Systematic error of the nose-to-nose sampling-oscilloscope calibration

被引:5
|
作者
Williams, Dylan F. [1 ]
Clement, Tracy S.
Remley, Kate A.
Hale, Tatil D.
Verbeyst, Frans
机构
[1] Natl Inst Stand & Technol, Boulder, CO 80305 USA
[2] NMDG Engn, B-2880 Bornem, Belgium
[3] Vrije Univ Brussel VIB, Fac Ingn, Vakgrp ELEC, B-1050 Brussels, Belgium
关键词
calibration; electrooptic sampling; impedance mismatch; mismatch correction; nose-to-nose calibration; sampling oscilloscope; swept-sine calibration; UNCERTAINTY; DISTORTION;
D O I
10.1109/TMTT.2007.904333
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We use traceable swept-sine and electrooptic-sampling-system-based sampling-oscilloscope calibrations to measure the systematic error of the nose-to-nose calibration, and compare the results to simulations. Our results show that the errors in the nose-to-nose calibration are small at low frequencies, but significant at high frequencies.
引用
收藏
页码:1951 / 1957
页数:7
相关论文
共 50 条
  • [41] Experimental infection by Salmonella enterica ssp. enterica serovar Panama and tentative of nose-to-nose transmission in weaned pigs
    Oliveira, L. G.
    Carvalho, L. F. O. S.
    Masson, G. C. I. H.
    Feliciano, M. A. R.
    ARQUIVO BRASILEIRO DE MEDICINA VETERINARIA E ZOOTECNIA, 2010, 62 (06) : 1340 - 1347
  • [42] 基于Nose-to-Nose校准法的取样示波器冲激响应的一种算法
    赵华
    林茂六
    电子学报, 2003, (03) : 365 - 367
  • [43] Rapid millimetre-wave sampler response characterization to well beyond 120GHz using an improved nose-to-nose method
    Scott, JB
    2003 IEEE MTT-S INTERNATIONAL MICROWAVE SYMPOSIUM DIGEST, VOLS 1-3, 2003, : 1511 - 1514
  • [44] Calibration of an Electronic Nose for Poultry Farm
    Abdullah, A. H.
    Shukor, S. A.
    Kamis, M. S.
    Shakaff, A. Y. M.
    Zakaria, A.
    Rahim, N. A.
    Mamduh, S. M.
    Kamarudin, K.
    Saad, F. S. A.
    Masnan, M. J.
    Mustafa, H.
    11TH ASIAN CONFERENCE ON CHEMICAL SENSORS (ACCS2015), 2017, 1808
  • [45] The effects of offset voltage on the amplitude and bandwidth of kick-out pulses used in the nose-to-nose sampler impulse response characterization method
    Larson, DR
    Paulter, NG
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2001, 50 (04) : 872 - 876
  • [46] VITAL CYTOLOGICAL SAMPLING FROM THE NOSE
    DEITMER, T
    ARCHIVES OF OTO-RHINO-LARYNGOLOGY, 1987, 244 (05): : 333 - 333
  • [47] A nose-to-nose comparison of the physiological effects of exposure to ionic silver versus silver chloride in the European eel (Anguilla anguilla) and the rainbow trout (Oncorhynchus mykiss)
    Grosell, M
    Hogstrand, C
    Wood, CM
    Hansen, HJM
    AQUATIC TOXICOLOGY, 2000, 48 (2-3) : 327 - 342
  • [48] PULSED WAVEMETER TIMING REFERENCE FOR SAMPLING OSCILLOSCOPE CALIBRATION
    ANDREWS, JR
    GANS, WL
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1975, IM24 (01) : 82 - 82
  • [49] Deviated Nose: A Systematic Approach for Correction
    Abu El-Wafa, Ahmed M.
    Emara, Sherif S.
    PLASTIC AND RECONSTRUCTIVE SURGERY-GLOBAL OPEN, 2020, 8 (08)
  • [50] AI-aided e-nose calibration
    Malec, A.
    Suchorab, Z.
    Guz, L.
    Caserta, C.
    IV INTERNATIONAL CONFERENCE OF COMPUTATIONAL METHODS IN ENGINEERING SCIENCE (CMES'19), 2019, 710