Beam modeling and commissioning for Monte Carlo photon beam on an Elekta Versa HD LINAC

被引:2
|
作者
Can, Sumeyra [1 ]
Karacetin, Didem [2 ]
Meric, Niyazi [1 ]
机构
[1] Ankara Univ, Inst Nucl Sci, TR-06100 Ankara, Turkey
[2] Istanbul Basaksehir Cam & Sakura City Hosp, Istanbul, Turkey
关键词
Beam data measurements; Beam matching; Elekta VersaHD LINAC; Monte Carlo simulation; Monaco Treatment planning system; Virtual source model; TREATMENT PLANNING SYSTEM; DOSE CALCULATION TOOL; VIRTUAL SOURCE MODEL; CLINICAL IMPLEMENTATION; VALIDATION; IMRT; THERAPY;
D O I
10.1016/j.apradiso.2021.110054
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
Purpose: This study aims at analyzing beam data commissioning along with verifying Monte Carlo-based treatment planning system on the basis of the manufacturer guidelines for Elekta Versa HD Linear Accelerator. Moreover, evaluating the beam match process in terms of quality index, photon profile and multi leaf collimator (MLC) offset is aimed as well. Materials and methods: The process of collecting beam data for Monaco 5.51 Treatment Planning System commissioning was done based on the instructions provided by the manufacturer as well as AAPM TG-106. Monte Carlo analysis was done for output factors in water, percent depth dose and beam profiles. A set of eight static and intensity modulated radiation therapy fields were used to verify the MLC parameters. Results: The difference between the measured and modeled penetrative quality (D-10) was achieved to be 0.54%. The output factors for 6 MV photon energy were measured and the difference between the measured and Monte Carlo output results was smaller than 1% for all the fields. The average percentage of passing the gamma criteria for commissioning test fields was (95+-4)%, however, the minimum agreement was 87.5% belonging to "7SEGA". Additionally, the agreement between both LINAC is 96%, however, the second LINAC reveals a positive offset in the point of approximately -4 cm on the x-axis at the crossplane. Conclusion: Test commissioning was successfully verified using a homogeneous phantom for point dose measurements, post modelling MLC parameters and patient QA plans. All plan parameters pass the gamma criteria. 6 MV photon beam was successfully commissioned for Elekta VersaHD LINAC and is ready for clinical implementation.
引用
收藏
页数:10
相关论文
共 50 条
  • [21] Monte carlo model and output factors of elekta Infinity™ 6 and 10 MV photon beam
    Yani, Sitti
    Budiansah, Indra
    Rhani, Mohamad Fahdillah
    Haryanto, Freddy
    REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2020, 25 (04) : 470 - 478
  • [22] Monte Carlo simulation of Varian Linac for 6 MV photon beam with BEAMnrc code
    Mohammed, Maged
    El Bardouni, T.
    Chakir, E.
    Boukhal, H.
    Saeed, M.
    Ahmed, Abdul-Aziz
    RADIATION PHYSICS AND CHEMISTRY, 2018, 144 : 69 - 75
  • [23] MONTE-CARLO CALCULATION OF THE PHOTON-BEAM SPECTRA OF A DUAL ENERGY LINAC
    BOYER, AL
    CASSADY, AE
    WANG, LL
    MEDICAL PHYSICS, 1987, 14 (03) : 452 - 452
  • [24] On the commissioning of Monte Carlo beam models and dose profile corrections on stereotactic photon beams
    Papaconstadopoulos, P.
    Tessier, F.
    Syme, A.
    Seuntjens., J.
    RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S107 - S108
  • [25] Electron beam commissioning for Monte Carlo dose calculation
    Deng, J
    Jiang, SB
    Pawlicki, T
    Li, JS
    Ma, CM
    USE OF COMPUTERS IN RADIATION THERAPY, 2000, : 431 - 433
  • [26] Commissioning of a Varian Clinac iX 6 MV Photon Beam using Monte Carlo Simulation
    Dirgayussa, I. Gde Eka
    Yani, Sitti
    Rhani, M. Fahdillah
    Haryanto, Freddy
    5TH INTERNATIONAL CONFERENCE ON MATHEMATICS AND NATURAL SCIENCES (ICMNS 2014), 2015, 1677
  • [27] Monte Carlo Commissioning of 6 MV Photon Beam of Varian Linear Accelerator Installed at INOR
    Bajwa, Saqib
    Gul, Attia
    Ahmed, Shahbaz
    Kakakhel, Muhammad Basim
    2016 INTERNATIONAL CONFERENCE ON EMERGING TECHNOLOGIES (ICET), 2016,
  • [28] Commissioning and Validation of Brainlab Elements Multiple Metastases with An Elekta Versa HD and ExacTrac
    Ruiz, B.
    Hill, J.
    Boles, L.
    MEDICAL PHYSICS, 2018, 45 (06) : E177 - E177
  • [29] An automated beam commissioning procedure for Monte Carlo treatment planning
    Qin, L
    Li, J
    Jiang, S
    Deng, J
    Ma, C
    MEDICAL PHYSICS, 2002, 29 (06) : 1232 - 1232
  • [30] Source modelling and beam commissioning for Monte Carlo treatment planning
    Ma, C
    Faddegon, B
    Curran, B
    MEDICAL PHYSICS, 2005, 32 (06) : 2119 - 2119