Tidal Volume Delivery From ICU Ventilators at BTPS Conditions: A Bench Study

被引:8
|
作者
Duchateau, Paul [1 ]
Guerin, Claude [2 ,3 ]
机构
[1] Haute Etud Ingn, Lille, France
[2] Hosp Civils Lyon, Hop Croix Rousse, Serv Reanimat Med, Lyon, France
[3] Univ Lyon, Lyon, France
关键词
tidal volume; ventilator; nebulization; external filter; MECHANICAL VENTILATION; HUMIDIFICATION; HEAT;
D O I
10.4187/respcare.01915
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
BACKGROUND: Even though it is not a common practice, an external filter to the expiratory limb of the breathing circuit may protect the expiratory valve from water saturation in case of nebulization, or from the environment in case of lung infection with multi-drug-resistant micro-organisms or H1N1 influenza. METHODS: We added an external filter to the expiratory limb and measured tidal volume (V-T) from 6 ICU ventilators: 2 with built-in expiratory filter (Avea, Puritan Bennett 840), and 4 without (Engstrom Carestation, Evita XL, Evita V500, and Servo-i), set in volume controlled mode, at BTPS (body temperature and pressure saturated) condition, with a heated humidifier and a lung model (compliance 16 mL/cm H2O, resistance 20 cm H2O/L/s) placed inside a neonatal incubator. The temperature was targeted at 37 degrees C for both the heated humidifier and the incubator. The setup was run continuously for 24 hours. In the latter 4 ICU ventilators, a Hygrobac or Sterivent S external filter was placed upstream from the expiratory valve for an additional 24-hour period for each. At the end of this period, V-T was measured at 4 nominal V-T values (300, 400, 500, and 800 mL) with a pneumotachograph. The volume error computed from the ratio of set to measured V-T (% set V-T) was the primary end point. RESULTS: In these warm and wet conditions, volume error averaged 96 +/- 3% for Avea, 100 +/- 7% for Puritan Bennett 840, 90 +/- 2% for Evita XL, 100 +/- 7% for Evita V500, 105 +/- 2% for Servo-i, and 108 +/- 4% for Engstrom Carestation (P < .001). With the Hygrobac the values were 93 +/- 1% for Evita XL, 94 +/- 4% for Evita V500, 110 +/- 4% for Servo-i, and 99 at 2% for Engstrom Carestation (P < .001). With the Sterivent the corresponding values were 95 +/- 2%, 105 +/- 2%, 112 +/- 5%, and 98 +/- 2%, respectively (P < .001). CONCLUSIONS: In BTPS condition, volume error differed substantially across ICU ventilators for V-T delivery, with further significant changes occurring after addition of a filter at the distal expiratory limb.
引用
收藏
页码:623 / 632
页数:10
相关论文
共 50 条
  • [1] TIDAL VOLUME DELIVERY FROM ICE VENTILATORS IN BTPS CONDITION. A BENCH MODEL
    Duchateau, P.
    Guerin, C.
    INTENSIVE CARE MEDICINE, 2012, 38 : S67 - S67
  • [2] Delivery of tidal volume from four anaesthesia ventilators during volume-controlled ventilation: a bench study
    Wallon, G.
    Bonnet, A.
    Guerin, C.
    BRITISH JOURNAL OF ANAESTHESIA, 2013, 110 (06) : 1045 - 1051
  • [3] DELIVERY OF TIDAL VOLUME FROM FOUR ANAESTHESIA VENTILATORS DURING VOLUME-CONTROLLED VENTILATION: A BENCH STUDY
    Wallon, G.
    Bonnet, A.
    Guerin, C.
    INTENSIVE CARE MEDICINE, 2012, 38 : S266 - S267
  • [4] Accuracy of tidal volume delivery by paediatric intensive care ventilators: A bench-model study
    Vedrenne-Cloquet, Meryl
    Tuffet, Samuel
    Louis, Bruno
    Khirani, Sonia
    Collignon, Charlotte
    Renolleau, Sylvain
    Fauroux, Brigitte
    Carteaux, Guillaume
    AUSTRALIAN CRITICAL CARE, 2024, 37 (04) : 600 - 605
  • [5] Correction to the paper 'Delivery of tidal volume from four anaesthesia ventilators during volume-controlled ventilation: a bench study'
    Hallback, M.
    Wallin, M. K. E. B.
    BRITISH JOURNAL OF ANAESTHESIA, 2016, 116 (05) : 720 - 721
  • [6] Accuracy of Delivery and Effects on Absolute Humidity of Low Tidal Volume by ICU Ventilators
    Moro, Barbara
    Baboi, Loredana
    Yonis, Hodane
    Subtil, Fabien
    Louis, Bruno
    Guerin, Claude
    RESPIRATORY CARE, 2018, 63 (10) : 1253 - 1263
  • [7] Accuracy of ventilators for intermediate care to deliver tidal volume. A bench study
    Baboi, Loredana
    Guegan, Sarah
    Guerin, Claude
    EUROPEAN RESPIRATORY JOURNAL, 2015, 46
  • [8] A bench evaluation of fraction of oxygen in air delivery and tidal volume accuracy in home care ventilators available for hospital use
    Baboi, Loredana
    Subtil, Fabien
    Guerin, Claude
    JOURNAL OF THORACIC DISEASE, 2016, 8 (12) : 3639 - 3647
  • [9] Effect of Leak and Breathing Pattern on the Accuracy of Tidal Volume Estimation by Commercial Home Ventilators: A Bench Study
    Lujan, Manel
    Sogo, Ana
    Pomares, Xavier
    Monso, Eduard
    Sales, Bemat
    Blanch, Lluis
    RESPIRATORY CARE, 2013, 58 (05) : 770 - 777
  • [10] Delivering Low Tidal Volume With Anesthesia and ICU Ventilators in a Neonatal Lung Model
    Medeiros, Kyle J.
    Morais, Caio A.
    Winterton, Dario
    Rub, David M.
    Santiago, Roberta De Santis
    Shekhar, Nitika
    Chipman, Daniel
    Monaghan, Thomas T.
    Bittner, Edward A.
    Carroll, Ryan
    Berra, Lorenzo
    RESPIRATORY CARE, 2023, 68 (03) : 384 - 391