Beam modeling and commissioning for Monte Carlo photon beam on an Elekta Versa HD LINAC

被引:2
|
作者
Can, Sumeyra [1 ]
Karacetin, Didem [2 ]
Meric, Niyazi [1 ]
机构
[1] Ankara Univ, Inst Nucl Sci, TR-06100 Ankara, Turkey
[2] Istanbul Basaksehir Cam & Sakura City Hosp, Istanbul, Turkey
关键词
Beam data measurements; Beam matching; Elekta VersaHD LINAC; Monte Carlo simulation; Monaco Treatment planning system; Virtual source model; TREATMENT PLANNING SYSTEM; DOSE CALCULATION TOOL; VIRTUAL SOURCE MODEL; CLINICAL IMPLEMENTATION; VALIDATION; IMRT; THERAPY;
D O I
10.1016/j.apradiso.2021.110054
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
Purpose: This study aims at analyzing beam data commissioning along with verifying Monte Carlo-based treatment planning system on the basis of the manufacturer guidelines for Elekta Versa HD Linear Accelerator. Moreover, evaluating the beam match process in terms of quality index, photon profile and multi leaf collimator (MLC) offset is aimed as well. Materials and methods: The process of collecting beam data for Monaco 5.51 Treatment Planning System commissioning was done based on the instructions provided by the manufacturer as well as AAPM TG-106. Monte Carlo analysis was done for output factors in water, percent depth dose and beam profiles. A set of eight static and intensity modulated radiation therapy fields were used to verify the MLC parameters. Results: The difference between the measured and modeled penetrative quality (D-10) was achieved to be 0.54%. The output factors for 6 MV photon energy were measured and the difference between the measured and Monte Carlo output results was smaller than 1% for all the fields. The average percentage of passing the gamma criteria for commissioning test fields was (95+-4)%, however, the minimum agreement was 87.5% belonging to "7SEGA". Additionally, the agreement between both LINAC is 96%, however, the second LINAC reveals a positive offset in the point of approximately -4 cm on the x-axis at the crossplane. Conclusion: Test commissioning was successfully verified using a homogeneous phantom for point dose measurements, post modelling MLC parameters and patient QA plans. All plan parameters pass the gamma criteria. 6 MV photon beam was successfully commissioned for Elekta VersaHD LINAC and is ready for clinical implementation.
引用
收藏
页数:10
相关论文
共 50 条
  • [31] Determination of the initial beam parameters in Monte Carlo linac simulation
    Aljarrah, K
    Sharp, GC
    Neicu, T
    Jiang, SB
    MEDICAL PHYSICS, 2006, 33 (04) : 850 - 858
  • [32] Beam dynamics for beam commissioning of RAON linac and KoBRA beam line
    Ji-Ho Jang
    Dong-O. Jeon
    Hyunchang Jin
    Journal of the Korean Physical Society, 2023, 82 : 1062 - 1077
  • [33] Beam dynamics for beam commissioning of RAON linac and KoBRA beam line
    Jang, Ji-Ho
    Jeon, Dong-O.
    Jin, Hyunchang
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2023, 82 (11) : 1062 - 1077
  • [34] Commissioning and validation of a single photon beam model in RayStation for multiple matched Elekta Linacs
    Su, Lin
    Huang, Ellen
    Miles, Devin A.
    Farjam, Reza
    Marsh, Ian R.
    Li, Qiongge
    Moore, Joseph A.
    McNutt, Todd R.
    Ding, Kai
    Wang, Ken Kang-Hsin
    Robinson, Adam
    Kuri, Gregory
    Seabrease, Richard
    Adam, David P.
    Oglesby, Ryan
    Shen, Bin
    Wu, Binbin
    Lee, Junghoon
    Jia, Xun
    Han-Oh, Sarah
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2024, 25 (10):
  • [35] Effects of collimator backscatter in an Elekta linac by Monte Carlo simulation
    Kairn, T.
    Crowe, S. B.
    Poole, C. M.
    Fielding, A. L.
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2009, 32 (03) : 129 - 135
  • [36] Beam commissioning of BEPCII injector linac
    Chen, Yan-Wei
    Wang, Shu-Hong
    Pei, Guo-Xi
    Liu, Wei-Bin
    Ye, Qiang
    Kao Neng Wu Li Yu Ho Wu Li/High Energy Physics and Nuclear Physics, 2006, 30 (06): : 562 - 565
  • [37] Commissioning of a medical accelerator photon beam Monte Carlo simulation using wide-field profiles
    Pena, J
    Franco, L
    Gómez, F
    Iglesias, A
    Lobato, R
    Mosquera, J
    Pazos, A
    Pardo, J
    Pombar, M
    Rodríguez, A
    Sendón, J
    PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (21): : 4929 - 4942
  • [38] Beam commissioning for a superconducting proton linac
    Wang, Zhi-Jun
    He, Yuan
    Jia, Huan
    Dou, Wei-ping
    Chen, Wei-long
    Zhang, X. L.
    Liu, Shu-hui
    Feng, Chi
    Tao, Yue
    Wang, Wang-sheng
    Wu, Jian-qiang
    Zhang, Sheng-hu
    Zhao, Hong-Wei
    PHYSICAL REVIEW ACCELERATORS AND BEAMS, 2016, 19 (12):
  • [39] Development and commissioning of a Monte Carlo photon beam model for the forthcoming clinical trials in microbeam radiation therapy
    Martinez-Rovira, I.
    Sempau, J.
    Prezado, Y.
    MEDICAL PHYSICS, 2012, 39 (01) : 119 - 131
  • [40] PHOTON BEAM CHARACTERISTICS AT MONITOR CHAMBER LEVEL IN A FLATTENING FILTER FREE LINAC: A MONTE CARLO STUDY
    Lind, M.
    Knoos, T.
    Ceberg, C.
    Wieslander, E.
    McClean, B.
    Georg, D.
    RADIOTHERAPY AND ONCOLOGY, 2009, 92 : S57 - S57