Comparison of Pneumatonometry and Transpalpebral Tonometry Measurements of Intraocular Pressure during Scleral Lens Wear

被引:13
|
作者
Fogt, Jennifer Swingle [1 ]
Nau, Cherie B. [2 ]
Schornack, Muriel [2 ]
Shorter, Ellen [3 ]
Nau, Amy [4 ]
Harthan, Jennifer S. [5 ]
机构
[1] Ohio State Univ, Coll Optometry, Columbus, OH 43210 USA
[2] Mayo Clin, Dept Ophthalmol, Rochester, MN USA
[3] Univ Illinois, Dept Ophthalmol & Visual Sci, Chicago, IL USA
[4] Korb & Associates, Boston, MA USA
[5] Illinois Coll Optometry, Chicago, IL USA
关键词
GOLDMANN APPLANATION TONOMETRY; CONTACT-LENSES; TGDC-01;
D O I
10.1097/OPX.0000000000001574
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
SIGNIFICANCE: As scleral lens wear becomes more common, understanding the impact of these lenses upon ocular physiology is critically important. Studies on the effect of scleral lens wear upon intraocular pressure (IOP) have used different instruments and have reported conflicting results. PURPOSE: The purpose of this study was to compare assessment of IOP during scleral lens wear using pneumatonometry and transpalpebral tonometry. METHODS: Twenty healthy subjects wore a small-diameter (15.2 mm) and a large-diameter (18.0 mm) scleral lens on the right eye, each for 1 hour in randomized order. IOP was assessed with pneumatonometry and transpalpebral tonometry on both eyes before lens application, immediately after lens application, after 1 hour of lens wear, and immediately after lens removal. Paired t test compared mean IOP in the study eye to the control eye. Repeated-measures ANOVA was performed to take instrumentation, lens diameter, and their interaction into account in an analysis of the change in IOP in the study eye. RESULTS: Mean peripheral IOP measured with pneumatonometry was not significantly different from baseline at any subsequentmeasurement. Measurements with transpalpebral tonometry, however, were significantly different during scleral lens wear immediately after application and after 1 hour of wear with both diameter lenses (P < .005), but were not significantly different after either sized lens was removed. Repeated-measures ANOVA revealed that the instrument used to measure IOP was a significant factor in IOP changes found during lens wear (P = .001). CONCLUSIONS: Assessment of IOP during scleral lens wear varies based upon the instrument that is used. Although further studies are clearly needed to further elucidate this issue, clinicians should continue to monitor optic nerve structure and function in scleral lens wearers, as they do in all patients.
引用
收藏
页码:711 / 719
页数:9
相关论文
共 50 条
  • [21] Agreement of rebound and applanation tonometry intraocular pressure measurements during atmospheric pressure change
    Vercellin, Alice Verticchio
    Harris, Alon
    Siesky, Brent
    Zukerman, Ryan
    Tanga, Lucia
    Carnevale, Carmela
    Scarinci, Fabio
    Oddone, Francesco
    PLOS ONE, 2021, 16 (10):
  • [22] Effect of Short-term Scleral Lens Wear of Small and Large Diameter on Intraocular Pressure
    Fogt, Jennifer Swingle
    Harthan, Jennifer
    Nau, Amy Catherine
    Shorter, Ellen
    Schornack, Muriel
    Nau, Cherie B.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2020, 61 (07)
  • [23] A comparison of intraocular pressure measurements with the XPERT noncontact tonometer and Goldmann applanation tonometry
    Popovich, KS
    Shields, MB
    JOURNAL OF GLAUCOMA, 1997, 6 (01) : 44 - 46
  • [24] Changes in Scleral Tonometry and Anterior Chamber Angle after Short-term Scleral Lens Wear
    Obinwanne, Chukwuemeka Junior
    Echendu, Damian C.
    Agbonlahor, Osazee
    Dike, Stella
    OPTOMETRY AND VISION SCIENCE, 2020, 97 (09) : 720 - 725
  • [25] Transpalpebral measurement of intraocular pressure using the Diaton tonometer versus standard Goldmann applanation tonometry
    Yuehua Li
    Jingming Shi
    Xuanchu Duan
    Fang Fan
    Graefe's Archive for Clinical and Experimental Ophthalmology, 2010, 248 : 1765 - 1770
  • [26] Transpalpebral measurement of intraocular pressure using the Diaton tonometer versus standard Goldmann applanation tonometry
    Li, Yuehua
    Shi, Jingming
    Duan, Xuanchu
    Fan, Fang
    GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2010, 248 (12) : 1765 - 1770
  • [27] Intraocular Pressures Changes at the Central Cornea and Sclera during 5 hours of Scleral Lens Wear
    Tse, Vivien
    Litvin, Taras
    Chung, Lisa
    Zhou, Yixiu
    Han, Ying
    Lin, Meng C.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2022, 63 (07)
  • [28] Measurements of Intraocular Pressure by Goldmann Tonometry, Tonopen XL, and the Transpalpebral Tonometer, TGDc-01. After Penetrating Keratoplasty: A Comparativye Study
    Shemesh, Gabi
    Waisbourd, Michael
    Varssano, David
    Michaeli, Adi
    Lazar, Moshe
    Kurtz, Shimon
    CORNEA, 2009, 28 (07) : 724 - 727
  • [29] Comparison of three types of scleral fixation of the intraocular lens
    Byun, Zeeyoon
    Hwang, Sungsoon
    Kong, Mingui
    INDIAN JOURNAL OF OPHTHALMOLOGY, 2025, 73 (03) : 422 - 428
  • [30] Comparison of three types of scleral fixation of intraocular lens
    Byun, Zeeyoon
    Kong, Mingui
    ACTA OPHTHALMOLOGICA, 2022, 100