Continuous ASL (CASL) perfusion MRI with an array coil and parallel imaging at 3T

被引:67
|
作者
Wang, Z
Wang, JJ
Connick, TJ
Wetmore, GS
Detre, JA
机构
[1] Univ Penn, Sch Med, Ctr Funct Neuroimaging, Dept Neurol, Philadelphia, PA 19104 USA
[2] Univ Penn, Sch Med, Dept Radiol, Philadelphia, PA 19104 USA
关键词
ASL perfusion imaging; parallel imaging; GRAPPA; CBF; SNR;
D O I
10.1002/mrm.20574
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The purpose of this work was to assess the feasibility and efficacy of using an array coil and parallel imaging in continuous arterial spin labeling (CASL) perfusion MRI. An 8-channel receive-only array head coil was used in conjunction with a surrounding detunable volume transmit coil. The signal to noise ratio (SNR), temporal stability, cerebral blood flow (CBF), and perfusion image coverage were measured from steady state CASL scans using: a standard volume coil, array coil, and array coil with 2- and 3-fold accelerated parallel imaging. Compared to the standard volume coil, the array coil provided 3 times the average SNR increase and higher temporal stability for the perfusion weighted images, even with threefold acceleration. Although perfusion images of the array coil were affected by the inhomogeneous coil sensitivities, this effect was invisible in the quantitative CBF images, which showed highly reproducible perfusion values compared to the standard volume coil. The unfolding distortions of parallel imaging were suppressed in the perfusion images by pairwise subtraction, though they sharply degraded the raw EPI images. Moreover, parallel imaging provided the potential of acquiring more slices due to the shortened acquisition time and improved coverage in brain regions with high static field inhomogeneity. Such results highlight the potential utility of array coils and parallel imaging in ASL perfusion MRI.
引用
收藏
页码:732 / 737
页数:6
相关论文
共 50 条
  • [1] A 64-channel 3T array coil for accelerated brain MRI
    Keil, Boris
    Blau, James N.
    Biber, Stephan
    Hoecht, Philipp
    Tountcheva, Veneta
    Setsompop, Kawin
    Triantafyllou, Christina
    Wald, Lawrence L.
    MAGNETIC RESONANCE IN MEDICINE, 2013, 70 (01) : 248 - 258
  • [2] Prostate MR imaging at 3T with a longitudinal array endorectal surface coil and phased array body coil
    Kim, David Y.
    Schnall, Mitchell D.
    Rosen, Mark A.
    Connick, Thomas
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2008, 27 (06) : 1327 - 1330
  • [3] A Novel Flexible Electrotextile 3T MRI RF Coil Array for Carotid Artery Imaging: Design, Characterization, and Prototyping
    Zhang, Daisong
    Rahmat-Samu, Yahya
    IEEE TRANSACTIONS ON ANTENNAS AND PROPAGATION, 2019, 67 (08) : 5115 - 5125
  • [4] Flexible RF Coil Array System Utilizing Electro-textiles for 3T MRI Carotid Artery Imaging
    Zhang, Daisong
    Rahmat-Samii, Yahya
    2017 IEEE INTERNATIONAL SYMPOSIUM ON ANTENNAS AND PROPAGATION & USNC/URSI NATIONAL RADIO SCIENCE MEETING, 2017, : 1015 - 1016
  • [5] Prostate MRI using an external phased array wearable pelvic coil at 3T: comparison with an endorectal coil
    O'Donohoe, Rory L.
    Dunne, Ruth M.
    Kimbrell, Vera
    Tempany, Clare M.
    ABDOMINAL RADIOLOGY, 2019, 44 (03) : 1062 - 1069
  • [6] Prostate MRI using an external phased array wearable pelvic coil at 3T: comparison with an endorectal coil
    Rory L. O’Donohoe
    Ruth M. Dunne
    Vera Kimbrell
    Clare M. Tempany
    Abdominal Radiology, 2019, 44 : 1062 - 1069
  • [7] Whole-brain 3D perfusion MRI at 3.0 T using CASL with a separate labeling coil
    Talagala, SL
    Ye, FQ
    Ledden, PJ
    Chesnick, S
    MAGNETIC RESONANCE IN MEDICINE, 2004, 52 (01) : 131 - 140
  • [8] QUANTITATIVE PERFUSION MEASUREMENTS IN RENAL MASSES WITH ASL AND DCE MRI AT 3T CORRELATE WITH MICROVASCULAR DENSITY AT HISTOPATHOLOGY
    Zhang, Yue
    Kapur, Payal
    Yuan, Qing
    Madhuranthakam, Ananth
    Carvo, Ingrid
    Signoretti, Sabina
    Dimitrov, Ivan
    Xi, Yin
    Wicks, Katherine
    Margulis, Vitaly
    Cadeddu, Jeffrey
    Lay, Aaron
    Brugarolas, James
    Pedrosa, Ivan
    JOURNAL OF UROLOGY, 2015, 193 (04): : E124 - E124
  • [9] Design of 8 channel insertional pTx array coil for 3T body imaging
    Kim, Young Beom
    Ryu, Yeunchul
    Transactions of the Korean Institute of Electrical Engineers, 2014, 63 (04): : 546 - 550
  • [10] Automatic tuned MRI RF coil for multinuclear imaging of small animals at 3T
    Muftuler, LT
    Gulsen, G
    Sezen, KD
    Nalcioglu, O