Accuracy of tidal volume delivery by paediatric intensive care ventilators: A bench-model study

被引:0
|
作者
Vedrenne-Cloquet, Meryl [1 ,2 ,3 ]
Tuffet, Samuel [4 ,5 ,6 ]
Louis, Bruno [4 ]
Khirani, Sonia [1 ,2 ,7 ]
Collignon, Charlotte [3 ]
Renolleau, Sylvain [2 ,3 ]
Fauroux, Brigitte [1 ,2 ]
Carteaux, Guillaume [4 ,5 ,6 ]
机构
[1] Hop Necker Enfants Malad, AP HP, Pediat Sleep & Noninvas Ventilat Unit, EA 7330,VIFASOM, Paris, France
[2] Univ Paris, Paris, France
[3] CHU Necker Enfants Malad, AP HP, Reanimat Medicochirurg Pediat, Paris, France
[4] Univ Paris Est Creteil IMRB, INSERM, CNRS EMR, F-7000 Creteil, France
[5] Univ Paris Est Creteil, Fac Sante Creteil, GRC CARMAS, IMRB, Creteil, France
[6] Ctr Hosp Univ Henri Mondor, AP HP, Reanimat Med, Creteil, France
[7] ASV Sante, Gennevilliers, France
关键词
Volume-controlled ventilation; Bench study; Heatedhumidifier; Heat moisture exchanger; Paediatric intensive care unit; RESPIRATORY-DISTRESS-SYNDROME; ACUTE LUNG INJURY; ICU VENTILATORS; INFANTS; CHILDREN; PERFORMANCE; MECHANICS;
D O I
10.1016/j.aucc.2023.12.007
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Background: Tidal volume (Vt) delivery during mechanical ventilation is influenced by gas compression, humidity, and temperature. Objectives: This bench study aimed at assessing the accuracy of Vt delivery by paediatric intensive care ventilators according to the humidification system. Secondary objectives were to assess the following: (i) the accuracy of Vt delivery in ventilators with an integrated Y-piece pneumotachograph and (ii) the ability of ventilators to deliver and maintain a preset positive end-expiratory pressure. Methods: Six latest-generation intensive care ventilators equipped with a paediatric mode were tested on the ASL5000 test lung in four simulated paediatric bench models (full-term neonate, infant, preschool-age chile, and school-age child), under volume-controlled mode with a heated humidifier (HH) or a heat moisture exchanger, with various loading conditions. Three ventilators equipped with a Y-piece pneumotachograph were tested with or without the pneumotachograph in the neonatal and infant models. "Accurate Vt" delivery was defined as a volume error (percentage of the preset Vt under body temperature and pressure and saturated water vapour conditions) being <= 10 % of the absolute preset value. Results: Vt accuracy varied significantly across ventilators but was acceptable in almost all the ventilators and all the models, except the neonatal model. The humidification system had an impact on Vt delivery in the majority of the tested conditions (p < 0.05). The use of an HH was associated with a better Vt accuracy in four ventilators (V500, V800, R860, and ServoU) and allowed to achieve an acceptable level of volume error in the neonatal model as compared to the use of heat moisture exchanger. The use of an integrated pneumotachograph was associated with lower volume error in only one ventilator (p < 0.01). All the tested ventilators were able to maintain adequate positive end-expiratory pressure levels. Conclusion: The humidification system affects Vt accuracy of paediatric intensive care ventilators, especially in the youngest patients for whom the HH should be preferred. (c) 2024 Australian College of Critical Care Nurses Ltd. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:600 / 605
页数:6
相关论文
共 50 条
  • [1] Accuracy of ventilators for intermediate care to deliver tidal volume. A bench study
    Baboi, Loredana
    Guegan, Sarah
    Guerin, Claude
    [J]. EUROPEAN RESPIRATORY JOURNAL, 2015, 46
  • [2] Tidal Volume Delivery From ICU Ventilators at BTPS Conditions: A Bench Study
    Duchateau, Paul
    Guerin, Claude
    [J]. RESPIRATORY CARE, 2013, 58 (04) : 623 - 632
  • [3] A bench evaluation of fraction of oxygen in air delivery and tidal volume accuracy in home care ventilators available for hospital use
    Baboi, Loredana
    Subtil, Fabien
    Guerin, Claude
    [J]. JOURNAL OF THORACIC DISEASE, 2016, 8 (12) : 3639 - 3647
  • [4] TIDAL VOLUME DELIVERY FROM ICE VENTILATORS IN BTPS CONDITION. A BENCH MODEL
    Duchateau, P.
    Guerin, C.
    [J]. INTENSIVE CARE MEDICINE, 2012, 38 : S67 - S67
  • [5] Evaluation of transport ventilators used by a paediatric intensive care transport services: a bench study
    Manning, G.
    Ray, S.
    Chigaru, L.
    [J]. ANAESTHESIA, 2018, 73 : 31 - 31
  • [6] Delivery of tidal volume from four anaesthesia ventilators during volume-controlled ventilation: a bench study
    Wallon, G.
    Bonnet, A.
    Guerin, C.
    [J]. BRITISH JOURNAL OF ANAESTHESIA, 2013, 110 (06) : 1045 - 1051
  • [7] DELIVERY OF TIDAL VOLUME FROM FOUR ANAESTHESIA VENTILATORS DURING VOLUME-CONTROLLED VENTILATION: A BENCH STUDY
    Wallon, G.
    Bonnet, A.
    Guerin, C.
    [J]. INTENSIVE CARE MEDICINE, 2012, 38 : S266 - S267
  • [8] Effect of Leak and Breathing Pattern on the Accuracy of Tidal Volume Estimation by Commercial Home Ventilators: A Bench Study
    Lujan, Manel
    Sogo, Ana
    Pomares, Xavier
    Monso, Eduard
    Sales, Bemat
    Blanch, Lluis
    [J]. RESPIRATORY CARE, 2013, 58 (05) : 770 - 777
  • [9] Accuracy of volume and pressure delivery by mechanical ventilators in use in neonatal intensive care units: A quality control study
    Ottaviani, Valeria
    Veneroni, Chiara
    Dell'Orto, Valentina
    Lavizzari, Anna
    Ventura, Maria Luisa
    Tossici, Marta
    Maroszynska, Iwona
    Mosca, Fabio
    Tagliabue, Paolo
    De Luca, Daniele
    Dellaca, Raffaele L.
    [J]. PEDIATRIC PULMONOLOGY, 2020, 55 (08) : 1955 - 1962
  • [10] Accuracy of Delivery and Effects on Absolute Humidity of Low Tidal Volume by ICU Ventilators
    Moro, Barbara
    Baboi, Loredana
    Yonis, Hodane
    Subtil, Fabien
    Louis, Bruno
    Guerin, Claude
    [J]. RESPIRATORY CARE, 2018, 63 (10) : 1253 - 1263