Performance evaluation of an LED flatbed scanner for triple channel film dosimetry with EBT3 and EBT-XD film

被引:4
|
作者
Van der Walt, Marchant [1 ]
Marsh, Linda [1 ]
Baines, John [1 ]
Gibson, Stephen [1 ]
Shoobridge, Ariadne [1 ]
de Vine, Glenn [1 ]
机构
[1] Townsville Hosp & Hlth Serv, Dept Med Phys, Townsville, Qld 4814, Australia
关键词
Radiochromic film dosimetry; Radiotherapy; Triple channel corrections; LED; RADIOCHROMIC FILM; DOCUMENT SCANNER; CALIBRATION;
D O I
10.1007/s13246-022-01161-z
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We investigate the properties of a light emitting diode (LED) flatbed scanner for use with EBT3 and EBT-XD film types in a clinical radiochromic film (RCF) dosimetry program with modern treatment techniques. The flatbed scanner was characterised in terms of lateral and longitudinal response, X-Y scaling integrity, scanning reproducibility, scanner warm up dependence and film orientation dependence. The preferred lateral response artefact (LRA) corrections are investigated for the LED light source. Supporting evidence is provided regarding the dose independent nature of the corrections while also providing results suggesting a potential film type independence. Results from 2D gamma analysis of four patient treatments were compared between the new 12000XL and existing 10000XL model. Lastly, a dose uncertainty analysis was performed for the film-scanner system combination. It may be concluded that the lateral response variation requires correction while the longitudinal response variation is insignificant. The linear scaling in the lateral and longitudinal directions are within 0.5% and the scanner reproducibility is stable. Scanner warm up dependence no longer exists, and effort should be made to maintain all film orientation in a study set within 15 degrees. The LRA corrections are as reported substantially dose independent and there is evidence to support film type independence. Comparative gamma analysis of patient specific dose maps between the EPSON 10000XL (xenon fluorescent lamp) and 12000XL (LED) scanners showed that results are indistinguishable for both film types across the two scanner models when the necessary corrections are applied. Dose uncertainty is in agreement with the literature and can be kept below 3% with necessary corrections applied.
引用
收藏
页码:901 / 914
页数:14
相关论文
共 50 条
  • [41] Technical Note: On GAFChromic EBT-XD film and the lateral response artifact
    Lewis, David F.
    Chan, Maria F.
    MEDICAL PHYSICS, 2016, 43 (02) : 643 - 649
  • [42] Technical Note: Initial characterization of the new EBT-XD Gafchromic film
    Grams, Michael P.
    Gustafson, Jon M.
    Long, Kenneth M.
    de los Santos, Luis E. Fong
    MEDICAL PHYSICS, 2015, 42 (10) : 5782 - 5786
  • [43] Evaluation of EBT3 Film Response in Different Batches
    Escarcia, F.
    Herrera, J.
    Garcia, O.
    MEDICAL PHYSICS, 2015, 42 (06) : 3422 - 3422
  • [44] A protocol for EBT3 radiochromic film dosimetry using reflection scanning
    Papaconstadopoulos, Pavlos
    Hegyi, Gyorgy
    Seuntjens, Jan
    Devic, Slobodan
    MEDICAL PHYSICS, 2014, 41 (12)
  • [45] Radiochromic film dosimetry: Considerations on precision and accuracy for EBT2 and EBT3 type films
    Dreindl, Ralf
    Georg, Dietmar
    Stock, Markus
    ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2014, 24 (02): : 153 - 163
  • [46] Evaluation of the uncertainty in an EBT3 film dosimetry system utilizing net optical density
    Leon Marroquin, Elsa Y.
    Herrera Gonzalez, Jose A.
    Camacho Lopez, Miguel A.
    Villarreal Barajas, Jose E.
    Garcia-Garduno, Olivia A.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2016, 17 (05): : 466 - 481
  • [47] Dosimetry evaluation of 192Ir Flexisource using gafchromic EBT3 film
    Oustous, Aziz
    Airouss, Zouhair
    EL Attaoui, Yacine
    Oustous, Yassine
    Khalis, Mohammed
    Sebihi, Rajaa
    JOURNAL OF RADIOTHERAPY IN PRACTICE, 2024, 23
  • [48] Artifact free EBT film scanner for radiotherapy dosimetry
    Ranade, M.
    DuBose, R.
    Li, J.
    Kozelka, J.
    Simon, W.
    Dempsey, J.
    MEDICAL PHYSICS, 2007, 34 (06) : 2609 - 2609
  • [49] A comparison between Gafchromic EBT3 and EBT-XD films for pre-treatment QA of SBRT/SRS plans
    Santos, T.
    Ventura, T.
    Capela, M.
    Mateus, J.
    Lopes, M. D. C.
    RADIOTHERAPY AND ONCOLOGY, 2021, 161 : S1343 - S1344
  • [50] Analysis of different procedures for absolute dosimetry with EBT3 radiochromic film
    San Miguel, F. J.
    Clemente-Gutierrez, F.
    Perez-Vara, C.
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2018, 4 (06):