A prototype small CdTe gamma camera for radioguided surgery and other imaging applications

被引:52
|
作者
Tsuchimochi, M
Sakahara, H
Hayama, K
Funaki, M
Ohno, R
Shirahata, T
Orskaug, T
Maehlum, G
Yoshioka, K
Nygard, E
机构
[1] Nippon Dent Univ, Sch Dent Niigata, Dept Oral & Maxillofacial Radiol, Niigata 9518580, Japan
[2] Hamamatsu Univ Sch Med, Dept Radiol, Hamamatsu, Shizuoka 43131, Japan
[3] Acrorad Co Ltd, Tokyo, Japan
[4] Integrated Detector & Elect AS, Hovik, North Ireland
关键词
CdTe; gamma camera; semiconductor; sentinel lymph node biopsy; radioguided surgery;
D O I
10.1007/s00259-003-1301-3
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Gamma probes have been used for sentinel lymph node biopsy in melanoma and breast cancer. However, these probes can provide only radioactivity counts and variable pitch audio output based on the intensity of the detected radioactivity. We have developed a small semiconductor gamma camera (SSGC) that allows visualisation of the size, shape and location of the target tissues. This study is designed to characterise the performance of the SSGC for radioguided surgery of metastatic lesions and for other imaging applications amenable to the smaller format of this prototype imaging system. The detector head had 32 cadmium telluride semiconductor arrays with a total of 1,024 pixels, and with application-specific integrated circuits (ASICs) and a tungsten collimator. The entire assembly was encased in a lead housing measuring 152 mmx166 mmx65 rum. The effective visual field was 44.8 mmx44.8 mm. The energy resolution and imaging aspects were tested. Two spherical 5-mm- and 15-mm-diameter technetium-99m radioactive sources that had activities of 0.15 MBq and 100 MBq, respectively, were used to simulate a sentinel lymph node and an injection site. The relative detectability of these foci by the new detector and a conventional scintillation camera was studied. The prototype was also examined in a variety of clinical applications. Energy resolution [full-width at half-maximum (FWHM)] for a single element at the centre of the field of view was 4.2% at 140 keV (Tc-99m), and the mean energy resolution of the CdTe detector arrays was approximately 7.8%. The spatial resolution, represented by FWHM had a mean value of 1.56+/-0.05 mm. Simulated node foci could be visualised clearly by the SSGC using a 15-s acquisition time. In preliminary clinical tests, the SSGC successfully imaged diseases in a variety of tissues, including salivary and thyroid glands, temporomandibular joints and sentinel lymph nodes. The SSGC has significant potential for diagnosing diseases and facilitating subsequent radioguided surgery.
引用
收藏
页码:1605 / 1614
页数:10
相关论文
共 50 条
  • [1] A prototype small CdTe gamma camera for radioguided surgery and other imaging applications
    Makoto Tsuchimochi
    Harumi Sakahara
    Kazuhide Hayama
    Minoru Funaki
    Ryoichi Ohno
    Takashi Shirahata
    Terje Orskaug
    Gunnar Maehlum
    Koki Yoshioka
    Einar Nygard
    European Journal of Nuclear Medicine and Molecular Imaging, 2003, 30 : 1605 - 1614
  • [2] Performance of a small CdTe gamma camera for radioguided surgery
    Makoto, T
    Sakahara, H
    Hayama, K
    Funaki, M
    Shirahata, T
    Orskaug, T
    Maehlum, G
    Yoshioka, K
    Nygard, E
    PENETRATING RADIATION SYSTEMS AND APPLICATIONS III, 2001, 4508 : 74 - 87
  • [3] CdTe compact gamma camera for coded aperture imaging in radioguided surgery
    Russo, Paolo
    Di Lillo, Francesca
    Corvino, Veronica
    Frallicciardi, Paola Maria
    Sarno, Antonio
    Mettivier, Giovanni
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2020, 69 : 223 - 232
  • [4] Sensitivity of a prototype cadmium telluride gamma camera for radioguided surgery.
    Tsuchimochi, M
    Sakahara, H
    Hayama, K
    Funaki, M
    Shirahata, T
    Orskaug, T
    Maehlum, G
    Yoshioka, K
    Nygard, E
    JOURNAL OF NUCLEAR MEDICINE, 2001, 42 (05) : 101P - 101P
  • [5] Image synthesis of small semiconductor gamma camera and optical camera for identifying sentinel lymph nodes in radioguided surgery
    Haneishi, Hideaki
    Onishi, Yoshimi
    Hayashi, Hideki
    2005 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2005, : 2165 - 2169
  • [6] Applications and imaging techniques of a Si/CdTe Compton gamma-ray camera
    Takeda, Shin'ichiro
    Ichinohe, Yuto
    Hagino, Kouichi
    Odaka, Hirokazu
    Yuasa, Takayuki
    Ishikawa, Shin-nosuke
    Fukuyama, Taro
    Saito, Shinya
    Sato, Tamotsu
    Sato, Goro
    Watanabe, Shin
    Kokubun, Motohide
    Takahashi, Tadayuki
    Yamaguchi, Mitsutaka
    Tajima, Hiroyasu
    Tanaka, Takaaki
    Nakazawa, Kazuhiro
    Fukazawa, Yasushi
    Nakano, Takashi
    PROCEEDINGS OF THE 2ND INTERNATIONAL CONFERENCE ON TECHNOLOGY AND INSTRUMENTATION IN PARTICLE PHYSICS (TIPP 2011), 2012, 37 : 859 - 866
  • [7] Small scale prototype of a CdTe imaging system
    Lambropoulos, H
    Papadakis, I
    Loukas, D
    Zervakis, M
    2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7, 2004, : 4540 - 4542
  • [8] Evaluation of Compton gamma camera prototype based on pixelated CdTe detectors
    Calderon, Y.
    Chmeissani, M.
    Kolstein, M.
    De Lorenzo, G.
    JOURNAL OF INSTRUMENTATION, 2014, 9
  • [9] Clinical applications of gamma-detection probes – radioguided surgery
    Schlomo Schneebaum
    Einat Even-Sapir
    Meir Cohen
    Hedva Shacham-Lehrman
    Andrea Gat
    Eli Brazovsky
    Gennady Livshitz
    Jona Stadler
    Yehuda Skornick
    European Journal of Nuclear Medicine, 1999, 26 (Suppl 1) : S26 - S35
  • [10] A Compton camera prototype for prompt gamma medical imaging
    Thirolf, P. G.
    Aldawood, S.
    Boehmer, M.
    Bortfeldt, J.
    Castelhano, I.
    Dedes, G.
    Fiedler, F.
    Gernhaeuser, R.
    Golnik, C.
    Helmbrecht, S.
    Hueso-Gonzalez, F.
    Kolff, H. V. D.
    Kormoll, T.
    Lang, C.
    Liprandi, S.
    Lutter, R.
    Marinsek, T.
    Maier, L.
    Pausch, G.
    Petzoldt, J.
    Roemer, K.
    Schaart, D.
    Parodi, K.
    12TH INTERNATIONAL CONFERENCE ON NUCLEUS-NUCLEUS COLLISIONS 2015, 2016, 117