High-resolution monolithic LYSO detector with 6-layer depth-of-interaction for clinical PET

被引:21
|
作者
Stockhoff, Mariele [1 ]
Decuyper, Milan [1 ]
Van Holen, Roel [1 ]
Vandenberghe, Stefaan [1 ]
机构
[1] Univ Ghent, Dept Elect & Informat Syst, Med Image & Signal Proc MEDISIP, Ghent, Belgium
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2021年 / 66卷 / 15期
关键词
PET; TB-PET; monolithic detector; high-resolution; nearest neighbour; neural networks; CALIBRATION PROCEDURE; POSITION ESTIMATION; SPATIAL-RESOLUTION; NEURAL-NETWORK; TOF-PET; CRYSTALS; DOI;
D O I
10.1088/1361-6560/ac1459
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The system spatial resolution of whole-body positron emission tomography (PET) is limited to around 2 mm due to positron physics and the large diameter of the bore. To stay below this 'physics'-limit a scintillation detector with an intrinsic spatial resolution of around 1.3 mm is needed. Currently used detector technology consists of arrays of 2.6-5 mm segmented scintillator pixels which are the dominant factor contributing to the system resolution. Pixelated detectors using smaller pixels exist but face major drawbacks in sensitivity, timing, energy resolution and cost. Monolithic continuous detectors, where the spatial resolution is determined by the shape of the light distribution on the photodetector array, are a promising alternative. Without having the drawbacks of pixelated detectors, monolithic ones can also provide depth-of-interaction (DOI) information. In this work we present a monolithic detector design aiming to serve high-resolution clinical PET systems while maintaining high sensitivity. A 50 x 50 x 16 mm(3) Lutetium-Yttrium oxyorthosilicate scintillation crystal with silicon photomultiplier (SiPM) backside readout is calibrated in singles mode by a collimated beam obtaining a reference dataset for the event positioning. A mean nearest neighbour (MNN) algorithm and an artificial neural network for positioning are compared. The targeted intrinsic detector resolution of 1.3 mm needed to reach a 2 mm resolution on system level was accomplished with both algorithms. The neural network achieved a mean spatial resolution of 1.14 mm FWHM for the whole detector and 1.02 mm in the centre (30 x 30 mm(2)). The MNN algorithm performed slightly worse with 1.17 mm for the whole detector and 1.13 mm in the centre. The intrinsic DOI information will also result in uniform system spatial resolution over the full field of view.
引用
收藏
页数:15
相关论文
共 50 条
  • [31] Investigation of crystal surface finish and geometry on single LYSO scintillator detector performance for depth-of-interaction measurement with silicon photomultipliers
    Bircher, Chad
    Shao, Yiping
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2012, 693 : 236 - 243
  • [32] Development of a micro PET system with improved spatial resolution through depth-of-interaction measurement
    Correia, P. M. M.
    Castro, I. F. C.
    Veloso, J. F. C. A.
    SECOND INTERNATIONAL CONFERENCE ON APPLICATIONS OF OPTICS AND PHOTONICS, 2014, 9286
  • [33] New depth-of-interaction PET detector using a single-layer crystal array with a stair-shaped reflector arrangement
    Son, Jeong-Whan
    Lee, Min Sun
    Lee, Jae Sung
    JOURNAL OF NUCLEAR MEDICINE, 2016, 57
  • [34] Performance of a high-resolution depth encoding PET detector using barium sulfate reflector
    Kuang, Zhonghua
    Wang, Xiaohui
    Li, Cheng
    Deng, Xinhan
    Feng, Kai
    Hu, Zhanli
    Fu, Xin
    Ren, Ning
    Zhang, Xianming
    Zheng, Yunfei
    Liang, Dong
    Liu, Xin
    Zheng, Hairong
    Yang, Yongfeng
    PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (15): : 5945 - 5958
  • [35] Preclinical Studies Using a Prototype High-Resolution PET System with Depth of Interaction
    Safavi-Naeini, M.
    Franklin, D. R.
    Petasecca, M.
    Lerch, M. L. F.
    Kirkham, R.
    Moorhead, G.
    Dunn, P.
    De Geronimo, G.
    Rosenfeld, A. B.
    2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, : 3305 - 3307
  • [36] Depth of interaction resolution measurements for a high resolution PET detector using position sensitive avalanche photodiodes
    Yang, Yongfeng
    Dokhale, Purushottam A.
    Silverman, Robert W.
    Shah, Kanai S.
    McClish, Mickel A.
    Farrell, Richard
    Entine, Gerald
    Cherry, Simon R.
    PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (09): : 2131 - 2142
  • [37] Development of high-resolution detector module with depth of interaction identification for positron emission tomography
    Niknejad, Tahereh
    Pizzichemi, Marco
    Stringhini, Gianluca
    Auffray, Etiennette
    Bugalho, Ricardo
    Da Silva, Jose Carlos
    Di Francesco, Agostino
    Ferramacho, Luis
    Lecoq, Paul
    Leong, Carlos
    Paganoni, Marco
    Rolo, Manuel
    Silva, Rui
    Silveira, Miguel
    Tavernier, Stefaan
    Varela, Joao
    Zorraquino, Carlos
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2017, 845 : 684 - 688
  • [38] Design of High-Resolution Brain PET Using Dual-Layer DOI Detector
    Lim, W.
    Lee, S.
    Jung, J. H.
    Choi, Y.
    INTERNAL MEDICINE JOURNAL, 2022, 52 : 33 - 34
  • [39] Depth Encoding PET Detector Using Two and Four Reflector Arrangements along Depth of Single Layer LYSO Array
    Zhao B.
    Fu C.
    Kuang Z.
    Wang X.
    Ren N.
    Wu S.
    Yang Q.
    Sun M.
    Zhang C.
    Zhang X.
    Gao J.
    Sang Z.
    Hu Z.
    Du J.
    Yang Y.
    Yuanzineng Kexue Jishu/Atomic Energy Science and Technology, 2020, 54 (03): : 533 - 541
  • [40] Performance evaluation of continuous depth-of-interaction encoding PET detector using a single-ended readout
    Ito, Mikiko
    Lee, Min Sun
    Lee, Jae Sung
    JOURNAL OF NUCLEAR MEDICINE, 2012, 53