Sensitivity of Electronic Portal Imaging Device (EPID) Based Transit Dosimetry to Detect Inter-fraction Patient Variations

被引:4
|
作者
Bawazeer, Omemh [1 ,2 ]
Sarasanandarajah, Sivananthan [1 ,3 ,4 ]
Herath, Sisira [3 ]
Kron, Tomas [3 ,5 ]
Deb, Pradip [1 ]
机构
[1] RMIT Univ, Discipline Med Radiat, Melbourne, Vic, Australia
[2] Umm Al Qura Univ, Discipline Sci, Mecca, Saudi Arabia
[3] Peter MacCallum Canc Ctr, Dept Phys Sci, Melbourne, Vic, Australia
[4] IAEA, Div Human Hlth, Vienna, Austria
[5] Univ Melbourne, Sir Peter MacCallum Dept Oncol, Melbourne, Vic, Australia
关键词
Transit EPID dosimetry; Patient variation SSIM index; QUALITY-ASSURANCE; RADIOTHERAPY; INCIDENTS; ERRORS;
D O I
10.1007/978-981-10-9023-3_86
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The sensitivity of EPID-based transit dosimetry to detect patient variations between treatment fractions is examined using gamma analysis and a structural similarity (SSIM) index. EPID images were acquired for 3-dimensional conformal (3DCRT) and dynamic intensity modulated (dIMRT) radiation therapy fields in multiple fractions. Transit images were converted to doses, transit dose in the first fraction considered the reference dose. Variations in patient position or weight were then introduced in the subsequent fractions. Positional variations were examined using a lung and a head and neck phantoms. Anatomical variations were examined using a slab phantom in three scenarios, with solid water simulating tissue, medium-density fiberboard simulating fat, and Styrofoam simulating lung. The dose difference between the first and subsequent fractions was computed using various gamma criteria and the SSIM index. Using a criterion of 3%/3 mm, EPID can detect positional variations >= 4 mm, and tissue and fat variations >= 1 cm, whereas it cannot detect lung variations up to 4 cm. The sensitivity for 3DCRT is higher than for dIMRT. EPID can detect the most variations when using 3%/1 mm. With the SSIM index, EPID can detect a 2 mm positional variation and 1 cm of lung variation. The factor that optimized the sensitivity of EPID was a reduction in the distance to the agreement criteria. Our study introduces the SSIM as an alternative analysis with high sensitivity for minimal variations.
引用
下载
收藏
页码:477 / 480
页数:4
相关论文
共 50 条
  • [31] Electronic portal cine imaging reveals only minimal intra- and inter-fraction motion intangential breast irradiation
    Reist, M
    Vetterli, D
    Greiner, RH
    Gruber, G
    STRAHLENTHERAPIE UND ONKOLOGIE, 2005, 181 : 139 - 139
  • [32] Feasibility study of patient positioning verification in electron beam radiotherapy with an electronic portal imaging device (EPID)
    Ramm, U.
    Koehn, J.
    Dominguez, R. Rodriguez
    Licher, J.
    Koch, N.
    Kara, E.
    Scherf, C.
    Roedel, C.
    Weiss, C.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2014, 30 (02): : 215 - 220
  • [33] Use of in vivo transit portal images to detect gross inter-fraction patient geometry changes on an O-ring type linear accelerator for pelvis and head/neck patients
    Aland, Trent
    Jarema, Talia
    Spalding, Myles
    Kairn, Tanya
    Trapp, Jamie
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2021, 7 (06):
  • [34] Simple Proposal for Dosimetry with an Elekta iViewGTTM Electronic Portal Imaging Device (EPID) Using Commercial Software ModulesEinfacher Vorschlag für die Dosimetrie mit einem Elekta iViewGTTM Electronic Portal Imaging Device (EPID) mit kommerziellen Softwaremodulen
    Janett Liebich
    Jörg Licher
    Christian Scherf
    Eugen Kara
    Nadine Koch
    Claus Rödel
    Ulla Ramm
    Strahlentherapie und Onkologie, 2011, 187 (5)
  • [35] Fast transit portal dosimetry using density-scaled layer modeling of aSi-based electronic portal imaging device and Monte Carlo method
    Jung, Jae Won
    Kim, Jong Oh
    Yeo, Inhwan Jason
    Cho, Young-Bin
    Kim, Sun Mo
    DiBiase, Steven
    MEDICAL PHYSICS, 2012, 39 (12) : 7593 - 7602
  • [36] Assessing the accuracy of electronic portal imaging device (EPID)-based dosimetry: II. Evaluation of a dosimetric uncertainty budget and development of a new film-in-EPID absorbed dose calibration methodology
    Renaud, James
    Muir, Bryan
    MEDICAL PHYSICS, 2022, 49 (02) : 1238 - 1247
  • [37] Evaluation of patient set-up variation in isocentric breast radiotherapy using an electronic portal imaging device (EPID)
    Petillion, S.
    Swinnen, A.
    Van den Heuvel, F.
    RADIOTHERAPY AND ONCOLOGY, 2006, 81 : S205 - S205
  • [38] Comparison of Patient-Specific IMRT QA with the Integral Quality Monitor (IQM) and the Electronic Portal Imaging Device (EPID)
    Ghafarian, M.
    Morales-Paliza, M.
    Price, M.
    MEDICAL PHYSICS, 2019, 46 (06) : E612 - E612
  • [39] Initial Clinical Experience Performing Patient Treatment Verification With an Electronic Portal Imaging Device Transit Dosimeter
    Berry, Sean L.
    Polvorosa, Cynthia
    Cheng, Simon
    Deutsch, Israel
    Chao, K. S. Clifford
    Wuu, Cheng-Shie
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2014, 88 (01): : 204 - 209
  • [40] Commissioning and Evaluation of an Electronic Portal Imaging Device-Based In-Vivo Dosimetry Software
    Held, Mareike
    Cheung, Joey
    Andujar, Angelica Perez
    Husson, Francois
    Morin, Olivier
    CUREUS, 2018, 10 (02):