Evaluation of focused multipolar stimulation for cochlear implants in long-term deafened cats

被引:31
|
作者
George, Shefin S. [1 ,2 ]
Wise, Andrew K. [1 ,2 ]
Fallon, James B. [1 ,2 ]
Shepherd, Robert K. [1 ,2 ]
机构
[1] Bion Inst, East Melbourne 3002, Australia
[2] Univ Melbourne, Dept Med Bion, Melbourne, Vic 3002, Australia
基金
澳大利亚国家健康与医学研究理事会;
关键词
cochlear implant; current focusing; electrode configuration; spatial tuning curve; inferior colliculus; deafness; ELECTRICAL-STIMULATION; SPIRAL GANGLION; SPATIAL SELECTIVITY; SPEECH RECOGNITION; INFERIOR COLLICULUS; SPECTRAL CHANNELS; AUDITORY-NERVE; TRIPOLAR; PERCEPTION; EXCITATION;
D O I
10.1088/1741-2560/12/3/036003
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective. Focused multipolar (FMP) stimulation has been shown to produce restricted neural activation using intracochlear stimulation in animals with a normal population of spiral ganglion neurons (SGNs). However, in a clinical setting, the widespread loss of SGNs and peripheral fibres following deafness is expected to influence the effectiveness of FMP. Approach. We compared the efficacy of FMP stimulation to both monopolar (MP) and tripolar (TP) stimulation in long-term deafened cat cochleae (n = 8). Unlike our previous study, these cochleae contained <10% of the normal SGN population adjacent to the electrode array. We also evaluated the effect of electrode position on stimulation modes by using either modiolar facing or lateral wall facing half-band electrodes. The spread of neural activity across the inferior colliculus, a major nucleus within the central auditory pathway, was used as a measure of spatial selectivity. Main results. In cochleae with significant SGN degeneration, we observed that FMP and TP stimulation resulted in greater spatial selectivity than MP stimulation (p < 0.001). However, thresholds were significantly higher for FMP and TP stimulation compared to MP stimulation (p < 0.001). No difference between FMP and TP stimulation was found in any measures. The high threshold levels for FMP stimulation was significantly reduced without compromising spatial selectivity by varying the degree of current focusing (referred as 'partial-FMP' stimulation). Spatial selectivity of all stimulation modes was unaffected by the electrode position. Finally, spatial selectivity in long-term deafened cochleae was significantly less than that of cochleae with normal SGN population (George S S et al 2014 J. Neural Eng. 11 065003). Significance. The present results indicate that the greater spatial selectivity of FMP and TP stimulation over MP stimulation is maintained in cochleae with significant neural degeneration and is not adversely affected by electrode position. The greater spatial selectivity of FMP and TP stimulation would be expected to result in improved clinical performance.
引用
收藏
页数:15
相关论文
共 50 条
  • [11] Behavioral frequency discrimination ability of partially deafened cats using cochlear implants
    Benovitski, Yuri B.
    Blamey, Peter J.
    Rathbone, Graeme D.
    Fallon, James B.
    HEARING RESEARCH, 2014, 315 : 61 - 66
  • [12] An Open Expandable Data Protocol for Multipolar Stimulation in Cochlear Implants
    Akinin, Abraham
    Haque, Razi-ul
    2023 45TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE & BIOLOGY SOCIETY, EMBC, 2023,
  • [13] LONG-TERM RESULTS OF EARLY COCHLEAR IMPLANTATION IN CONGENITALLY AND PRELINGUALLY DEAFENED CHILDREN
    WALTZMAN, SB
    COHEN, NL
    GOMOLIN, RH
    SHAPIRO, WH
    OZDAMAR, SR
    HOFFMAN, RA
    AMERICAN JOURNAL OF OTOLOGY, 1994, 15 : 9 - 13
  • [14] Long-term functional outcomes of cochlear implants in children
    Laszig, R.
    Aschendorff, A.
    Beck, R.
    Schild, C.
    Kroeger, S.
    Wesarg, T.
    Arndt, S.
    HNO, 2009, 57 (07) : 657 - 662
  • [15] Stimulation of the cochlear nucleus with multichannel auditory brainstem implants and long-term results: Freiburg patients
    Marangos, N
    Stecker, M
    Sollmann, WP
    Laszig, R
    JOURNAL OF LARYNGOLOGY AND OTOLOGY, 2000, 114 : 27 - 31
  • [16] LONG-TERM EVALUATION OF TITANIUM IMPLANTS
    ALBREKTSSON, T
    BRANEMARK, PI
    TJELLSTROM, A
    ACTA ORTHOPAEDICA SCANDINAVICA, 1988, 59 (03): : 345 - 346
  • [17] Long-term speech perception performance in prelingually deafened adult cochlear implant recipients
    Yoshida, Haruo
    Kanda, Yukihiko
    Satoh, Chisei
    Kumai, Yoshihiko
    Takahashi, Haruo
    COCHLEAR IMPLANTS INTERNATIONAL, 2023, 24 (05) : 243 - 249
  • [18] COCHLEAR PATHOLOGY OF LONG-TERM NEOMYCIN INDUCED DEAFNESS IN CATS
    LEAKE, PA
    HRADEK, GT
    HEARING RESEARCH, 1988, 33 (01) : 11 - 34
  • [19] Long-term results of cochlear implants in children with residual hearing
    Gantz, BJ
    Rubinstein, JT
    Tyler, RS
    Teagle, HFB
    Cohen, NL
    Waltzman, SB
    Miyamoto, RT
    Kirk, KI
    ANNALS OF OTOLOGY RHINOLOGY AND LARYNGOLOGY, 2000, 109 (12): : 33 - 36
  • [20] Long-term monaural auditory deprivation and bilateral cochlear implants
    Boisvert, Isabelle
    McMahon, Catherine M.
    Dowell, Richard C.
    NEUROREPORT, 2012, 23 (03) : 195 - 199