Simple Proposal for Dosimetry with an Elekta iViewGT™ Electronic Portal Imaging Device (EPID) Using Commercial Software Modules

被引:1
|
作者
Liebich, Janett [1 ]
Licher, Joerg [1 ]
Scherf, Christian [1 ]
Kara, Eugen [1 ]
Koch, Nadine [1 ]
Roedel, Claus [1 ]
Ramm, Ulla [1 ]
机构
[1] Klinikum Johann Wolfgang Goethe Univ Frankfurt Ma, Zentrum Radiol, Klin Strahlentherapie & Onkol, Frankfurt, Germany
关键词
EPID dosimetry; Amorphous silicon EPID; IMRT verification; Dose response; INTENSITY-MODULATED RADIOTHERAPY; DOSE-RESPONSE; IMRT; FIELD; VERIFICATION; CALIBRATION; RADIATION; MATRIX; MODEL;
D O I
10.1007/s00066-011-2176-z
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background: An electronic portal imaging device (EPID) is used to control for patient setup and positioning during fractionated radiotherapy. Due to the rising complexity and conformity of irradiation techniques, the demand for an accurate verification of the dose delivered to the patient has also increased. The purpose of this study was to investigate a simple guidance for dosimetry with an Elekta iViewGT (TM) EPID using commercial software modules. Material and Methods: EPID measurements were performed using an Elekta iViewGT (TM) EPID on a linear accelerator with 6 MV x-ray beam. The EPID signal was studied for reproducibility, as well as characteristics as a function of dose, dose rate, and field size. A series of experiments, comparing the response of the flat panel imager and ionization chamber measurements of dose, determine the parameters for the calibration model. EPID measurements were also compared with calculations of the treatment planning system. Results: We found a stable response of the EPID signal over a period of 14 months. It showed nonlinearity depending on dose up to 6.8%. There were low oscillations up to 1.2% depending on dose rate. For all fields, the calibrated flat panel profiles match the measured and calculated dose profiles with maximum deviation of 2-3% for the in-field region. In the high gradient areas, higher differences up to 6% were found. Conclusions: The gamma evaluation indicates good correlation between predicted and acquired EPID images. The EPID-based pretreatment IMRT verification method will help to improve the quality assurance procedure.
引用
收藏
页码:316 / 321
页数:6
相关论文
共 50 条
  • [21] Using An Electronic Portal Imaging Device (EPID) for Correlating Linac Photon Beam Energies
    Yaddanapudi, S.
    Cai, B.
    Sun, B.
    Noel, C.
    Goddu, S.
    Mutic, S.
    MEDICAL PHYSICS, 2015, 42 (06) : 3517 - 3517
  • [22] CT, laser and Linac axes alignment using an electronic portal imaging device (EPID)
    Samant, S
    Xia, J
    Patra, P
    Sobczak, D
    MEDICAL PHYSICS, 2004, 31 (06) : 1876 - 1876
  • [23] Electronic portal imaging device-based dosimetry
    Pinol, R. De Blas
    Farre, I. Modolell
    Olmos, C. Picon
    Kolster, I. Sancho
    Vaque, J. Puxeu
    Arroyo, M. Lizuain
    RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S187 - S187
  • [24] Evaluation of an electronic portal imaging device for transit dosimetry
    He, XD
    Van Esch, A
    Reymen, R
    Huyskens, D
    ACTA ONCOLOGICA, 1999, 38 (05) : 591 - 596
  • [25] Evaluation of an electronic portal imaging device for transit dosimetry
    He, Xiaodong
    Esch, Ann Van
    Reymen, Rita
    Huyskens, Dominique
    Acta Oncologica, 38 (05): : 591 - 596
  • [26] The use of an electronic portal imaging device for exit dosimetry
    Conte, L
    Stucchi, P
    Mordacchini, C
    Bianchi, C
    Monciardini, M
    Novario, R
    Cassani, E
    Mantovani, R
    Antognoni, P
    Tordiglione, M
    PROCEEDINGS OF THE XIITH INTERNATIONAL CONFERENCE ON THE USE OF COMPUTERS IN RADIATION THERAPY, 1997, : 279 - 281
  • [27] Development of an Electronic Portal Imaging Device Dosimetry Method
    Zhang, Jun
    Fan, Ziting
    Zhang, Xile
    Yang, Ruijie
    Wen, Junhai
    DIAGNOSTICS, 2021, 11 (09)
  • [28] Evaluation of quality assurance phantom for electronic portal imaging device (EPID)
    Das, I.
    Cheng, C.
    Chopra, K.
    Misic, V.
    Schule, E.
    MEDICAL PHYSICS, 2007, 34 (06) : 2371 - 2372
  • [29] Reliability analysis of an Electronic Portal Imaging Device (EPID) of a linear accelerator
    Gouveia, Clesia
    Sobral, Jose
    2023 IEEE 7TH PORTUGUESE MEETING ON BIOENGINEERING, ENBENG, 2023, : 52 - 55
  • [30] Sensitivity of Electronic Portal Imaging Device (EPID) Based Transit Dosimetry to Detect Inter-fraction Patient Variations
    Bawazeer, Omemh
    Sarasanandarajah, Sivananthan
    Herath, Sisira
    Kron, Tomas
    Deb, Pradip
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING 2018, VOL 3, 2019, 68 (03): : 477 - 480