Comparison of Pneumatonometry and Transpalpebral Tonometry Measurements of Intraocular Pressure during Scleral Lens Wear

被引:13
|
作者
Fogt, Jennifer Swingle [1 ]
Nau, Cherie B. [2 ]
Schornack, Muriel [2 ]
Shorter, Ellen [3 ]
Nau, Amy [4 ]
Harthan, Jennifer S. [5 ]
机构
[1] Ohio State Univ, Coll Optometry, Columbus, OH 43210 USA
[2] Mayo Clin, Dept Ophthalmol, Rochester, MN USA
[3] Univ Illinois, Dept Ophthalmol & Visual Sci, Chicago, IL USA
[4] Korb & Associates, Boston, MA USA
[5] Illinois Coll Optometry, Chicago, IL USA
关键词
GOLDMANN APPLANATION TONOMETRY; CONTACT-LENSES; TGDC-01;
D O I
10.1097/OPX.0000000000001574
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
SIGNIFICANCE: As scleral lens wear becomes more common, understanding the impact of these lenses upon ocular physiology is critically important. Studies on the effect of scleral lens wear upon intraocular pressure (IOP) have used different instruments and have reported conflicting results. PURPOSE: The purpose of this study was to compare assessment of IOP during scleral lens wear using pneumatonometry and transpalpebral tonometry. METHODS: Twenty healthy subjects wore a small-diameter (15.2 mm) and a large-diameter (18.0 mm) scleral lens on the right eye, each for 1 hour in randomized order. IOP was assessed with pneumatonometry and transpalpebral tonometry on both eyes before lens application, immediately after lens application, after 1 hour of lens wear, and immediately after lens removal. Paired t test compared mean IOP in the study eye to the control eye. Repeated-measures ANOVA was performed to take instrumentation, lens diameter, and their interaction into account in an analysis of the change in IOP in the study eye. RESULTS: Mean peripheral IOP measured with pneumatonometry was not significantly different from baseline at any subsequentmeasurement. Measurements with transpalpebral tonometry, however, were significantly different during scleral lens wear immediately after application and after 1 hour of wear with both diameter lenses (P < .005), but were not significantly different after either sized lens was removed. Repeated-measures ANOVA revealed that the instrument used to measure IOP was a significant factor in IOP changes found during lens wear (P = .001). CONCLUSIONS: Assessment of IOP during scleral lens wear varies based upon the instrument that is used. Although further studies are clearly needed to further elucidate this issue, clinicians should continue to monitor optic nerve structure and function in scleral lens wearers, as they do in all patients.
引用
收藏
页码:711 / 719
页数:9
相关论文
共 50 条
  • [1] Intraocular pressure during scleral lens wear using pneumatonometry
    Nau, Cherie B.
    Harthan, Jennifer
    Nau, Amy Catherine
    Shorter, Ellen
    Schornack, Muriel
    Fogt, Jennifer Swingle
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2020, 61 (07)
  • [2] Comparison of Corneal and Scleral Intraocular Pressure Measurements Obtained by Pneumatonometry, TonoPen, and iCare Tonometers
    Chen, Rebecca
    He, Christina
    Levy, Richard
    Oltra, Erica
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2019, 60 (09)
  • [3] Correlation Between Serial Scleral and Corneal Intraocular Pressure Measurements by Pneumatonometry
    Kuo, Debbie S.
    Ou, Yvonne
    Jeng, Bennie H.
    Bhisitkul, Robert B.
    Stewart, Jay M.
    Duncan, Jacque L.
    Han, Ying
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2014, 55 (13)
  • [4] Effect of Scleral Lens wear on Intraocular Pressure
    Aitsebaomo, A. Philip
    Wong-Powell, Jeannette
    Miller, William
    Amir, Farshid
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2018, 59 (09)
  • [5] Scleral Tonometry Precision During Scleral Lens Wear: A Pilot Study
    Ganjei, Allen Y.
    Shlager, Gabriel G. L.
    Brocks, Daniel
    CLINICAL OPTOMETRY, 2021, 13 : 341 - 350
  • [6] Comparison of Scleral Tono-Pen Intraocular Pressure Measurements with Goldmann Applanation Tonometry
    Badakere, Swathi Vallabh
    Choudhari, Nikhil S.
    Rao, Harsha L.
    Chary, C. Raghava
    Garudadri, Chandra Sekhar
    Senthil, Sirisha
    OPTOMETRY AND VISION SCIENCE, 2018, 95 (02) : 129 - 135
  • [7] Intraocular Pressure Changes During One Week of Scleral Contact Lens Wear
    Alshamrani, Amane
    Yudcovitch, Lorne
    Caroline, Patrick
    Chase, Chris
    Alanazi, Muteb
    Lampa, Matthew
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2024, 65 (07)
  • [8] Comparison of intraocular pressure measurements between Easyton transpalpebral tonometry and Perkins, iCare iC100 and Corvis ST, and the influence of corneal and anterior scleral thickness
    Salazar-Quinones, Liseth
    Fernandez-Vigo, Jose Ignacio
    Perez-Quinones, Yadhira
    Montolio-Marzo, Elena
    Garcia-Bella, Javier
    Morales-Fernandez, Laura
    Sanchez-del-Hoyo, Rafael
    Garcia-Feijoo, Julian
    Martinez-de-la-Casa, Jose Maria
    INTERNATIONAL OPHTHALMOLOGY, 2023, 43 (11) : 4121 - 4129
  • [9] Comparison of intraocular pressure measurements between Easyton transpalpebral tonometry and Perkins, iCare iC100 and Corvis ST, and the influence of corneal and anterior scleral thickness
    Liseth Salazar-Quiñones
    José Ignacio Fernández-Vigo
    Yadhira Pérez-Quiñones
    Elena Montolío-Marzo
    Javier García-Bella
    Laura Morales-Fernandez
    Rafael Sánchez-del-Hoyo
    Julián García-Feijóo
    José María Martínez-de-la-Casa
    International Ophthalmology, 2023, 43 : 4121 - 4129
  • [10] Anatomical and physiological considerations in scleral lens wear: Intraocular pressure
    Schornack, Muriel M.
    Vincent, Stephen J.
    Walker, Maria K.
    CONTACT LENS & ANTERIOR EYE, 2023, 46 (01):