size while increasing in performance. We bench-tested the triggering, battery duration, and tidal volume (V(T)) of 7 portable ventilators: LTV 1000, LTV 1200, Puritan Bennett 540, Trilogy, Vela, iVent 101, and HT50. METHODS: We tested triggering with a modified dual-chamber test lung to simulate spontaneous breathing with weak, normal, and strong inspiratory effort. We measured battery duration by fully charging the battery and operating the ventilator with a V(T) of 500 mL, a respiratory rate of 20 breaths/min, and PEEP of 5 cm H(2)O until breath-delivery ceased. We tested V(T) accuracy with pediatric ventilation scenarios (V(T) 50 mL or 100 mL, respiratory rate 50 breaths/min, inspiratory time 0.3 s, and PEEP 5 cm H(2)O) and an adult ventilation scenario (V(T) 400 mL, respiratory rate 30 breaths/min, inspiratory time 0.5 s, and PEEP 5 cm H(2)O). We measured and analyzed airway pressure, volume, and flow signals. RESULTS: At the adult settings the measured V(T) range was 362-426 mL. On the pediatric settings the measured V(T) range was 51-182 mL at the set V(T) of 50 mL, and 90-141 mL at the set V(T) of 100 mL. The V(T) delivered by the Vela at both the 50 mL and 100 mL, and by the HT50 at 100 mL, did not meet the American Society for Testing and Materials standard for V(T) accuracy. Triggering response and battery duration ranged widely among the tested ventilators. CONCLUSIONS: There was wide variability in battery duration and triggering sensitivity. Five of the ventilators performed adequately in V(T) delivery across several settings. The combination of high respiratory rate and low V(T) presented problems for 2 of the ventilators. Key words: mechanical ventilation; portable ventilator; home care; triggering; battery duration. [Respir Care 2011;56(11): 1791-1798. (C) 2011 Daedalus Enterprises]
引用
收藏
页码:1791 / 1798
页数:8
相关论文
共 50 条
  • [1] A Bench Evaluation of Eight Home-Care Ventilators
    Martinez Castro, Sara
    Belda Nacher, Francisco Javier
    Delgado Navarro, Carlos
    Puig Bernabeu, Jaume
    [J]. RESPIRATORY CARE, 2021, 66 (10) : 1531 - 1541
  • [2] Efficiency of quality of triggering system and pressurization of home care ventilators. A bench
    Baboi L.
    Guegan S.
    Guérin C.
    [J]. Intensive Care Medicine Experimental, 3 (Suppl 1)
  • [3] Multifaceted bench comparative evaluation of latest intensive care unit ventilators
    Garnier, M.
    Quesnel, C.
    Fulgencio, J. -P.
    Degrain, M.
    Carteaux, G.
    Bonnet, F.
    Similowski, T.
    Demoule, A.
    [J]. BRITISH JOURNAL OF ANAESTHESIA, 2015, 115 (01) : 89 - 98
  • [4] HOME-CARE PROGRAMS
    GUTHRIE, EH
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 1964, 270 (10): : 537 - &
  • [5] HOME-CARE PROGRAMS
    不详
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1958, 166 (15): : 1868 - 1868
  • [6] THE HOME-CARE TEAM
    不详
    [J]. LANCET, 1955, 1 (JAN22): : 191 - 192
  • [7] The Personal Impact of Home-Care Nursing An Alternative Perspective to Home-Care Satisfaction
    Porter, Eileen J.
    [J]. RESEARCH IN GERONTOLOGICAL NURSING, 2008, 1 (02) : 105 - 115
  • [8] A bench evaluation of fraction of oxygen in air delivery and tidal volume accuracy in home care ventilators available for hospital use
    Baboi, Loredana
    Subtil, Fabien
    Guerin, Claude
    [J]. JOURNAL OF THORACIC DISEASE, 2016, 8 (12) : 3639 - 3647
  • [9] Multifaceted Bench Comparative Evaluation Of Latest Intensive Care Unit Ventilators Performance
    Garnier, M.
    Quesnel, C.
    Fulgencio, J. -P.
    Degrain, M.
    Carteaux, G.
    Bonnet, F.
    Similowski, T.
    Demoule, A.
    [J]. AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2014, 189
  • [10] HOME-CARE FOR DIABETIC CHILDREN
    VANELLI, M
    [J]. RIVISTA ITALIANA DI PEDIATRIA-ITALIAN JOURNAL OF PEDIATRICS, 1993, 19 (04): : 380 - 383