Erroneous Pressure Reading from an Unpowered Intra-aortic Balloon Pump: Case Report

被引:0
|
作者
Belway, Dean [1 ]
Corrigan, Andrea [1 ]
Manchuk, Derek [2 ]
Bittira, Bindu [3 ,4 ]
机构
[1] Perfus Serv, Sudbury, ON P3E 5J1, Canada
[2] Northern Ontario Sch Med, Dept Anesthesia, Sudbury, ON P3E 5J1, Canada
[3] Northern Ontario Sch Med, Div Cardiac Surg, Sudbury, ON P3E 5J1, Canada
[4] Hlth Sci North, 41 Ramsey Lake Rd, Sudbury, ON P3E 5J1, Canada
关键词
case report; equipment malfunctions; intra-aortic balloon pump; CARDIAC-SURGERY; COUNTERPULSATION;
D O I
10.1097/MAT.0000000000001695
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Intra-aortic balloon pump (IABP) technology is an established treatment modality for patients with acute cardiac failure. Advances in IABP technology have simplified use and reliability such that electromechanical malfunctions are relatively rare. We present a case of an unanticipated pressure output signal from a powered off Cardiosave IABP console resulting in an erroneous mean arterial blood pressure waveform being displayed during cardiopulmonary bypass. The patient suffered no ill effects and made a full recovery. As a result of this incident, our policies have been revised to ensure that all patients with an IABP brought to the operating room for surgery have at least two arterial lines, one of which is not off the IABP, to prevent future occurrences.
引用
收藏
页码:E211 / E214
页数:4
相关论文
共 50 条
  • [21] INTRA-AORTIC BALLOON PUMP: INDICATIONS FOR USE
    Schampaert, S.
    van 't Veer, M.
    de Vosse, F. N. van
    Pills, N. H. J.
    Rutten, M. C. M.
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2011, 34 (08): : 631 - 631
  • [22] PREOPERATIVE INTRA-AORTIC BALLOON PUMP ASSISTANCE
    MUNDTH, ED
    ANNALS OF THORACIC SURGERY, 1976, 22 (06): : 603 - 604
  • [23] Radiology of Intra-Aortic Balloon Pump Catheters
    Rhodes, Nicholas G.
    Johnson, Tucker F.
    Boyum, James H.
    Khandelwal, Ashish
    Howell, Barrett D.
    Froemming, Adam T.
    Behfar, Atta
    RADIOLOGY-CARDIOTHORACIC IMAGING, 2022, 4 (02):
  • [24] Principles of intra-aortic balloon pump counterpulsation
    Krishna, Murli
    Zacharowski, Kai
    BJA EDUCATION, 2009, 9 (01) : 24 - 28
  • [25] Fourier analysis of the intra-aortic balloon pump
    Grossi, E
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1999, 118 (03): : 568 - 568
  • [26] Fourier analysis of the intra-aortic balloon pump
    Poullis, M
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1999, 117 (04): : 827 - 828
  • [27] Intra-aortic balloon pump: is the tide turning?
    Sandro Gelsomino
    Daniel M. Johnson
    Roberto Lorusso
    Critical Care, 22
  • [28] A complication of an axillary intra-aortic balloon pump
    Kodack, Eileen
    Patel, Riya
    Bonner, Carain
    Pantin, Enrique
    JOURNAL OF SURGICAL CASE REPORTS, 2024, 2024 (03):
  • [29] Mathematical modelling of intra-aortic balloon pump
    Abdolrazaghi, Mona
    Navidbakhsh, Mahdi
    Hassani, Kamran
    COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING, 2010, 13 (05) : 567 - 576
  • [30] INTRA-AORTIC BALLOON ASSISTANCE WITHOUT A PUMP
    SIDERIS, DA
    NANAS, JN
    CHRYSOS, DN
    MOULOPOULOS, SD
    EUROPEAN HEART JOURNAL, 1983, 4 (08) : 536 - 546