Multifaceted bench comparative evaluation of latest intensive care unit ventilators

被引:26
|
作者
Garnier, M. [1 ,5 ]
Quesnel, C. [1 ,5 ]
Fulgencio, J. -P. [1 ]
Degrain, M. [4 ]
Carteaux, G. [8 ]
Bonnet, F. [1 ,5 ]
Similowski, T. [2 ,3 ,5 ,6 ]
Demoule, A. [2 ,3 ,5 ,7 ]
机构
[1] Hop Tenon, Anaesthesiol & Intens Care Dept, F-75020 Paris, France
[2] Grp Hosp Pitie Salpetriere, Med Intens Care Unit, F-75634 Paris, France
[3] Grp Hosp Pitie Salpetriere, Div Resp, F-75634 Paris, France
[4] AP HP, Agence Gen Equipements & Prod Sante, Paris, France
[5] Univ Paris 06, Fac Med Pierre & Marie Curie, Paris, France
[6] Univ Paris 06, ER10, Paris, France
[7] Univ Paris 06, INSERM, U974, Paris, France
[8] Ctr Hosp Albert Chenevier Henri Mondor, AP HP, Med Intens Care Unit, Creteil, France
关键词
bench study; mechanical ventilation; non-invasive ventilation; quality improvement; ventilator performance; RESPIRATORY-DISTRESS-SYNDROME; PRESSURE SUPPORT VENTILATION; END-EXPIRATORY PRESSURE; ACUTE LUNG INJURY; NONINVASIVE VENTILATION; PERFORMANCE-CHARACTERISTICS; ICU VENTILATORS; TIDAL VOLUME; ANESTHESIA VENTILATORS; MECHANICAL VENTILATORS;
D O I
10.1093/bja/aev028
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Background: Independent bench studies using specific ventilation scenarios allow testing of the performance of ventilators in conditions similar to clinical settings. The aims of this study were to determine the accuracy of the latest generation ventilators to deliver chosen parameters in various typical conditions and to provide clinicians with a comprehensive report on their performance. Methods: Thirteen modern intensive care unit ventilators were evaluated on the ASL5000 test lung with and without leakage for: (i) accuracy to deliver exact tidal volume (VT) and PEEP in assist-control ventilation (ACV); (ii) performance of trigger and pressurization in pressure support ventilation (PSV); and (iii) quality of non-invasive ventilation algorithms. Results: In ACV, only six ventilators delivered an accurate VT and nine an accurate PEEP. Eleven devices failed to compensate VT and four the PEEP in leakage conditions. Inspiratory delays differed significantly among ventilators in invasive PSV (range 75-149 ms, P=0.03) and non-invasive PSV (range 78-165 ms, P<0.001). The percentage of the ideal curve (concomitantly evaluating the pressurization speed and the levels of pressure reached) also differed significantly (range 57-86% for invasive PSV, P=0.04; and 60-90% for non-invasive PSV, P<0.001). Non-invasive ventilation algorithms efficiently prevented the decrease in pressurization capacities and PEEP levels induced by leaks in, respectively, 10 and 12 out of the 13 ventilators. Conclusions: We observed real heterogeneity of performance amongst the latest generation of intensive care unit ventilators. Although non-invasive ventilation algorithms appear to maintain adequate pressurization efficiently in the case of leakage, basic functions, such as delivered VT in ACV and pressurization in PSV, are often less reliable than the values displayed by the device suggest.
引用
收藏
页码:89 / 98
页数:10
相关论文
共 50 条
  • [11] DIMENSIONING THE PARK OF MECHANICAL VENTILATORS ON AN INTENSIVE CARE UNIT OF A HOSPITAL
    Assis, Rui
    Matos, Manuel
    [J]. 2019 6TH IEEE PORTUGUESE MEETING IN BIOENGINEERING (ENBENG), 2019,
  • [12] Knowledge translation and the multifaceted intervention in the intensive care unit
    Black, Martin D.
    Schorr, Christa
    Levy, Mitchell M.
    [J]. CRITICAL CARE MEDICINE, 2012, 40 (04) : 1324 - 1328
  • [13] Effect of Jet Nebulization on Noninvasive Positive-Pressure Ventilation Administered with Noninvasive or Intensive Care Unit Ventilators: A Bench Study
    Xu, Meng-Jiao
    Dai, Bing
    Peng, Yun
    Su, Jia
    Tan, Wei
    Zhao, Hong-Wen
    [J]. RESPIRATION, 2019, 97 (04) : 355 - 362
  • [14] Evaluation of a Multifaceted Educational Intervention to Improve Palliative Care in the Intensive Care Unit Adopting the Care and Communication Bundle
    Constantine, Lori
    [J]. JOURNAL OF HOSPICE & PALLIATIVE NURSING, 2016, 18 (06) : 519 - 525
  • [15] Probiotics in the Intensive Care Unit: from Bench to Bedside
    Barraud, D.
    Gibot, S.
    [J]. REANIMATION, 2016, 25 (03): : 328 - 339
  • [16] Performance of Current Intensive Care Unit Ventilators During Pressure and Volume Ventilation
    Marchese, Andrew D.
    Sulemanji, Demet
    Chipman, Daniel
    Villar, Jesus
    Kacmarek, Robert M.
    [J]. RESPIRATORY CARE, 2011, 56 (07) : 928 - 940
  • [17] The Urge to Build More Intensive Care Unit Beds and Ventilators: Intuitive but Errant
    Halpern, Scott D.
    Miller, Franklin G.
    [J]. ANNALS OF INTERNAL MEDICINE, 2020, 173 (04) : 302 - +
  • [18] Accuracy of tidal volume delivery by paediatric intensive care ventilators: A bench-model study
    Vedrenne-Cloquet, Meryl
    Tuffet, Samuel
    Louis, Bruno
    Khirani, Sonia
    Collignon, Charlotte
    Renolleau, Sylvain
    Fauroux, Brigitte
    Carteaux, Guillaume
    [J]. AUSTRALIAN CRITICAL CARE, 2024, 37 (04) : 600 - 605
  • [19] Bone Conduction Noise Exposure via Ventilators in the Neonatal Intensive Care Unit
    Gol, Mohammad Abraham Kazemizadeh
    Black, Angela
    Sidman, James
    [J]. LARYNGOSCOPE, 2015, 125 (10): : 2388 - 2392
  • [20] Public Preferences for Allocating Ventilators in an Intensive Care Unit: A Discrete Choice Experiment
    Norman, Richard
    Robinson, Suzanne
    Dickinson, Helen
    Williams, Iestyn
    Meshcheriakova, Elena
    Manipis, Kathleen
    Anstey, Matthew
    [J]. PATIENT-PATIENT CENTERED OUTCOMES RESEARCH, 2021, 14 (03): : 319 - 330