摘要
The aim of this study was to be able to implement the field-in-field intensity-modulated radiotherapy (FiF) technique in our daily practice for breast radiotherapy. To do this, we performed a dosimetric comparison. Treatment plans were produced for 20 consecutive patients. FiF plans and conformal radiotherapy (CRT) plans were compared for doses in the planning target volume (PTV), the dose homogeneity index (DHI), doses in irradiated soft tissue outside the target volume (SST), ipsilateral lung and heart doses for left breast irradiation, and the monitor unit counts (MU) required for treatment. Averaged values were compared using Student's t-test. With FiF, the DHI is improved 7.0% and 5.7%, respectively (P < 0.0001) over the bilateral and lateral wedge CRT techniques. When the targeted volumes received 105% and 110% of the prescribed dose in the PTV were compared, significant decreases are found with the FiF technique. With the 105% dose, the SST, heart, and ipsilateral lung doses and the MU counts were also significantly lower with the FiF technique. The FiF technique, compared to CRT, for breast radiotherapy enables significantly better dose distribution in the PTV. Significant differences are also found for soft tissue volume, the ipsilateral lung dose, and the heart dose. Considering the decreased MUs needed for treatment, the FiF technique is preferred over tangential CRT.
引用
收藏
页码:283 / 289
页数:7
相关论文
共 50 条
  • [1] Dosimetric comparison of field in field intensity-modulated radiotherapy technique with conformal radiotherapy techniques in breast cancer
    Tülay Ercan
    Şefik İğdem
    Gül Alço
    Funda Zengin
    Selin Atilla
    Maktav Dinçer
    Sait Okkan
    [J]. Japanese Journal of Radiology, 2010, 28 : 283 - 289
  • [2] DOSIMETRIC COMPARISON OF FIELD IN FIELD INTENSITY MODULATED RADIOTHERAPY TECHNIQUE WITH CONFORMAL RADIOTHERAPY TECHNIQUES IN BREAST CANCER
    Ercan, T.
    Igdem, S.
    Alco, G.
    Zengin, F.
    Ozgules, R.
    Atilla, S.
    Dincer, M.
    Okkan, S.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2009, 92 : S201 - S202
  • [3] Determination and comparison of dosimetric parameters of three-dimensional conformal radiotherapy, field in field, and intensity-modulated radiotherapy techniques in radiotherapy of breast conserving patients
    Takabi, Fatemeh Shirani
    Broomand, Mohammad Ali
    Nickfarjam, Abolfazl
    Asadi, Amin
    Namiranian, Nasim
    [J]. JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2023, 19 (03) : 624 - 632
  • [4] Adjuvant radiotherapy for gallbladder cancer: A dosimetric comparison of conformal radiotherapy and intensity-modulated radiotherapy
    Sun, Xiao-Nan
    Wang, Qi
    Gu, Ben-Xing
    Zhu, Yan-Hong
    Hu, Jian-Bin
    Shi, Guo-Zhi
    Zheng, Shu
    [J]. WORLD JOURNAL OF GASTROENTEROLOGY, 2011, 17 (03) : 397 - 402
  • [6] Dosimetric Comparison of Intensity-Modulated Radiotherapy, Volumetric Modulated Arc Therapy and Hybrid Three-Dimensional Conformal Radiotherapy/Intensity-Modulated Radiotherapy Techniques for Right Breast Cancer
    Liu, Yi-Chi
    Chang, Hung-Ming
    Lin, Hsin-Hon
    Lu, Chia-Chun
    Lai, Lu-Han
    [J]. JOURNAL OF CLINICAL MEDICINE, 2020, 9 (12) : 1 - 14
  • [7] Dosimetric Comparison of 3-Dimensional Conformal and Intensity-Modulated Radiotherapy Planning Techniques in Breast Cancer
    Sirin Ozdemir, Beyza
    Kirecci, Orkun
    [J]. UHOD-ULUSLARARASI HEMATOLOJI-ONKOLOJI DERGISI, 2019, 29 (04): : 191 - 197
  • [8] Breast-conserving radiotherapy with simultaneous integrated boost; field-in-field three-dimensional conformal radiotherapy versus inverse intensity-modulated radiotherapy - A dosimetric comparison: Do we need intensity-modulated radiotherapy?
    Joseph, Bindhu
    Farooq, Nisma
    Kumar, Sabari
    Vijay, C. R.
    Puthur, Kurian J.
    Ramesh, C.
    Lokesh, Vishwanath
    [J]. SOUTH ASIAN JOURNAL OF CANCER, 2018, 7 (03) : 163 - 166
  • [9] Dosimetric comparison of whole breast radiotherapy using field in field and conformal radiotherapy techniques in early stage breast cancer
    Yavas, G.
    Yavas, C.
    Acar, H.
    [J]. IRANIAN JOURNAL OF RADIATION RESEARCH, 2012, 10 (3-4): : 131 - 138
  • [10] A Dosimetric Comparison of Inverse Intensity-modulated Radiotherapy and Forward Intensity-modulated Radiotherapy for Breast Cancer After Radical Mastectomy
    Zhang, G.
    Yin, Y.
    Lu, J.
    [J]. EUROPEAN JOURNAL OF CANCER, 2011, 47 : S341 - S342