Sensitivity of Electronic Portal Imaging Device (EPID) Based Transit Dosimetry to Detect Inter-fraction Patient Variations

被引:4
|
作者
Bawazeer, Omemh [1 ,2 ]
Sarasanandarajah, Sivananthan [1 ,3 ,4 ]
Herath, Sisira [3 ]
Kron, Tomas [3 ,5 ]
Deb, Pradip [1 ]
机构
[1] RMIT Univ, Discipline Med Radiat, Melbourne, Vic, Australia
[2] Umm Al Qura Univ, Discipline Sci, Mecca, Saudi Arabia
[3] Peter MacCallum Canc Ctr, Dept Phys Sci, Melbourne, Vic, Australia
[4] IAEA, Div Human Hlth, Vienna, Austria
[5] Univ Melbourne, Sir Peter MacCallum Dept Oncol, Melbourne, Vic, Australia
关键词
Transit EPID dosimetry; Patient variation SSIM index; QUALITY-ASSURANCE; RADIOTHERAPY; INCIDENTS; ERRORS;
D O I
10.1007/978-981-10-9023-3_86
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The sensitivity of EPID-based transit dosimetry to detect patient variations between treatment fractions is examined using gamma analysis and a structural similarity (SSIM) index. EPID images were acquired for 3-dimensional conformal (3DCRT) and dynamic intensity modulated (dIMRT) radiation therapy fields in multiple fractions. Transit images were converted to doses, transit dose in the first fraction considered the reference dose. Variations in patient position or weight were then introduced in the subsequent fractions. Positional variations were examined using a lung and a head and neck phantoms. Anatomical variations were examined using a slab phantom in three scenarios, with solid water simulating tissue, medium-density fiberboard simulating fat, and Styrofoam simulating lung. The dose difference between the first and subsequent fractions was computed using various gamma criteria and the SSIM index. Using a criterion of 3%/3 mm, EPID can detect positional variations >= 4 mm, and tissue and fat variations >= 1 cm, whereas it cannot detect lung variations up to 4 cm. The sensitivity for 3DCRT is higher than for dIMRT. EPID can detect the most variations when using 3%/1 mm. With the SSIM index, EPID can detect a 2 mm positional variation and 1 cm of lung variation. The factor that optimized the sensitivity of EPID was a reduction in the distance to the agreement criteria. Our study introduces the SSIM as an alternative analysis with high sensitivity for minimal variations.
引用
下载
收藏
页码:477 / 480
页数:4
相关论文
共 50 条
  • [1] Transit dosimetry with an electronic portal imaging device (EPID) for 115 prostate cancer patients
    Pasma, KL
    Kroonwijk, M
    Quint, S
    Visser, AG
    Heijmen, BJM
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1999, 45 (05): : 1297 - 1303
  • [2] Transit Dosimetry for Verification of Dose Delivery Using Electronic Portal Imaging Device (EPID)
    Baek, T.
    Chung, E.
    Lee, S.
    Yoon, M.
    MEDICAL PHYSICS, 2014, 41 (06) : 301 - 301
  • [3] An electronic portal imaging device for transit dosimetry
    SymondsTayler, JRN
    Partridge, M
    Evans, PM
    PHYSICS IN MEDICINE AND BIOLOGY, 1997, 42 (11): : 2273 - 2283
  • [4] Using Varian Portal Dosimetry to Detect Inter-Fraction Patient Separation Changes On a Varian Halcyon Linear Accelerator
    Aland, T.
    Spalding, M.
    Jarema, T.
    MEDICAL PHYSICS, 2020, 47 (06) : E874 - E875
  • [5] Evaluation of an electronic portal imaging device for transit dosimetry
    He, XD
    Van Esch, A
    Reymen, R
    Huyskens, D
    ACTA ONCOLOGICA, 1999, 38 (05) : 591 - 596
  • [6] Evaluation of an electronic portal imaging device for transit dosimetry
    He, Xiaodong
    Esch, Ann Van
    Reymen, Rita
    Huyskens, Dominique
    Acta Oncologica, 38 (05): : 591 - 596
  • [7] Using Varian Portal Dosimetry to Detect Inter-Fraction Patient Geometry Changes On a Varian Halcyon Linear Accelerator
    Aland, T.
    Jarema, T.
    MEDICAL PHYSICS, 2018, 45 (06) : E182 - E182
  • [8] Radiotherapy delivery error detection with electronic portal imaging device (EPID)-based in vivo dosimetry
    Yang Xueying
    Geng Lisheng
    Huang David
    Li Kaiwen
    Zhuang Hongqing
    Cai Jing
    Yang Ruijie
    中华医学杂志(英文版), 2023, 136 (08)
  • [9] Radiotherapy delivery error detection with electronic portal imaging device (EPID)-based in vivo dosimetry
    Yang, Xueying
    Geng, Lisheng
    Huang, David
    Li, Kaiwen
    Zhuang, Hongqing
    Cai, Jing
    Yang, Ruijie
    CHINESE MEDICAL JOURNAL, 2023, 136 (08) : 998 - 1000
  • [10] SIMPLE PROPOSAL FOR DOSIMETRY WITH AN ELEKTA IVIEWGTTM ELECTRONIC PORTAL IMAGING DEVICE (EPID)
    Liebich, J.
    Licher, J.
    Scherf, C.
    Kara, E.
    Koch, N.
    Roedel, C.
    Ramm, U.
    RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S419 - S419