Sensitivity of Electronic Portal Imaging Device (EPID) Based Transit Dosimetry to Detect Inter-fraction Patient Variations

被引:4
|
作者
Bawazeer, Omemh [1 ,2 ]
Sarasanandarajah, Sivananthan [1 ,3 ,4 ]
Herath, Sisira [3 ]
Kron, Tomas [3 ,5 ]
Deb, Pradip [1 ]
机构
[1] RMIT Univ, Discipline Med Radiat, Melbourne, Vic, Australia
[2] Umm Al Qura Univ, Discipline Sci, Mecca, Saudi Arabia
[3] Peter MacCallum Canc Ctr, Dept Phys Sci, Melbourne, Vic, Australia
[4] IAEA, Div Human Hlth, Vienna, Austria
[5] Univ Melbourne, Sir Peter MacCallum Dept Oncol, Melbourne, Vic, Australia
关键词
Transit EPID dosimetry; Patient variation SSIM index; QUALITY-ASSURANCE; RADIOTHERAPY; INCIDENTS; ERRORS;
D O I
10.1007/978-981-10-9023-3_86
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The sensitivity of EPID-based transit dosimetry to detect patient variations between treatment fractions is examined using gamma analysis and a structural similarity (SSIM) index. EPID images were acquired for 3-dimensional conformal (3DCRT) and dynamic intensity modulated (dIMRT) radiation therapy fields in multiple fractions. Transit images were converted to doses, transit dose in the first fraction considered the reference dose. Variations in patient position or weight were then introduced in the subsequent fractions. Positional variations were examined using a lung and a head and neck phantoms. Anatomical variations were examined using a slab phantom in three scenarios, with solid water simulating tissue, medium-density fiberboard simulating fat, and Styrofoam simulating lung. The dose difference between the first and subsequent fractions was computed using various gamma criteria and the SSIM index. Using a criterion of 3%/3 mm, EPID can detect positional variations >= 4 mm, and tissue and fat variations >= 1 cm, whereas it cannot detect lung variations up to 4 cm. The sensitivity for 3DCRT is higher than for dIMRT. EPID can detect the most variations when using 3%/1 mm. With the SSIM index, EPID can detect a 2 mm positional variation and 1 cm of lung variation. The factor that optimized the sensitivity of EPID was a reduction in the distance to the agreement criteria. Our study introduces the SSIM as an alternative analysis with high sensitivity for minimal variations.
引用
收藏
页码:477 / 480
页数:4
相关论文
共 50 条
  • [21] A Clinical Objective IMRT QA Method Based on Portal Dosimetry and Electronic Portal Imager Device (EPID) Measurement
    Zhang, Miao
    Qin, Songbing
    Chen, Ting
    Kim, Sung
    Jabbour, Salma
    Haffty, Bruce
    Yue, Ning J.
    TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2013, 12 (02) : 145 - 150
  • [22] Electronic portal imaging device (EPID) based on a novel camera with avalanche multiplication
    Pang, G
    Zhao, W
    Luhta, R
    Rowlands, JA
    PHYSICS OF MEDICAL IMAGING, 1998, 3336 : 195 - 201
  • [23] A new approach for the pixel map sensitivity (PMS) evaluation of an electronic portal imaging device (EPID)
    Boriano, Alberto
    Lucio, Francesco
    Calamia, Elisa
    Russi, Elvio
    Marchetto, Flavio
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2013, 14 (06): : 234 - 250
  • [24] A dual two dimensional electronic portal imaging device transit dosimetry model based on an empirical quadratic formalism
    Tan, Y. I.
    Metwaly, M.
    Glegg, M.
    Baggarley, S. P.
    Elliott, A.
    BRITISH JOURNAL OF RADIOLOGY, 2015, 88 (1051):
  • [25] Simple Proposal for Dosimetry with an Elekta iViewGT™ Electronic Portal Imaging Device (EPID) Using Commercial Software Modules
    Liebich, Janett
    Licher, Joerg
    Scherf, Christian
    Kara, Eugen
    Koch, Nadine
    Roedel, Claus
    Ramm, Ulla
    STRAHLENTHERAPIE UND ONKOLOGIE, 2011, 187 (05) : 316 - 321
  • [26] EPIDEEP: Using a Deep Learning Model to Predict In Vivo Electronic Portal Imaging Device (EPID) Transit Images
    Anderson, B. M.
    Moore, L.
    Moore, K. L.
    Bojechko, C.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2023, 117 (02): : E645 - E645
  • [27] Assessing the accuracy of electronic portal imaging device (EPID)-based dosimetry: I. Quantities influencing long-term stability
    Renaud, James
    Muir, Bryan
    MEDICAL PHYSICS, 2022, 49 (02) : 1231 - 1237
  • [28] Relative dosimetry using a scanning liquid-filled ion chamber (SLIC) electronic portal imaging device (EPID)
    Ortega, JFC
    Seoane, A
    Garrido, L
    Casals, J
    RADIOTHERAPY AND ONCOLOGY, 2004, 73 : S415 - S415
  • [29] In vivo dosimetry for prostate cancer patients using an electronic portal imaging device (EPID); demonstration of internal organ motion
    Kroonwijk, M
    Pasma, KL
    Quint, S
    Koper, PCM
    Visser, AG
    Heijmen, BJM
    RADIOTHERAPY AND ONCOLOGY, 1998, 49 (02) : 125 - 132
  • [30] A Monte Carlo calculation model of electronic portal imaging device for transit dosimetry through heterogeneous media
    Yoon, Jihyung
    Jung, Jae Won
    Kim, Jong Oh
    Yeo, Inhwan
    MEDICAL PHYSICS, 2016, 43 (05) : 2242 - 2250