High-resolution monolithic LYSO detector with 6-layer depth-of-interaction for clinical PET

被引:21
|
作者
Stockhoff, Mariele [1 ]
Decuyper, Milan [1 ]
Van Holen, Roel [1 ]
Vandenberghe, Stefaan [1 ]
机构
[1] Univ Ghent, Dept Elect & Informat Syst, Med Image & Signal Proc MEDISIP, Ghent, Belgium
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2021年 / 66卷 / 15期
关键词
PET; TB-PET; monolithic detector; high-resolution; nearest neighbour; neural networks; CALIBRATION PROCEDURE; POSITION ESTIMATION; SPATIAL-RESOLUTION; NEURAL-NETWORK; TOF-PET; CRYSTALS; DOI;
D O I
10.1088/1361-6560/ac1459
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The system spatial resolution of whole-body positron emission tomography (PET) is limited to around 2 mm due to positron physics and the large diameter of the bore. To stay below this 'physics'-limit a scintillation detector with an intrinsic spatial resolution of around 1.3 mm is needed. Currently used detector technology consists of arrays of 2.6-5 mm segmented scintillator pixels which are the dominant factor contributing to the system resolution. Pixelated detectors using smaller pixels exist but face major drawbacks in sensitivity, timing, energy resolution and cost. Monolithic continuous detectors, where the spatial resolution is determined by the shape of the light distribution on the photodetector array, are a promising alternative. Without having the drawbacks of pixelated detectors, monolithic ones can also provide depth-of-interaction (DOI) information. In this work we present a monolithic detector design aiming to serve high-resolution clinical PET systems while maintaining high sensitivity. A 50 x 50 x 16 mm(3) Lutetium-Yttrium oxyorthosilicate scintillation crystal with silicon photomultiplier (SiPM) backside readout is calibrated in singles mode by a collimated beam obtaining a reference dataset for the event positioning. A mean nearest neighbour (MNN) algorithm and an artificial neural network for positioning are compared. The targeted intrinsic detector resolution of 1.3 mm needed to reach a 2 mm resolution on system level was accomplished with both algorithms. The neural network achieved a mean spatial resolution of 1.14 mm FWHM for the whole detector and 1.02 mm in the centre (30 x 30 mm(2)). The MNN algorithm performed slightly worse with 1.17 mm for the whole detector and 1.13 mm in the centre. The intrinsic DOI information will also result in uniform system spatial resolution over the full field of view.
引用
收藏
页数:15
相关论文
共 50 条
  • [41] Compact detector modules for high resolution PET Imaging with LYSO and avalanche photodiode arrays
    Clajus, M
    Cajipe, VB
    Calderwood, RF
    Cherry, SR
    Hayakawa, S
    Tümer, TO
    Yossifor, O
    2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7, 2004, : 2890 - 2892
  • [42] Investigation of a high-resolution detector for wholebody PET imaging
    Surti, Suleman
    Kuhn, Austin
    Karp, Joel S.
    2005 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2005, : 1601 - 1605
  • [43] A high-resolution PET detector based on continuous scintillators
    Tavernier, S
    Bruyndonckx, P
    Leonard, S
    Devroede, O
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2005, 537 (1-2): : 321 - 325
  • [44] Resolution improvement in PET using detector depth of interaction information.
    Dahlbom, M
    MacDonald, LR
    Cherry, SR
    Eriksson, L
    Casey, ME
    JOURNAL OF NUCLEAR MEDICINE, 1997, 38 (05) : 162 - 162
  • [45] Computer simulation of four-layer depth of interaction detector for PET
    Sato, M
    Haneishi, H
    Inadama, N
    Murayama, H
    2003 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORD, VOLS 1-5, 2004, : 1913 - 1917
  • [46] Study of two-layer tapered depth of interaction PET detector
    Behnamian, Hadi
    Yousefnejad, Sirous
    Shafiee, Mehdi
    Rafiei, Alireza
    APPLIED RADIATION AND ISOTOPES, 2021, 174
  • [47] Development of dual-layer GSO depth-of-interaction block detector using angled optical fiber
    Okumura, Satoshi
    Yamamoto, Seiichi
    Watabe, Hiroshi
    Kato, Natsuki
    Hamamura, Huka
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2015, 781 : 65 - 70
  • [48] Monolithic PET detectors with sub-mm transverse and 6 layer DOI identification for next generation high resolution clinical PET scanners
    Stockhoff, M.
    Van Holen, R.
    Vandenberghe, S.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2019, 46 (SUPPL 1) : S808 - S808
  • [49] Calibration Methodology for a Dual-ended Readout Silicon Photomultiplier Based Depth-of-interaction PET Detector Module
    Xi, W.
    McKisson, J. E.
    Weisenberger, A. G.
    Lee, S.
    Taylor, M.
    Stepanyan, A.
    Zorn, C.
    2012 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE RECORD (NSS/MIC), 2012, : 3493 - 3495
  • [50] Lower-cost Depth-of-Interaction PET Detector Designs Using Dichotomous-3D Decoding
    Zhang, Yuxuan
    Yan, Han
    Baghaei, Hossain
    Wong, Wai-Hoi
    2015 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2015,