A MONTE CARLO CALIBRATION OF A WHOLE BODY COUNTER USING THE ICRP COMPUTATIONAL PHANTOMS

被引:6
|
作者
Nilsson, Jenny [1 ]
Isaksson, Mats [1 ]
机构
[1] Univ Gothenburg, Dept Radiat Phys, Inst Clin Sci, Sahlgrenska Acad, Gothenburg, Sweden
关键词
ABSORPTION; SIMULATION; RETENTION;
D O I
10.1093/rpd/ncu256
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
A fast and versatile calibration of a whole body counter (WBC) is presented. The WBC, consisting of four large plastic scintillators, is to be used for measurements after accident or other incident involving ionising radiation. The WBC was calibrated using Monte Carlo modelling and the ICRP computational phantoms. The Monte Carlo model of the WBC was made in GATE, v6.2 (Geant4 Application for Tomographic Emission) and MATLAB. The Monte Carlo model was verified by comparing simulated energy spectrum and simulated counting efficiency with experimental energy spectrum and experimental counting efficiency for high-energy monoenergetic gamma-emitting point sources. The simulated results were in good agreement with experimental results except when compared with experimental results from high dead-time (DT) measurements. The Monte Carlo calibration was made for a heterogeneous source distribution of Cs-137 and K-40, respectively, inside the ICRP computational phantoms. The source distribution was based on the biokinetic model for Cs-137.
引用
收藏
页码:458 / 467
页数:10
相关论文
共 50 条
  • [31] SELECTED ORGAN DOSE CONVERSION COEFFICIENTS FOR EXTERNAL PHOTONS CALCULATED USING ICRP ADULT VOXEL PHANTOMS AND MONTE CARLO CODE FLUKA
    Patni, H. K.
    Nadar, M. Y.
    Akar, D. K.
    Bhati, S.
    Sarkar, P. K.
    [J]. RADIATION PROTECTION DOSIMETRY, 2011, 147 (03) : 406 - 416
  • [32] Application of voxel phantoms in whole-body counting for the validation of calibration phantoms and the assessment of uncertainties
    de Carlan, L.
    Roch, P.
    Blanchardon, E.
    Franck, D.
    [J]. RADIATION PROTECTION DOSIMETRY, 2007, 125 (1-4) : 477 - 482
  • [33] Calibration of a phoswich type partial body counter by Monte Carlo simulation of low-energy photon transport
    Doerfel, H.
    Heide, B.
    [J]. RADIATION PROTECTION DOSIMETRY, 2007, 123 (04) : 464 - 472
  • [34] Monte Carlo simulation of the dose distribution of ICRP adult reference computational phantoms for acquisitions with a 320 detector-row cone-beam CT scanner
    Salvado, M.
    Cros, M.
    Joemai, R. M. S.
    Calzado, A.
    Geleijns, J.
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2015, 31 (05): : 452 - 462
  • [35] Monte Carlo comparison of the St Petersburg phantom with a BOMAB phantom in the HML's whole-body counter
    Kramer, Gary H.
    Capello, Kevin
    Sung, Jeremy
    [J]. RADIATION PROTECTION DOSIMETRY, 2008, 128 (02) : 245 - 250
  • [36] NCINM: an internal dosimetry software using ICRP and NCI pediatric and adult computational phantoms
    Villoing, D.
    Lee, C.
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2019, 46 (SUPPL 1) : S364 - S364
  • [37] DEVELOPING COMPUTATIONAL METHODS TO VIEW DISTRIBUTIONS OF SECOND CANCER RISK USING MONTE CARLO AND VOXEL HUMAN PHANTOMS
    Alsaleh, W.
    Hugtenburg, R.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2012, 102 : S75 - S75
  • [38] A DXA Whole Body Composition Cross-Calibration Experience: Evaluation With Humans, Spine, and Whole Body Phantoms
    Krueger, Diane
    Libber, Jessie
    Sanfilippo, Jennifer
    Yu, Hui Jing
    Horvath, Blaine
    Miller, Colin G.
    Binkley, Neil
    [J]. JOURNAL OF CLINICAL DENSITOMETRY, 2016, 19 (02) : 220 - 225
  • [39] Monte Carlo simulations using whole-body pediatric and adult phantoms as virtual patients to assess secondary organ doses in proton radiation therapy
    Zacharatou-Jarlskog, C.
    Lee, C.
    Jiang, H.
    Bolch, W.
    Xu, X.
    Paganetti, H.
    [J]. MEDICAL PHYSICS, 2006, 33 (06) : 2123 - 2123
  • [40] CALIBRATION OF WHOLE-BODY COUNTER DETECTORS BY DATA TRANSFER
    FULOP, M
    LAHHAM, A
    RAGAN, P
    [J]. RADIATION PROTECTION DOSIMETRY, 1995, 59 (01) : 59 - 62