A Comparison of Laparoscopic Energy Devices on Charges in Thermal Power After Application to Porcine Mesentery

被引:9
|
作者
Eto, Ken [1 ]
Omura, Nobuo [1 ]
Haruki, Koichiro [1 ]
Uno, Yoshiko [1 ]
Ohkuma, Masahisa [1 ]
Nakajima, Shintaro [1 ]
Anan, Tadashi [1 ]
Kosuge, Makoto [1 ]
Fujita, Tetsuji [1 ]
Yanaga, Katsuhiko [1 ]
机构
[1] Jikei Univ, Sch Med, Dept Surg, Minato Ku, Tokyo 1058461, Japan
关键词
laparoscopy; energy device; thermal injury; HARMONIC SCALPEL; ULTRASONIC ENERGY; MONOPOLAR; LIGASURE; ARTERIES; ELECTROCOAGULATION; CHOLECYSTECTOMY; HEMOSTASIS; DISSECTION; DIATHERMY;
D O I
10.1097/SLE.0000000000000059
中图分类号
R61 [外科手术学];
学科分类号
摘要
Introduction: Advances in energy devices have played a major role in the rapid expansion of laparoscopic surgery. However, complications due to these energy devices are occasionally reported, and if the characteristics of these devices are not well understood, serious complications may occur. This study evaluated various typical energy devices and measured temperature rises in the adjacent tissue and in the devices themselves. Equipment and Methods: We used the following 7 types of energy devices: AutoSonix (AU), SonoSurg (SS), Harmonic Scalpel (HS), LigaSure Atlas (LA), LigaSure Dolphin Tip (LD), monopolar diathermy (Mono), and bipolar scissors (Bi). Laparoscopy was performed under general anesthesia in pigs, and the mesentery was dissected using each energy device. Tissue temperature at a distance of 1mm from the energy device blade before and after dissection was measured. Temperature of the device blade both before and after dissection, time required for dissection, and interval until the temperature fell to 100 degrees C, 75 degrees C, and 50 degrees C were documented. Results: Temperature of the surrounding tissue using each device rose the most with the Mono (50.5 +/- 8.0 degrees C) and the least with the HS in full mode (6.2 +/- 0.7 degrees C). Device temperature itself rose the highest with the AU in full mode (318.2 +/- 49.6 degrees C), and the least with the Bi (61.9 +/- 4.8 degrees C). All ultrasonic coagulation and cutting devices (AU, SS, and HS) had device temperatures increase up to >= 100 degrees C, and even at 8 seconds after completing dissection, temperatures remained at >= 100 degrees C. Conclusions: Because the adjacent tissue temperature peaked with the Mono, cautious use near the intestine and blood vessels is necessary. In addition, the active blades of all ultrasonic coagulation and cutting devices, regardless of model, developed high temperatures exceeding 100 degrees C. Therefore, an adequate cooling period after using these devices is therefore necessary between applications.
引用
收藏
页码:E37 / E41
页数:5
相关论文
共 50 条
  • [31] POWER-DELAY ENERGY COMPARISON OF BIPOLAR AND IGFET DIGITAL DEVICES AND CIRCUITS
    JOHNSON, EO
    IEEE TRANSACTIONS ON ELECTRON DEVICES, 1975, 22 (11) : 1044 - 1045
  • [32] Postoperative adhesion formation after laparoscopic uterine horn resection in a porcine model: Comparison of five instruments
    Hirota, Y
    Tsukada, K
    Nishio, E
    Yoshida, M
    Tada, S
    Udagawa, Y
    JOURNAL OF LAPAROENDOSCOPIC & ADVANCED SURGICAL TECHNIQUES, 2005, 15 (06): : 581 - 585
  • [33] Energy efficiency in China: optimization and comparison between hydropower and thermal power
    Wang, Ruo-Mei
    Tian, Ze
    Ren, Fang-rong
    ENERGY SUSTAINABILITY AND SOCIETY, 2021, 11 (01)
  • [34] Energy efficiency in China: optimization and comparison between hydropower and thermal power
    Ruo-Mei Wang
    Ze Tian
    Fang-rong Ren
    Energy, Sustainability and Society, 11
  • [35] Application of thermal video system in energy saving of heating power pipe
    Tuan, Yuiyang
    Hongwai jishu, 1993, 15 (01): : 27 - 31
  • [36] Test and Analysis of the Heat Exchanger for Small Ocean Thermal Energy Power Generation Devices
    Wu, Xiao
    Wang, Xiangnan
    Wang, Bingzhen
    ENERGIES, 2023, 16 (22)
  • [37] Application of computer-based thermography to thermal measurements of integrated circuits and power devices
    Grecki, M
    Pacholik, J
    Wiecek, B
    Napieralski, A
    MICROELECTRONICS JOURNAL, 1997, 28 (03) : 337 - 347
  • [38] Comparison of effects of energy based devices on quality of life after sutureless thyroidectomy
    Arikan, Akif Enes
    Ozyegin, Mustafa Ates
    MEDICINE, 2022, 101 (02) : E28518
  • [39] Thermal transport measurement of three-dimensional graphene powders for application in energy devices
    Li, C.
    Liu, Z.
    Zhang, X.
    Hasan, S. W.
    Tian, Z. Q.
    Zhou, J.
    Yin, Z.
    MATERIALS TODAY ENERGY, 2021, 19
  • [40] From chip to inverter: Electro-thermal modeling and design for paralleled power devices in high power application
    Fan, Peng
    Huang, Shoudao
    Wang, Huai
    Li, Huimin
    Luo, Derong
    MICROELECTRONICS RELIABILITY, 2018, 87 : 271 - 277