Quantification of Cold Spots Caused by Geometrical Uncertainty in Field-in-field Techniques for Whole Breast Radiotherapy

被引:8
|
作者
Nakamura, Naoki [1 ]
Hatanaka, Shogo
Shikama, Naoto [2 ]
Akahane, Keiko
Sekiguchi, Kenji
机构
[1] St Lukes Int Hosp, Dept Radiat Oncol, Chuoh Ku, Tokyo 1048560, Japan
[2] Saitama Med Univ Hosp, Dept Radiat Oncol, Hidaka, Japan
关键词
breast cancer; radiotherapy; whole breast radiotherapy; field-in-field technique; geometrical uncertainty; INTENSITY MODULATION; DOSE DISTRIBUTIONS; IRRADIATION; IMRT; MOTION;
D O I
10.1093/jjco/hyr112
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Objective: To quantify the cold spot under geometrical uncertainties in field-in-field techniques for whole breast radiotherapy. Methods: Ten consecutive patients from both the left-and right-sided treatment site groups who received whole breast radiotherapy with the field-in-field technique were included. Virtual plans were made with moving isocenters to the posterior direction having two amplitudes (5 and 10 mm) and prescribing the same monitor unit as the original plan (FIF_5 and FIF_10). The planning target volume for evaluation was defined by subtracting the areas within 5 mm from the skin and within 5 mm from the lung from the whole breast. The differences in V90, V95 and D98 of planning target volume for evaluation were measured between the original and virtual plans. As a reference, the same measurements were taken for the wedge techniques (Wedge_5 and Wedge_10). Results: The differences in V95 were -0.2% on FIF_5, -1.7% on FIF_10, -0.5% on Wedge_5 and -1.5% on Wedge_10. The differences in V90 were -0.02% on FIF_5, -0.3% on FIF_10, 0.05% on Wedge_5 and 0.1% on Wedge_10. The differences in D98 were 0 Gy on FIF_5, -0.1 Gy on FIF_10, -0.2 Gy on Wedge_5 and -0.4 Gy on Wedge_10. The differences in D98 between the original plans and virtual scenarios for field-in-field techniques were significantly smaller than those for wedge techniques, but there were no statically significant differences in V90 and V95. Conclusions: The quantity of the cold spots caused by the geometrical uncertainties in field-in-field techniques was similar to that for the wedge techniques and was acceptable.
引用
收藏
页码:1127 / 1131
页数:5
相关论文
共 50 条
  • [41] Comparison of Field in Field and Tangential Wedged Beams techniques in radiotherapy of breast cancer
    Yarahmadi, M.
    Haghparast, A.
    Faramarzi, B.
    Saalehi, Z.
    RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S1074 - S1074
  • [42] Dosimetric comparison of 3-dimensional conformal and field-in-field radiotherapy techniques for the adjuvant treatment of early stage endometrial cancer
    Yavas, Guler
    Yavas, Cagdas
    Acar, Hilal
    Buyukyoruk, Ahmet
    Cobanoglu, Gokcen
    Kerimoglu, Ozlem Secilmis
    Yavas, Ozlem
    Celik, Cetin
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2013, 29 (06): : 577 - 582
  • [43] DOSE DISTRIBUTION ANALYSIS OF AXILLARY LYMPH NODES FOR THREE-DIMENSIONAL CONFORMAL RADIOTHERAPY WITH A FIELD-IN-FIELD TECHNIQUE FOR BREAST CANCER
    Ohashi, Toshio
    Takeda, Atsuya
    Shigematsu, Naoyuki
    Fukada, Junichi
    Sanuki, Naoko
    Amemiya, Atsushi
    Kubo, Atsushi
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 73 (01): : 80 - 87
  • [44] DOSIMETRIC COMPARISON OF FIELD IN FIELD INTENSITY MODULATED RADIOTHERAPY TECHNIQUE WITH CONFORMAL RADIOTHERAPY TECHNIQUES IN BREAST CANCER
    Ercan, T.
    Igdem, S.
    Alco, G.
    Zengin, F.
    Ozgules, R.
    Atilla, S.
    Dincer, M.
    Okkan, S.
    RADIOTHERAPY AND ONCOLOGY, 2009, 92 : S201 - S202
  • [45] Dose distribution analysis of axillary lymph nodes for three-dimensional conformal radiotherapy with a field-in-field technique for breast cancer
    Ohashi, T.
    Takeda, A.
    Shigematsu, N.
    Fukuda, J.
    Sanuki, N.
    Ameniya, A.
    Kubo, A.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 72 (01): : S177 - S177
  • [46] Comparison between the four-field box and field-in-field techniques for conformal radiotherapy of the esophagus using dose–volume histograms and normal tissue complication probabilities
    Farzaneh Allaveisi
    Amir Nami Moghadam
    Japanese Journal of Radiology, 2017, 35 : 327 - 334
  • [47] Assessment of planning reproducibility in three-dimensional field-in-field radiotherapy technique for breast cancer: impact of surgery-simulation interval
    Dong Soo Lee
    Young Kyu Lee
    Young Nam Kang
    Yong Gyun Won
    Seung Hwan Park
    Yong Seok Kim
    Jeong Soo Kim
    Hye Sung Won
    Scientific Reports, 11
  • [48] Dosimetric comparison of field in field intensity-modulated radiotherapy technique with conformal radiotherapy techniques in breast cancer
    Ercan, Tuelay
    Igdem, Sefik
    Alco, Guel
    Zengin, Funda
    Atilla, Selin
    Dincer, Maktav
    Okkan, Sait
    JAPANESE JOURNAL OF RADIOLOGY, 2010, 28 (04) : 283 - 289
  • [49] Dosimetric comparison of field in field intensity-modulated radiotherapy technique with conformal radiotherapy techniques in breast cancer
    Tülay Ercan
    Şefik İğdem
    Gül Alço
    Funda Zengin
    Selin Atilla
    Maktav Dinçer
    Sait Okkan
    Japanese Journal of Radiology, 2010, 28 : 283 - 289
  • [50] Review of the three-field techniques in breast cancer radiotherapy
    Kagkiouzis, John
    Platoni, Kalliopi
    Kantzou, Ioanna
    Dilvoi, Maria
    Patatoukas, George
    Kypraiou, Efrosyni
    Efstathopoulos, Efstathios
    Sarris, George
    Trogkanis, Nikolaos
    Kouloulias, Vassilis
    JOURNAL OF BUON, 2017, 22 (03): : 599 - 605