Intraocular Pressure Measurement with Pneumatonometry and a Tonometer Tip Cover During Negative Pressure Application

被引:2
|
作者
Brambilla, Enrico [1 ]
Ferguson, Tanner J. [2 ]
Chu, Nathan [1 ]
Ammar, David [3 ]
Yoo, Paul [1 ]
机构
[1] Equinox Ophthalm Inc, Newport Beach, CA USA
[2] Vance Thompson Vis, 3101 W 57th St, Sioux Falls, SD 57108 USA
[3] Lions Eye Inst Transplant & Res, Tampa, FL USA
来源
CLINICAL OPHTHALMOLOGY | 2022年 / 16卷
关键词
tonometry; intraocular pressure; TOP; multi-pressure dial; MPD; ophthalmic techniques; diagnostics; GLAUCOMA PROGRESSION;
D O I
10.2147/OPTH.S359605
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: This is a 2-part study to investigate the agreement between pneumatonometry and direct pressure transducer intraocular pressure ( IOP) measurements in a perfusion organ culture (POC) model where (1) the perfusion fluid column is open to atmospheric pressure, holding IOP constant to permit evaluation of the impact of negative pressure (NP) on IOP measurements, and (2) the perfusion fluid column is a closed system, allowing IOP to vary with NP application. Methods: The first part incorporated a fluid column open to atmospheric pressure, maintaining IOP constant to permit evaluation of the effect of applied NP on IOP measurement accuracy. In the second part, the POC column was closed, allowing IOP to vary with NP application and permit evaluation of agreement between pneumatonometry and pressure transducer measurements. In each part, four perfused tissues were used in thirteen paired pre-set IOP (10, 20, 25, 30 mmHg) and NP (0, 5, 10, 15, 20 mmHg) combinations, resulting in a total of 1040 paired measurements (520 per study). The difference in IOP measurements (Delta IOP = Excursion tonometry - pressure transducer) was calculated at each paired configuration. Results: During the first part, the mean Delta IOP was -0.7 +/- 1.6 mmHg across all measurements. During the second part, the mean Delta IOP across all measurements was +0.7 +/- 1.4 mmHg. At NP settings of -5, -10, -15, and -20 mmHg, across all pre-set IOPs, the mean TOP reduction via Excursion tonometry was 3.1 +/- 0.3, 5.6 +/- 1.3, 8.5 +/- 1.7 and 11.2 +/- 1.8 mmHg, respectively. Conclusion: Measurement of IOP via Excursion tonometry yields results within the accuracy range of the pneumatonometry device (per manufacturer) and is minimally impacted by NP application. The IOP-lowering results are consistent with previous studies and further support the effectiveness of the Multi-Pressure Dial in lowering IOP relative to atmospheric pressure.
引用
收藏
页码:1289 / 1300
页数:12
相关论文
共 50 条
  • [21] Noninvasive measurement of intraocular pressure in the conscious rat with the TonoPen tonometer.
    Pan, J
    Tremblay, F
    Archibald, ML
    Kelly, MEM
    Chauhan, BC
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2000, 41 (04) : S897 - S897
  • [22] Does patient comfort influence the choice of tonometer for the measurement of intraocular pressure?
    Ugalahi, Mary O.
    Seidu, Mukaila A.
    Olusanya, Bolutife A.
    Baiyeroju, Aderonke M.
    INTERNATIONAL OPHTHALMOLOGY, 2016, 36 (03) : 385 - 389
  • [23] Effect of Intraocular Pressure on Aerosol Density Generated by Noncontact Tonometer Measurement
    Tang, Yuan
    Li, Chunchun
    Chen, Yanyan
    Chen, Zhangyan
    Zhang, Peihua
    Wang, Aisun
    Huang, Xiaoqiong
    Qu, Jia
    Li, Mengchen
    Ma, Siwen
    Vasudevan, Balamurali
    JOURNAL OF GLAUCOMA, 2020, 29 (11) : 1001 - 1005
  • [24] Comparison of intraocular pressure measurement with Schiotz tonometer and Tono-Pen Vet tonometer in healthy dogs
    Wrzesniewska, Karolina
    Madany, Jacek
    Winiarczyk, Dagmara
    JOURNAL OF VETERINARY RESEARCH, 2018, 62 (02) : 243 - 247
  • [25] PRESSURE MEASUREMENT - WHICH TONOMETER
    KAUFMAN, HE
    INVESTIGATIVE OPHTHALMOLOGY, 1972, 11 (02): : 80 - &
  • [26] Comparison of Pneumatonometry and Transpalpebral Tonometry Measurements of Intraocular Pressure during Scleral Lens Wear
    Fogt, Jennifer Swingle
    Nau, Cherie B.
    Schornack, Muriel
    Shorter, Ellen
    Nau, Amy
    Harthan, Jennifer S.
    OPTOMETRY AND VISION SCIENCE, 2020, 97 (09) : 711 - 719
  • [27] Effects of baffie and intraocular pressure on aerosols generated in the noncontact tonometer measurement during COVID-19
    Yuan Tang
    Yan-Yan Chen
    Chun-Chun Li
    Zhang-Yan Chen
    Chen Chen
    Si-Qi Wen
    Xiao-Qiong Huang
    Jia Qu
    Yan-Miao Chen
    Ai-Ai Chen
    International Journal of Ophthalmology, 2022, (04) : 533 - 540
  • [28] Effects of baffle and intraocular pressure on aerosols generated in the noncontact tonometer measurement during COVID-19
    Tang, Yuan
    Chen, Yan-Yan
    Li, Chun-Chun
    Chen, Zhang-Yan
    Chen, Chen
    Wen, Si-Qi
    Huang, Xiao-Qiong
    Qu, Jia
    Chen, Yan-Miao
    Chen, Ai-Ai
    INTERNATIONAL JOURNAL OF OPHTHALMOLOGY, 2022, 15 (04) : 533 - 540
  • [29] Evaluation of monkey intraocular pressure by rebound tonometer
    Yu, Wenhan
    Cao, Guiqun
    Qiu, Jinghua
    Liu, Xuyang
    Ma, Jia
    Li, Ni
    Yu, Man
    Yan, Naihong
    Chen, Lei
    Pang, Iok-Hou
    MOLECULAR VISION, 2009, 15 (235-36): : 2196 - 2201
  • [30] The Measurement of Intraocular Pressure in Eyes with Keratoconus Using Pascal Dynamic Contour Tonometer
    Ocakoglu, Ozcan
    Iskeleli, Guzin
    Ucar, Didar
    TURK OFTALMOLOJI DERGISI-TURKISH JOURNAL OF OPHTHALMOLOGY, 2008, 38 (03): : 185 - 190