Continuous ASL (CASL) perfusion MRI with an array coil and parallel imaging at 3T

被引:67
|
作者
Wang, Z
Wang, JJ
Connick, TJ
Wetmore, GS
Detre, JA
机构
[1] Univ Penn, Sch Med, Ctr Funct Neuroimaging, Dept Neurol, Philadelphia, PA 19104 USA
[2] Univ Penn, Sch Med, Dept Radiol, Philadelphia, PA 19104 USA
关键词
ASL perfusion imaging; parallel imaging; GRAPPA; CBF; SNR;
D O I
10.1002/mrm.20574
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The purpose of this work was to assess the feasibility and efficacy of using an array coil and parallel imaging in continuous arterial spin labeling (CASL) perfusion MRI. An 8-channel receive-only array head coil was used in conjunction with a surrounding detunable volume transmit coil. The signal to noise ratio (SNR), temporal stability, cerebral blood flow (CBF), and perfusion image coverage were measured from steady state CASL scans using: a standard volume coil, array coil, and array coil with 2- and 3-fold accelerated parallel imaging. Compared to the standard volume coil, the array coil provided 3 times the average SNR increase and higher temporal stability for the perfusion weighted images, even with threefold acceleration. Although perfusion images of the array coil were affected by the inhomogeneous coil sensitivities, this effect was invisible in the quantitative CBF images, which showed highly reproducible perfusion values compared to the standard volume coil. The unfolding distortions of parallel imaging were suppressed in the perfusion images by pairwise subtraction, though they sharply degraded the raw EPI images. Moreover, parallel imaging provided the potential of acquiring more slices due to the shortened acquisition time and improved coverage in brain regions with high static field inhomogeneity. Such results highlight the potential utility of array coils and parallel imaging in ASL perfusion MRI.
引用
收藏
页码:732 / 737
页数:6
相关论文
共 50 条
  • [21] A Passive and Conformal Magnetic Metasurface for 3T MRI Birdcage Coil
    Rotundo, Sabrina
    Brizi, Danilo
    Carluccio, Giuseppe
    Lakshmanan, Karthik
    Collins, Christopher M.
    Lattanzi, Riccardo
    Monorchio, Agostino
    2023 INTERNATIONAL CONFERENCE ON ELECTROMAGNETICS IN ADVANCED APPLICATIONS, ICEAA, 2023, : 222 - 224
  • [22] Eight-Channel Phased Array RF Coils Design for 3T Parallel MRI System
    Basari
    Rahardjo, Eko Tjipto
    Zulkifli, Fitri Yuli
    2015 IEEE 4TH ASIA-PACIFIC CONFERENCE ON ANTENNAS AND PROPAGATION (APCAP), 2015, : 56 - 57
  • [23] PERFUSION IMAGING OF FLOW EXPERIENCES WITH CONTINUOUS ARTERIAL SPIN LABELING (CASL)
    Hoenig, Klaus
    Ulrich, Martin
    Waller, Christiane
    Keller, Johannes
    Groen, Georg
    PSYCHOPHYSIOLOGY, 2013, 50 : S125 - S125
  • [24] Phased array coil for implementing parallel MRI in intravascular imaging: A feasibility study
    Omer, Hammad
    Dickinson, Robert J.
    Awan, Shakil A.
    CONCEPTS IN MAGNETIC RESONANCE PART A, 2014, 43 (06) : 267 - 276
  • [25] An introduction to coil array design for parallel MRI
    Ohliger, Michael A.
    Sodickson, Daniel K.
    NMR IN BIOMEDICINE, 2006, 19 (03) : 300 - 315
  • [26] An eight-channel, nonoverlapping phased array coil with capacitive decoupling for parallel MRI at 3 T
    Von Morze, Cornelius
    Tropp, James
    Banerjee, Suchandrima
    Xu, Duan
    Karpodinis, Kostas
    Carvajal, Lucas
    Hess, Christopher P.
    Mukherjee, Pratik
    Majumdar, Sharmila
    Vigneron, Daniel B.
    CONCEPTS IN MAGNETIC RESONANCE PART B-MAGNETIC RESONANCE ENGINEERING, 2007, 31B (01) : 37 - 43
  • [27] 3T MR imaging of the orbits, optic nerves and chiasm using a hybrid phased-array coil
    Lane, JI
    Kocharian, A
    Witte, RJ
    Felmlee, JP
    Watson, R
    RADIOLOGY, 2002, 225 : 564 - 564
  • [28] Improvements in Carotid Plaque Imaging Using a New Eight-Element Phased Array Coil at 3T
    Balu, Niranjan
    Yarnykh, Vasily L.
    Scholnick, Joshua
    Chu, Baocheng
    Yuan, Chun
    Hayes, Cecil
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2009, 30 (05) : 1209 - 1214
  • [29] 3T Multiparametric MRI (3T mMRI) With a Phased-Array Torso Coil in Staging of Prostate Cancer Patients (PCP) Prior to Radical Prostatectomy: Preliminary Results
    Counago, F.
    Recio, M.
    Del Cerro, E.
    Diaz, A. A.
    Marcos, F. J.
    Murillo, R.
    Rodriguez, J. M.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2014, 90 : S409 - S410
  • [30] To Evaluate the Damage of Renal Function in CIAKI Rats at 3T: Using ASL and BOLD MRI
    Chen, Wen-bo
    Liang, Long
    Zhang, Bin
    Liu, Chun-ling
    Liu, Hong-jun
    Luo, Hai-ying
    Zeng, Qiong-xin
    Liang, Chang-hong
    Liu, Guan-shu
    Zhang, Shui-xing
    BIOMED RESEARCH INTERNATIONAL, 2015, 2015