A MONTE CARLO CALIBRATION OF A WHOLE BODY COUNTER USING THE ICRP COMPUTATIONAL PHANTOMS

被引:6
|
作者
Nilsson, Jenny [1 ]
Isaksson, Mats [1 ]
机构
[1] Univ Gothenburg, Dept Radiat Phys, Inst Clin Sci, Sahlgrenska Acad, Gothenburg, Sweden
关键词
ABSORPTION; SIMULATION; RETENTION;
D O I
10.1093/rpd/ncu256
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
A fast and versatile calibration of a whole body counter (WBC) is presented. The WBC, consisting of four large plastic scintillators, is to be used for measurements after accident or other incident involving ionising radiation. The WBC was calibrated using Monte Carlo modelling and the ICRP computational phantoms. The Monte Carlo model of the WBC was made in GATE, v6.2 (Geant4 Application for Tomographic Emission) and MATLAB. The Monte Carlo model was verified by comparing simulated energy spectrum and simulated counting efficiency with experimental energy spectrum and experimental counting efficiency for high-energy monoenergetic gamma-emitting point sources. The simulated results were in good agreement with experimental results except when compared with experimental results from high dead-time (DT) measurements. The Monte Carlo calibration was made for a heterogeneous source distribution of Cs-137 and K-40, respectively, inside the ICRP computational phantoms. The source distribution was based on the biokinetic model for Cs-137.
引用
收藏
页码:458 / 467
页数:10
相关论文
共 50 条
  • [1] Application of voxel phantoms and Monte Carlo method to whole-body counter calibration
    Kinase, S.
    Takagi, S.
    Noguchi, H.
    Saito, K.
    [J]. RADIATION PROTECTION DOSIMETRY, 2007, 125 (1-4) : 189 - 193
  • [2] Monte Carlo calculation of whole body counter efficiency factors for different computational phantoms
    Ferreira, Carlos V. G.
    Piedade, Jennifer S.
    Prado, Max R. D.
    Benavente, Jhonny
    Mendes, Bruno M.
    Paixao, Lucas
    Fonseca, Telma C. F.
    [J]. APPLIED RADIATION AND ISOTOPES, 2023, 194
  • [3] MONTE CARLO SIMULATION OF A WHOLE-BODY COUNTER USING IGOR PHANTOMS
    Bochud, Francois O.
    Laedermann, Jean-Pascal
    Baechler, Sebastien
    Bailat, Claude J.
    Boschung, Markus
    Aroua, Abbas
    Mayer, Sabine
    [J]. RADIATION PROTECTION DOSIMETRY, 2014, 162 (03) : 280 - 288
  • [4] COMPARISON OF TWO ANTHROPOMORPHIC PHANTOMS AS A CALIBRATION TOOL FOR WHOLE-BODY COUNTER USING MONTE CARLO SIMULATIONS
    Manohari, M.
    Mathiyarasu, R.
    Rajagopal, V.
    Venkatraman, B.
    [J]. RADIATION PROTECTION DOSIMETRY, 2015, 164 (03) : 298 - 303
  • [5] EVALUATION OF COUNTING EFFICIENCIES OF A WHOLE-BODY COUNTER USING MONTE CARLO SIMULATION WITH VOXEL PHANTOMS
    Takahashi, Masa
    Kinase, Sakae
    Kramer, Richard
    [J]. RADIATION PROTECTION DOSIMETRY, 2011, 144 (1-4) : 407 - 410
  • [6] Monte Carlo calculations for efficiency calibration of a whole-body monitor using BOMAB phantoms of different sizes
    Bhati, S.
    Patni, H. K.
    Ghare, V. P.
    Singh, I. S.
    Nadar, M. Y.
    [J]. RADIATION PROTECTION DOSIMETRY, 2012, 148 (04) : 414 - 419
  • [7] STANDARDIZATION OF CALIBRATION METHOD OF WHOLE-BODY COUNTER .1. CALIBRATION BY USING ANTHROPOMETRIC PHANTOMS
    ISHIKAWA, T
    MATSUMOTO, M
    UCHIYAMA, M
    MIZUSHITA, S
    KOBAYASHI, S
    [J]. JOURNAL OF THE ATOMIC ENERGY SOCIETY OF JAPAN, 1995, 37 (05): : 420 - 429
  • [8] Monte-Carlo simulations with mathematical phantoms to investigate the effectiveness of a whole-body counter for thyroid measurement
    Tani, Kotaro
    Igarashi, Yu
    Kim, Eunjoo
    Iimoto, Takeshi
    Kurihara, Osamu
    [J]. RADIATION MEASUREMENTS, 2020, 135
  • [9] A calibration method for whole-body counters, using Monte Carlo simulation
    Ishikawa, T
    Matsumoto, M
    Uchiyama, M
    [J]. RADIATION PROTECTION DOSIMETRY, 1996, 64 (04) : 283 - 288
  • [10] Monte Carlo simulation of a scanning detector whole body counter and the effect of BOMAB phantom size on the calibration
    Kramer, GH
    Burns, LC
    Guerriere, S
    [J]. HEALTH PHYSICS, 2002, 83 (04): : 526 - 533