Intraocular Pressure Measurement with Pneumatonometry and a Tonometer Tip Cover During Negative Pressure Application

被引:2
|
作者
Brambilla, Enrico [1 ]
Ferguson, Tanner J. [2 ]
Chu, Nathan [1 ]
Ammar, David [3 ]
Yoo, Paul [1 ]
机构
[1] Equinox Ophthalm Inc, Newport Beach, CA USA
[2] Vance Thompson Vis, 3101 W 57th St, Sioux Falls, SD 57108 USA
[3] Lions Eye Inst Transplant & Res, Tampa, FL USA
来源
CLINICAL OPHTHALMOLOGY | 2022年 / 16卷
关键词
tonometry; intraocular pressure; TOP; multi-pressure dial; MPD; ophthalmic techniques; diagnostics; GLAUCOMA PROGRESSION;
D O I
10.2147/OPTH.S359605
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: This is a 2-part study to investigate the agreement between pneumatonometry and direct pressure transducer intraocular pressure ( IOP) measurements in a perfusion organ culture (POC) model where (1) the perfusion fluid column is open to atmospheric pressure, holding IOP constant to permit evaluation of the impact of negative pressure (NP) on IOP measurements, and (2) the perfusion fluid column is a closed system, allowing IOP to vary with NP application. Methods: The first part incorporated a fluid column open to atmospheric pressure, maintaining IOP constant to permit evaluation of the effect of applied NP on IOP measurement accuracy. In the second part, the POC column was closed, allowing IOP to vary with NP application and permit evaluation of agreement between pneumatonometry and pressure transducer measurements. In each part, four perfused tissues were used in thirteen paired pre-set IOP (10, 20, 25, 30 mmHg) and NP (0, 5, 10, 15, 20 mmHg) combinations, resulting in a total of 1040 paired measurements (520 per study). The difference in IOP measurements (Delta IOP = Excursion tonometry - pressure transducer) was calculated at each paired configuration. Results: During the first part, the mean Delta IOP was -0.7 +/- 1.6 mmHg across all measurements. During the second part, the mean Delta IOP across all measurements was +0.7 +/- 1.4 mmHg. At NP settings of -5, -10, -15, and -20 mmHg, across all pre-set IOPs, the mean TOP reduction via Excursion tonometry was 3.1 +/- 0.3, 5.6 +/- 1.3, 8.5 +/- 1.7 and 11.2 +/- 1.8 mmHg, respectively. Conclusion: Measurement of IOP via Excursion tonometry yields results within the accuracy range of the pneumatonometry device (per manufacturer) and is minimally impacted by NP application. The IOP-lowering results are consistent with previous studies and further support the effectiveness of the Multi-Pressure Dial in lowering IOP relative to atmospheric pressure.
引用
收藏
页码:1289 / 1300
页数:12
相关论文
共 50 条
  • [31] RELIABILITY OF INTRAOCULAR-PRESSURE MEASUREMENT WITH THE GOLDMANN APPLANATION TONOMETER IN EPIDEMIOLOGIC STUDIES
    DIELEMANS, I
    VINGERLING, JR
    HOFMAN, A
    GROBBEE, DE
    DEJONG, PTVM
    GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 1994, 232 (03) : 141 - 144
  • [32] Effect of manual upper eyelid elevation on intraocular pressure measurement by noncontact tonometer
    Gun, Raziye Donmez
    Simsek, Saban
    INTERNATIONAL OPHTHALMOLOGY, 2022, 42 (09) : 2915 - 2923
  • [33] Evaluation of the Perkins® handheld applanation tonometer in the measurement of intraocular pressure in dogs and cats
    Andrade, Silvia Franco
    Cremonezi, Tatiana
    Miranda Zachi, Cristiane Aparecida
    Lonchiati, Caroline Ferreira
    Amatuzzi, Juliana Dalarrosa
    Sakamoto, Keila Priscilla
    de Arruda Mello, Paulo Augusto
    VETERINARY OPHTHALMOLOGY, 2009, 12 (05) : 277 - 284
  • [34] STUDIES OF INTRAOCULAR PRESSURE WITH GOLDMANS APPLANATION TONOMETER
    BECHRAKIS, E
    OPHTHALMOLOGICA, 1970, 161 (2-3) : 353 - +
  • [35] Effect of manual upper eyelid elevation on intraocular pressure measurement by noncontact tonometer
    Raziye Dönmez Gün
    Şaban Şimşek
    International Ophthalmology, 2022, 42 : 2915 - 2923
  • [36] Measurement of intraocular pressure in pig's eyes using a new tonometer prototype
    Banobre, A
    Alvarez, TL
    Fechtner, RD
    Greene, RJ
    Thomas, GA
    Levi, O
    Ciampa, N
    2005 IEEE 31ST ANNUAL NORTHEAST BIOENGINEERING CONFERENCE, 2005, : 260 - 261
  • [37] Fluctuation of Intraocular Pressure with Noncontact Tonometer and Rebound Tonometer in the Normal Volunteers
    Tatara, Shunya
    Maeda, Fumiatsu
    Tsukahara, Yoshinosuke
    Yaoeda, Kiyoshi
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2021, 62 (08)
  • [38] Correlation Between Serial Scleral and Corneal Intraocular Pressure Measurements by Pneumatonometry
    Kuo, Debbie S.
    Ou, Yvonne
    Jeng, Bennie H.
    Bhisitkul, Robert B.
    Stewart, Jay M.
    Duncan, Jacque L.
    Han, Ying
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2014, 55 (13)
  • [39] Reproducibility of intraocular pressure self-measurement by ICare Home rebound tonometer and comparison with Goldman applanation tonometer
    Gaillard, R.
    Fenolland, J. R.
    Bonnel, S.
    Rosenberg, R.
    Marill, A. F.
    Bostanci, K.
    Giraud, J. M.
    Renard, J. P.
    ACTA OPHTHALMOLOGICA, 2015, 93