A novel tool for in vivo dosimetry in diagnostic and interventional radiology using plastic scintillation detectors

被引:5
|
作者
Boivin, J. [1 ,2 ,3 ]
Beddar, S. [4 ]
Guillemette, M. [2 ,3 ,5 ]
Beaulieu, L. [1 ,2 ,3 ]
机构
[1] CHU Quebec, Dept Radiooncol & Axe Oncol, Ctr Rech, Quebec City, PQ, Canada
[2] Univ Laval, Dept Phys Genie Phys & Opt, Quebec City, PQ, Canada
[3] Univ Laval, Ctr Rech Canc, Quebec City, PQ, Canada
[4] Univ Texas MD Anderson Canc Ctr, Houston, TX 77030 USA
[5] Inst Univ Cardiol & Pneumol Quebec, Quebec City, PQ, Canada
关键词
PSD; plastic scintillation detector; dosimetry; photodetector; interventional radiology;
D O I
10.1007/978-3-319-19387-8_167
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
A plastic scintillation detector (PSD) has been developed and optimized to measure dose rate in real-time for low energy beams. The PSD is composed of a scintillating fiber coupled to a clear optical fiber transmitting the scintillator's light to a photodetector. A spectrometer was used to characterize the scintillator's spectrum under low and high energy exposures. Seven photodetectors were then evaluated to identify their operating range and potential applications. They comprise a photomultiplier tube (PMT), an avalanche photodiode, two passive diodes of which, and a set of three CCD cameras. The scintillator was exposed to low energy potential beams (120 kVp, 180 kVp, and 220 kVp) of an orthovoltage unit. The source-to-detector distance was varied to explore a broad dose rate range likely to be used in radiology and superficial treatment. Every detector could measure dose rate down to 10 mGy/s while keeping a relative standard deviation below 2 %. The CCD cameras were the less sensitive devices, but they allow multiple fibers to be read simultaneously. Among the photodetectors, the PMT was found to be the most sensitive detector with a relative standard deviation of less than 1 % at the lowest dose rate available. The PMT was then included in the PSD design for an in vivo study in interventional radiology where low dose rate sensitivity is essential. The PSD was located inside a plastic water phantom to measure skin and depth dose from 1 mm down to 24 cm. There was less than a 2 % difference between the PSD measured dose rate and the ion chamber reading located at the same depth. These results indicate that a broad range of photodetectors can be used in the PSD design, but low dose rate measurements require very sensitive devices such as a PMT.
引用
收藏
页码:680 / 684
页数:5
相关论文
共 50 条
  • [31] Pre-clinical and clinical evaluation of the HYPERSCINT plastic scintillation dosimetry research platform for in vivo dosimetry during radiotherapy
    Schoepper, Imke
    Dieterich, Sonja
    Trestrail, Earl Alonzo
    Kent, Michael Sean
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2022, 23 (04):
  • [32] Measurement of Small Field Output Factors Using Plastic Scintillation Detectors
    Klein, D.
    Tailor, R.
    Archambault, L.
    Wang, L.
    Therriault-Proulx, F.
    Beddar, S.
    MEDICAL PHYSICS, 2010, 37 (06)
  • [33] Stemless plastic scintillation detectors-A novel radiation dosimeter with a bright future
    Hupman, Allan
    Syme, Alasdair
    Hill, Ian
    MEDICAL PHYSICS, 2021, 48 (08) : 4685 - 4685
  • [34] BEST IN PHYSICS (THERAPY) - A Novel Multi-Point Plastic Scintillation Detector for in Vivo Dosimetry and Quality Assurance in Radiation Therapy
    Therriault-Proulx, F.
    Beaulieu, L.
    Archambault, L.
    Beddar, S.
    MEDICAL PHYSICS, 2012, 39 (06) : 3967 - 3967
  • [35] Development of dosimetry using detectors of diagnostic digital radiography systems
    Ariga, Eiji
    Ito, Shigeki
    Deji, Shizuhiko
    Saze, Takuya
    Nishizawa, Kunihide
    MEDICAL PHYSICS, 2007, 34 (01) : 166 - 174
  • [36] Dosimetry of Occupational Eye Lens Dose Using a Novel Direct Eye Dosimeter, DOSIRIS, during Interventional Radiology Procedures
    Hirakawa, Masakazu
    Nakatake, Hiroshi
    Tsuruta, Satoru
    Matsuura, Shuji
    Motomura, Yuushi
    Hiraki, Yoshiki
    Mimori, Koshi
    Ishigami, Kousei
    INTERVENTIONAL RADIOLOGY, 2022, 7 (02) : 40 - 43
  • [37] Radon dosimetry in typical Indian dwellings using plastic track detectors
    Kant, K
    Upadhyay, SB
    Sharma, GS
    Chakarvart, SK
    INDOOR AND BUILT ENVIRONMENT, 2006, 15 (02) : 187 - 191
  • [38] DOSIMETRY OF PI--MESONS USING SILICON DETECTORS AND PLASTIC SCINTILLATORS
    RAJU, MR
    LAMPO, E
    CURTIS, SB
    RICHMAN, C
    PHYSICS IN MEDICINE AND BIOLOGY, 1971, 16 (04): : 599 - +
  • [39] Proton Therapy Dosimetry by Using the Scintillation of Glass and Plastic Bare Optical Fibers
    Darafsheh, A.
    Taleei, R.
    Kassaee, A.
    Finlay, J.
    MEDICAL PHYSICS, 2017, 44 (06)
  • [40] DOSIMETRY USING GAFCHROMIC XR-RV2 RADIOCHROMIC FILMS IN INTERVENTIONAL RADIOLOGY
    Neocleous, A.
    Yakoumakis, E.
    Gialousis, G.
    Dimitriadis, A.
    Yakoumakis, N.
    Georgiou, E.
    RADIATION PROTECTION DOSIMETRY, 2011, 147 (1-2) : 78 - 82