Monte carlo model and output factors of elekta Infinity™ 6 and 10 MV photon beam

被引:6
|
作者
Yani, Sitti [1 ,2 ]
Budiansah, Indra [2 ]
Rhani, Mohamad Fahdillah [3 ]
Haryanto, Freddy [2 ]
机构
[1] Bogor Agr Univ, IPB Univ, Fac Math & Nat Sci, Dept Phys, Bogor, Indonesia
[2] Inst Teknol Bandung, Fac Math & Nat Sci, Dept Phys, Jalan Ganesa 10, Bandung, Indonesia
[3] Concord Int Hosp, Radiol Dept, 19 Adam Rd, Singapore, Singapore
关键词
Commissioning; Elekta Infinity (TM); Photon beam; Output factors; SMALL-FIELD DOSIMETRY; SIMULATION; RADIOTHERAPY; VALIDATION; WATER; SPECTRA;
D O I
10.1016/j.rpor.2020.03.021
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Aim: This study aimed to commission the Elekta Infinity (TM) working in 6 and 10 MV photon beam installed in Concord International Hospital, Singapore, and compare the OFs between MC simulation and measurement using PTW semiflex and microDiamond detector for small field sizes. Material and Methods: There are two main steps in this study: modelling of Linac 6 and 10 MV photon beam and analysis of the output factors for field size 2 x 2-10 x 10 cm(2). The EGSnrc/BEAMnrc-DOSXYZnrc code was used to model and characterize the Linac and to calculate the dose distributions in a water phantom. The dose distribution and OFs were compared to the measurement data in the same condition. Results: The commissioning process was only conducted for a 10 x 10 cm(2) field size. The PDD obtained from MC simulation showed a good agreement with the measurement. The local dose difference of PDDs was less than 2% for 6 and 10 MV. The initial electron energy was 5.2 and 9.4 MeV for 6 and 10 MV photon beam, respectively. This Linac model can be used for dose calculation in other situations and different field sizes because this Linac has been commissioned and validated using Monte Carlo simulation. The 10 MV Linac produces higher electron contamination than that of 6 MV. Conclusions: The Linac model in this study was acceptable. The most important result in this work comes from OFs resulted from MC calculation. This value was more significant than the OFs from measurement using semiflex and microDiamond for all beam energy and field sizes because of the CPE phenomenon. (C) 2020 Greater Poland Cancer Centre. Published by Elsevier B.V. All rights reserved.
引用
收藏
页码:470 / 478
页数:9
相关论文
共 50 条
  • [41] Development of An Accurate Monte Carlo Model of the Standard Clinac 6MV Beam
    Egan, A.
    Laub, W.
    [J]. MEDICAL PHYSICS, 2013, 40 (06)
  • [42] Monte Carlo calculated output correction factors for nine small field detectors in Varian Clinac iX 6 MV photon beams
    Benmakhlouf, H.
    Sempau, J.
    Andreo, P.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S201 - S202
  • [43] Monte Carlo estimation of electron contamination in a 18 MV clinical photon beam
    Allahverdi, M.
    Zabihzadeh, M.
    Ay, M. R.
    Mahdavi, S. R.
    Shahriari, M.
    Mesbahi, A.
    Alijanzadeh, H.
    [J]. IRANIAN JOURNAL OF RADIATION RESEARCH, 2011, 9 (01): : 15 - 28
  • [44] Investigation of photon beam output factors for conformal radiation therapy -: Monte Carlo simulations and measurements
    Haryanto, F
    Fippel, M
    Laub, W
    Dohm, O
    Nüsslin, F
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (11): : N133 - N143
  • [45] Dosimetric properties of a flattening filter-free 6-MV photon beam: A Monte Carlo study
    Mesbahi A.
    Mehnati P.
    Keshtkar A.
    Farajollahi A.
    [J]. Radiation Medicine, 2007, 25 (7): : 315 - 324
  • [46] Validation of GEANT4 Monte Carlo Simulation Code for 6 MV Varian Linac Photon Beam
    Salama, E.
    Ali, A. S.
    Emad, N.
    Radi, A.
    [J]. ARAB JOURNAL OF NUCLEAR SCIENCES AND APPLICATIONS, 2015, 48 (02): : 70 - 76
  • [47] A Monte Carlo study of the variation of electron fluence in water from a 6 MV photon beam outside of the field
    Kirkby, C.
    Field, C.
    MacKenzie, M.
    Syme, A.
    Fallone, B. G.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (12): : 3563 - 3578
  • [48] Commissioning of Elekta InfinityTM 6 MV flattening filter-free using Monte Carlo simulation
    Maulana, Iswan Fadhil
    Yani, Sitti
    Sumaryada, Tony
    Rhani, Mohamad Fahdillah
    Haryanto, Freddy
    [J]. RADIATION PHYSICS AND CHEMISTRY, 2023, 210
  • [49] Optimization of Efficiency on DOSXYZnrc Monte Carlo Code Parameters for a 12 MV Photon Beam
    Zeghari, Abdelkrim
    Saaidi, Rahal
    El Moursli, Rajaa Cherkaoui
    Kaddouch, Said
    [J]. 2022 2ND INTERNATIONAL CONFERENCE ON INNOVATIVE RESEARCH IN APPLIED SCIENCE, ENGINEERING AND TECHNOLOGY (IRASET'2022), 2022, : 1033 - 1037
  • [50] Validation of a 10 MV photon beam Elekta Synergy linear accelerator using the BEAMnrc MC code
    Almatani, Turki
    [J]. JOURNAL OF KING SAUD UNIVERSITY SCIENCE, 2021, 33 (04)