High-resolution monolithic LYSO detector with 6-layer depth-of-interaction for clinical PET

被引:21
|
作者
Stockhoff, Mariele [1 ]
Decuyper, Milan [1 ]
Van Holen, Roel [1 ]
Vandenberghe, Stefaan [1 ]
机构
[1] Univ Ghent, Dept Elect & Informat Syst, Med Image & Signal Proc MEDISIP, Ghent, Belgium
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2021年 / 66卷 / 15期
关键词
PET; TB-PET; monolithic detector; high-resolution; nearest neighbour; neural networks; CALIBRATION PROCEDURE; POSITION ESTIMATION; SPATIAL-RESOLUTION; NEURAL-NETWORK; TOF-PET; CRYSTALS; DOI;
D O I
10.1088/1361-6560/ac1459
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The system spatial resolution of whole-body positron emission tomography (PET) is limited to around 2 mm due to positron physics and the large diameter of the bore. To stay below this 'physics'-limit a scintillation detector with an intrinsic spatial resolution of around 1.3 mm is needed. Currently used detector technology consists of arrays of 2.6-5 mm segmented scintillator pixels which are the dominant factor contributing to the system resolution. Pixelated detectors using smaller pixels exist but face major drawbacks in sensitivity, timing, energy resolution and cost. Monolithic continuous detectors, where the spatial resolution is determined by the shape of the light distribution on the photodetector array, are a promising alternative. Without having the drawbacks of pixelated detectors, monolithic ones can also provide depth-of-interaction (DOI) information. In this work we present a monolithic detector design aiming to serve high-resolution clinical PET systems while maintaining high sensitivity. A 50 x 50 x 16 mm(3) Lutetium-Yttrium oxyorthosilicate scintillation crystal with silicon photomultiplier (SiPM) backside readout is calibrated in singles mode by a collimated beam obtaining a reference dataset for the event positioning. A mean nearest neighbour (MNN) algorithm and an artificial neural network for positioning are compared. The targeted intrinsic detector resolution of 1.3 mm needed to reach a 2 mm resolution on system level was accomplished with both algorithms. The neural network achieved a mean spatial resolution of 1.14 mm FWHM for the whole detector and 1.02 mm in the centre (30 x 30 mm(2)). The MNN algorithm performed slightly worse with 1.17 mm for the whole detector and 1.13 mm in the centre. The intrinsic DOI information will also result in uniform system spatial resolution over the full field of view.
引用
收藏
页数:15
相关论文
共 50 条
  • [21] A finely segmented semi-monolithic detector tailored for high-resolution PET
    Kuhl, Yannick
    Mueller, Florian
    Naunheim, Stephan
    Bovelett, Matthias
    Lambertus, Janko
    Schug, David
    Weissler, Bjoern
    Gegenmantel, Eike
    Gebhardt, Pierre
    Schulz, Volkmar
    MEDICAL PHYSICS, 2024, 51 (05) : 3421 - 3436
  • [22] Impact of Depth-of-Interaction on Image Resolution in Long Axial-FOV PET
    Yu, Yunhan
    Wang, Shi
    Wang, Xinqiang
    Xia, Yan
    Shang, Hong
    Wu, Jing
    Wang, Qiulin
    Liu, Yaqiang
    Ma, Tianyu
    2014 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2014,
  • [23] Improving depth-of-interaction resolution in pixellated PET detectors using neural networks
    Zatcepin, Artem
    Pizzichemi, Marco
    Polesel, Andrea
    Paganoni, Marco
    Auffray, Etiennette
    Ziegler, Sibylle, I
    Omidvari, Negar
    PHYSICS IN MEDICINE AND BIOLOGY, 2020, 65 (17):
  • [24] A practical depth-of-interaction PET/MR detector with dichotomous-orthogonal-symmetry decoding
    Yuxuan Zhang
    Hossain Baghaei
    Han Yan
    Wai-Hoi Wong
    EJNMMI Physics, 2 (Suppl 1)
  • [25] A NEW DUAL CRYSTAL DEPTH SENSITIVE DETECTOR FOR HIGH-RESOLUTION PET CAMERAS
    LECOMTE, R
    JOURNAL OF NUCLEAR MEDICINE, 1986, 27 (06) : 974 - 974
  • [26] A Depth-of-interaction PET Detector Using Mutual Gain-equalized Silicon Photomultiplier
    Xi, W.
    Weisenberger, A. G.
    Dong, H.
    Kross, B.
    Lee, S.
    McKisson, J.
    McKisson, J. E.
    Zorn, C.
    2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, : 3166 - 3168
  • [27] Depth-of-Interaction and Coincidence Time Resolution Characterization of Ultrahigh Resolution Prism-PET Modules
    Cao, Xinjie
    LaBella, Andy
    Zhao, Wei
    Goldan, Amir
    JOURNAL OF NUCLEAR MEDICINE, 2020, 61
  • [28] Detector normalization and scatter correction for the jPET-D4: A 4-layer depth-of-interaction PET scanner
    Kitamura, Keishi
    Ishikawa, Akihiro
    Mizuta, Tetsuro
    Yamaya, Taiga
    Yoshida, Eiji
    Murayama, Hideo
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 571 (1-2): : 231 - 234
  • [29] A Light Sharing, Charge Multiplexed Time-of-Flight Depth-of-Interaction PET Detector
    Bieniosek, Matthew F.
    Cates, Joshua W.
    Levin, Craig S.
    2015 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2015,
  • [30] A novel method for γ photons depth of interaction discrimination on monolithic LYSO crystals for brain PET/MRI
    Roberto Pani
    Marco Bettiol
    Enrico Preziosi
    Christian Borrazzo
    Rosanna Pellegrini
    Antonio Gonzalez
    Pablo Conde
    Maria Nerina Cinti
    Andrea Fabbri
    Elisabetta Di Castro
    Stan Majewski
    EJNMMI Physics, 2 (Suppl 1)