Performance evaluation of an LED flatbed scanner for triple channel film dosimetry with EBT3 and EBT-XD film

被引:4
|
作者
Van der Walt, Marchant [1 ]
Marsh, Linda [1 ]
Baines, John [1 ]
Gibson, Stephen [1 ]
Shoobridge, Ariadne [1 ]
de Vine, Glenn [1 ]
机构
[1] Townsville Hosp & Hlth Serv, Dept Med Phys, Townsville, Qld 4814, Australia
关键词
Radiochromic film dosimetry; Radiotherapy; Triple channel corrections; LED; RADIOCHROMIC FILM; DOCUMENT SCANNER; CALIBRATION;
D O I
10.1007/s13246-022-01161-z
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We investigate the properties of a light emitting diode (LED) flatbed scanner for use with EBT3 and EBT-XD film types in a clinical radiochromic film (RCF) dosimetry program with modern treatment techniques. The flatbed scanner was characterised in terms of lateral and longitudinal response, X-Y scaling integrity, scanning reproducibility, scanner warm up dependence and film orientation dependence. The preferred lateral response artefact (LRA) corrections are investigated for the LED light source. Supporting evidence is provided regarding the dose independent nature of the corrections while also providing results suggesting a potential film type independence. Results from 2D gamma analysis of four patient treatments were compared between the new 12000XL and existing 10000XL model. Lastly, a dose uncertainty analysis was performed for the film-scanner system combination. It may be concluded that the lateral response variation requires correction while the longitudinal response variation is insignificant. The linear scaling in the lateral and longitudinal directions are within 0.5% and the scanner reproducibility is stable. Scanner warm up dependence no longer exists, and effort should be made to maintain all film orientation in a study set within 15 degrees. The LRA corrections are as reported substantially dose independent and there is evidence to support film type independence. Comparative gamma analysis of patient specific dose maps between the EPSON 10000XL (xenon fluorescent lamp) and 12000XL (LED) scanners showed that results are indistinguishable for both film types across the two scanner models when the necessary corrections are applied. Dose uncertainty is in agreement with the literature and can be kept below 3% with necessary corrections applied.
引用
收藏
页码:901 / 914
页数:14
相关论文
共 50 条
  • [21] Changes of the optical characteristics of radiochromic films in the transition from EBT3 to EBT-XD films
    Schoenfeld, Andreas A.
    Wieker, Soeren
    Harder, Dietrich
    Poppe, Bjoern
    PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (14): : 5426 - 5442
  • [22] Comparison of the Absorption Spectra of the EBT-XD and EBT3 Films Irradiated with Clinical Radiotherapy Beams
    Leon, Y.
    Mulrow, D.
    Khan, R.
    Darafsheh, A.
    MEDICAL PHYSICS, 2019, 46 (06) : E621 - E621
  • [23] Is EBT-XD film suitable for linac and Gamma Knife radiosurgery dosimetry verification and audit?
    Nisbet, A.
    Dimitriadis, A.
    Palmer, A. L.
    Clark, C. H.
    RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S723 - S723
  • [24] Investigating the method of selection of background pixel values for the calibration of EBT-XD film dosimetry
    Raj, Sathiya
    Shree, Nithya
    Kadirampatti, Ganesh
    PRECISION RADIATION ONCOLOGY, 2024, 8 (03): : 132 - 137
  • [25] Radiochromic film dosimetry for protons up to 10 MeV with EBT2, EBT3 and unlaminated EBT3 films
    Sanchez-Parcerisa, Daniel
    Sanz-Garcia, Irene
    Ibanez, Paula
    Espana, Samuel
    Espinosa, Andrea
    Gutierrez-Neira, Carolina
    Lopez, Alfonso
    Vera, Juan Antonio
    Mazal, Alejandro
    Fraile, Luis Mario
    Udias, Jose Manuel
    PHYSICS IN MEDICINE AND BIOLOGY, 2021, 66 (11):
  • [26] Dosimetric accuracy of Gafchromic EBT2 and EBT3 film for in vivo dosimetry
    Moylan, Rebecca
    Aland, Trent
    Kairn, Tanya
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2013, 36 (03) : 331 - 337
  • [27] Dosimetric accuracy of Gafchromic EBT2 and EBT3 film for in vivo dosimetry
    Rebecca Moylan
    Trent Aland
    Tanya Kairn
    Australasian Physical & Engineering Sciences in Medicine, 2013, 36 : 331 - 337
  • [28] Dosimetric response of Gafchromic™ EBT-XD film to therapeutic protons
    Guan, Fada
    Wang, Xiaochun
    Yang, Ming
    Draeger, Emily
    Han, Dae
    Iga, Kiminori
    Guo, Fanqing
    Perles, Luis
    Li, Yuting
    Sahoo, Narayan
    Mohan, Radhe
    Chen, Zhe
    PRECISION RADIATION ONCOLOGY, 2023, 7 (01): : 15 - 26
  • [29] In vivo dosimetry using EBT3 and EBT-XD radiochromic films for high-energy electron beam in hypofractionation keloid radiotherapy
    Lin, Yu -Fang
    Hsieh, Chen-Hsi
    Tien, Hui-Ju
    Lai, Lu-Han
    Hsu, Shih-Ming
    Shueng, Pei-Wei
    RADIATION PHYSICS AND CHEMISTRY, 2024, 218
  • [30] Evaluation of Gafchromic EBT-XD Film Using Vidar DosimetryPRO Advantage Red Scanner for SRS and SBRT QA
    Dumas, M.
    Wen, N.
    MEDICAL PHYSICS, 2017, 44 (06) : 3119 - 3120