Systematic error of the nose-to-nose sampling-oscilloscope calibration

被引:5
|
作者
Williams, Dylan F. [1 ]
Clement, Tracy S.
Remley, Kate A.
Hale, Tatil D.
Verbeyst, Frans
机构
[1] Natl Inst Stand & Technol, Boulder, CO 80305 USA
[2] NMDG Engn, B-2880 Bornem, Belgium
[3] Vrije Univ Brussel VIB, Fac Ingn, Vakgrp ELEC, B-1050 Brussels, Belgium
关键词
calibration; electrooptic sampling; impedance mismatch; mismatch correction; nose-to-nose calibration; sampling oscilloscope; swept-sine calibration; UNCERTAINTY; DISTORTION;
D O I
10.1109/TMTT.2007.904333
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We use traceable swept-sine and electrooptic-sampling-system-based sampling-oscilloscope calibrations to measure the systematic error of the nose-to-nose calibration, and compare the results to simulations. Our results show that the errors in the nose-to-nose calibration are small at low frequencies, but significant at high frequencies.
引用
收藏
页码:1951 / 1957
页数:7
相关论文
共 50 条
  • [1] Nose-to-nose oscilloscope calibration phase error inherent in the sampling circuitry
    Remley, KA
    ARFTG: AUTOMATIC RF TECHNIQUES GROUP, CONFERENCE DIGEST, FALL 2002: MEASUREMENTS NEEDS FOR EMERGING TECHNOLOGIES, 2002, : 85 - 97
  • [2] New progress in theory of nose-to-nose calibration of broadband sampling oscilloscope
    Lin, M.L.
    Zhao, H.
    Tien Tzu Hsueh Pao/Acta Electronica Sinica, 2001, 29 (03): : 409 - 411
  • [3] Auto-calibrating system of broadband sampling oscilloscope based on nose-to-nose calibration procedure
    Zhao, Ke-Jia
    Liu, Ming-Liang
    Gao, Xiao-Xun
    Lu, Feng
    Zhu, Jiang-Miao
    Miao, Jing-Yuan
    Jiliang Xuebao/Acta Metrologica Sinica, 2006, 27 (04): : 388 - 391
  • [5] Research on the uncertainty of transition duration of broadband sampling oscilloscope based on nose-to-nose technology
    Yu, Yue-Hua
    Liu, Ming-Liang
    Zhao, Ke-Jia
    Zhao, Yi
    Zhu, Jiang-Miao
    Miao, Jing-Yuan
    Jiliang Xuebao/Acta Metrologica Sinica, 2007, 28 (01): : 81 - 84
  • [6] The simulation of the nose-to-nose calibration procedure
    Zhu, JM
    Liu, ML
    Gao, Y
    ICEMI'2003: PROCEEDINGS OF THE SIXTH INTERNATIONAL CONFERENCE ON ELECTRONIC MEASUREMENT & INSTRUMENTS, VOLS 1-3, 2003, : 1217 - 1220
  • [7] Extended "Nose-to-Nose" calibration procedure
    Liu, ML
    Wang, YT
    Miao, JY
    Yu, YH
    ICEMI 2005: Conference Proceedings of the Seventh International Conference on Electronic Measurement & Instruments, Vol 2, 2005, : 641 - 645
  • [8] Novel research of nose-to-nose calibration foundation
    Zhu, JM
    Liu, ML
    Lu, F
    ISTM/2005: 6th International Symposium on Test and Measurement, Vols 1-9, Conference Proceedings, 2005, : 8183 - 8187
  • [9] INDIVIDUAL CHARACTERIZATION OF BROAD-BAND SAMPLING OSCILLOSCOPES WITH A NOSE-TO-NOSE CALIBRATION PROCEDURE
    VERSPECHT, J
    RUSH, K
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1994, 43 (02) : 347 - 354
  • [10] A calibration theory and method named nose-to-nose based on the agilent broadband sampling oscilloscopes
    Zhu, JM
    Liu, ML
    Gao, Y
    Deng, C
    ISTM/2003: 5TH INTERNATIONAL SYMPOSIUM ON TEST AND MEASUREMENT, VOLS 1-6, CONFERENCE PROCEEDINGS, 2003, : 3051 - 3054