Monte carlo model and output factors of elekta Infinity™ 6 and 10 MV photon beam

被引:6
|
作者
Yani, Sitti [1 ,2 ]
Budiansah, Indra [2 ]
Rhani, Mohamad Fahdillah [3 ]
Haryanto, Freddy [2 ]
机构
[1] Bogor Agr Univ, IPB Univ, Fac Math & Nat Sci, Dept Phys, Bogor, Indonesia
[2] Inst Teknol Bandung, Fac Math & Nat Sci, Dept Phys, Jalan Ganesa 10, Bandung, Indonesia
[3] Concord Int Hosp, Radiol Dept, 19 Adam Rd, Singapore, Singapore
关键词
Commissioning; Elekta Infinity (TM); Photon beam; Output factors; SMALL-FIELD DOSIMETRY; SIMULATION; RADIOTHERAPY; VALIDATION; WATER; SPECTRA;
D O I
10.1016/j.rpor.2020.03.021
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Aim: This study aimed to commission the Elekta Infinity (TM) working in 6 and 10 MV photon beam installed in Concord International Hospital, Singapore, and compare the OFs between MC simulation and measurement using PTW semiflex and microDiamond detector for small field sizes. Material and Methods: There are two main steps in this study: modelling of Linac 6 and 10 MV photon beam and analysis of the output factors for field size 2 x 2-10 x 10 cm(2). The EGSnrc/BEAMnrc-DOSXYZnrc code was used to model and characterize the Linac and to calculate the dose distributions in a water phantom. The dose distribution and OFs were compared to the measurement data in the same condition. Results: The commissioning process was only conducted for a 10 x 10 cm(2) field size. The PDD obtained from MC simulation showed a good agreement with the measurement. The local dose difference of PDDs was less than 2% for 6 and 10 MV. The initial electron energy was 5.2 and 9.4 MeV for 6 and 10 MV photon beam, respectively. This Linac model can be used for dose calculation in other situations and different field sizes because this Linac has been commissioned and validated using Monte Carlo simulation. The 10 MV Linac produces higher electron contamination than that of 6 MV. Conclusions: The Linac model in this study was acceptable. The most important result in this work comes from OFs resulted from MC calculation. This value was more significant than the OFs from measurement using semiflex and microDiamond for all beam energy and field sizes because of the CPE phenomenon. (C) 2020 Greater Poland Cancer Centre. Published by Elsevier B.V. All rights reserved.
引用
收藏
页码:470 / 478
页数:9
相关论文
共 50 条
  • [1] A Monte Carlo model of an agility head for a 6 MV Elekta photon beam
    Al-Saleh, Wafa M.
    Almatani, Turki
    [J]. RADIATION PHYSICS AND CHEMISTRY, 2024, 216
  • [2] An estimation of relative output factors of 6 MV photon beam using Monte Carlo simulation
    Saprangsi, W.
    Phaisangittisakul, N.
    [J]. MEDICAL PHYSICS, 2007, 34 (06) : 2416 - 2416
  • [3] A Monte Carlo model of an agility head for a 10-MV photon beam
    Almatani, Turki
    Hugtenburg, Richard P.
    Smakovs, Artjoms
    [J]. JOURNAL OF TAIBAH UNIVERSITY FOR SCIENCE, 2022, 16 (01): : 300 - 307
  • [4] Monte Carlo study on electron contamination and output factors of small field dosimetry in 6 MV photon beam
    Yani, Sitti
    Dirgayussa, I. Gde Eka
    Rhani, Mohamad Fahdillah
    Soh, Roger C. X.
    Haryanto, Freddy
    Arif, Idam
    [J]. SMART SCIENCE, 2016, 4 (02): : 87 - 94
  • [5] Optimisation of variance reduction techniques in EGSnrc Monte Carlo for a 6 MV photon beam of an Elekta Synergy linear accelerator
    Almatani, Turki
    [J]. JOURNAL OF KING SAUD UNIVERSITY SCIENCE, 2021, 33 (04)
  • [6] Monte Carlo model of the Elekta SLiplus accelerator:: validation of a new MLC component module in BEAM for a 6 MV beam
    Van de Walle, J
    Martens, C
    Reynaert, N
    Palmans, H
    Coghe, M
    De Neve, W
    De Wagter, C
    Thierens, H
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (03): : 371 - 385
  • [7] Verification of a Monte Carlo-Based Source Model for a Varian 10 MV Photon Beam
    Davidson, S.
    Cui, J.
    Deasy, J.
    Ibbott, G.
    Followill, D.
    [J]. MEDICAL PHYSICS, 2008, 35 (06)
  • [8] Validation of BEAMnrc Monte Carlo model for a 12 MV photon beam
    Mohammed, Maged
    El Bardouni, T.
    Chakir, E.
    Saeed, M.
    Jamal, Al Zain
    Mohamed, Landour
    [J]. JOURNAL OF KING SAUD UNIVERSITY SCIENCE, 2018, 30 (04) : 537 - 543
  • [9] Validation of PRIMO Monte Carlo Model of Clinac®iX 6MV Photon Beam
    Sarin, B.
    Bindhu, B.
    Saju, B.
    Nair, Raguram K.
    [J]. JOURNAL OF MEDICAL PHYSICS, 2020, 45 (01) : 24 - 35
  • [10] A multiple source model for 6 MV photon beam dose calculations using Monte Carlo
    Fix, MK
    Stampanoni, M
    Manser, P
    Born, EJ
    Mini, R
    Rüegsegger, P
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (05): : 1407 - 1427