Systematic error of the nose-to-nose sampling-oscilloscope calibration

被引:5
|
作者
Williams, Dylan F. [1 ]
Clement, Tracy S.
Remley, Kate A.
Hale, Tatil D.
Verbeyst, Frans
机构
[1] Natl Inst Stand & Technol, Boulder, CO 80305 USA
[2] NMDG Engn, B-2880 Bornem, Belgium
[3] Vrije Univ Brussel VIB, Fac Ingn, Vakgrp ELEC, B-1050 Brussels, Belgium
关键词
calibration; electrooptic sampling; impedance mismatch; mismatch correction; nose-to-nose calibration; sampling oscilloscope; swept-sine calibration; UNCERTAINTY; DISTORTION;
D O I
10.1109/TMTT.2007.904333
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We use traceable swept-sine and electrooptic-sampling-system-based sampling-oscilloscope calibrations to measure the systematic error of the nose-to-nose calibration, and compare the results to simulations. Our results show that the errors in the nose-to-nose calibration are small at low frequencies, but significant at high frequencies.
引用
收藏
页码:1951 / 1957
页数:7
相关论文
共 50 条
  • [21] Sampling-oscilloscope Rise Time.
    Bonizzato, Maurizio
    Schiaffino, Paolo
    Energia Elettrica, 1982, 59 (09): : 397 - 401
  • [22] Nose-to-Nose校准技术的仿真研究
    朱苏磊
    测试技术学报, 2005, (04) : 394 - 397
  • [23] Nose-to-nose with foul smells:: learning to live in the same place
    Remy, Elisabeth
    Estades, Jacqueline
    SOCIOLOGIE DU TRAVAIL, 2007, 49 (02) : 237 - 252
  • [24] Comparison of Nose-to-nose procedure and swept-sine method in broad-band system calibration
    Zhu, JM
    Liu, ML
    Wang, YT
    Lu, F
    Yu, YH
    Miao, JY
    ICEMI 2005: Conference Proceedings of the Seventh International Conference on Electronic Measurement & Instruments, Vol 2, 2005, : 646 - 649
  • [25] Microwave scope response characterization using a nose-to-nose method
    Lee, Joo-Gwang
    Kang, Jin-Seob
    Kang, Tae-Weon
    Won, Sung-Ho
    2009 34TH INTERNATIONAL CONFERENCE ON INFRARED, MILLIMETER, AND TERAHERTZ WAVES, VOLS 1 AND 2, 2009, : 51 - 52
  • [26] Nose-to-nose transmission of Salmonella Typhimurium between weaned pigs
    Oliveira, Celso J. B.
    Garcia, Thiago B.
    Carvalho, Luiz Fernando O. S.
    Givisiez, Patricia E. N.
    VETERINARY MICROBIOLOGY, 2007, 125 (3-4) : 355 - 361
  • [27] HOW TO EXTEND SAMPLING-OSCILLOSCOPE VERSATILITY
    ZIMMERMAN, HA
    IEEE SPECTRUM, 1969, 6 (04) : 79 - +
  • [28] COMPUTER PROCESSING OF SAMPLING-OSCILLOSCOPE DATA
    COHN, CE
    TRANSACTIONS OF THE AMERICAN NUCLEAR SOCIETY, 1968, 11 (01): : 72 - &
  • [29] Measuring step geometry using the Nose-to-Nose method: Validity and repeatability
    Johnson, Daniel
    Sloan, Gary
    WORK-A JOURNAL OF PREVENTION ASSESSMENT & REHABILITATION, 2012, 41 : 3343 - 3348
  • [30] Contamination of pigs by nose-to-nose contact or airborne transmission of Salmonella Typhimurium
    Proux, K
    Cariolet, R
    Fravalo, P
    Houdayer, C
    Keranflech, A
    Madec, F
    VETERINARY RESEARCH, 2001, 32 (06) : 591 - 600