Erroneous Pressure Reading from an Unpowered Intra-aortic Balloon Pump: Case Report

被引:0
|
作者
Belway, Dean [1 ]
Corrigan, Andrea [1 ]
Manchuk, Derek [2 ]
Bittira, Bindu [3 ,4 ]
机构
[1] Perfus Serv, Sudbury, ON P3E 5J1, Canada
[2] Northern Ontario Sch Med, Dept Anesthesia, Sudbury, ON P3E 5J1, Canada
[3] Northern Ontario Sch Med, Div Cardiac Surg, Sudbury, ON P3E 5J1, Canada
[4] Hlth Sci North, 41 Ramsey Lake Rd, Sudbury, ON P3E 5J1, Canada
关键词
case report; equipment malfunctions; intra-aortic balloon pump; CARDIAC-SURGERY; COUNTERPULSATION;
D O I
10.1097/MAT.0000000000001695
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Intra-aortic balloon pump (IABP) technology is an established treatment modality for patients with acute cardiac failure. Advances in IABP technology have simplified use and reliability such that electromechanical malfunctions are relatively rare. We present a case of an unanticipated pressure output signal from a powered off Cardiosave IABP console resulting in an erroneous mean arterial blood pressure waveform being displayed during cardiopulmonary bypass. The patient suffered no ill effects and made a full recovery. As a result of this incident, our policies have been revised to ensure that all patients with an IABP brought to the operating room for surgery have at least two arterial lines, one of which is not off the IABP, to prevent future occurrences.
引用
收藏
页码:E211 / E214
页数:4
相关论文
共 50 条
  • [41] Intra-Aortic Balloon Pump (IABP) in cardiogenic shock
    Fuernau, Georg
    Thiele, Holger
    CURRENT OPINION IN CRITICAL CARE, 2013, 19 (05) : 404 - 409
  • [42] COMPLICATIONS ASSOCIATED WITH USE OF THE INTRA-AORTIC BALLOON PUMP
    GOLDMAN, BS
    HILL, TJ
    ROSENTHAL, GA
    SCULLY, HE
    WEISEL, RD
    BAIRD, RJ
    CANADIAN JOURNAL OF SURGERY, 1982, 25 (02) : 153 - 156
  • [43] Intra-aortic balloon pump: is the technique really outdated?
    Akodad, Mariama
    Delmas, Clement
    Bonello, Laurent
    Duflos, Claire
    Roubille, Francois
    ESC HEART FAILURE, 2020, 7 (03): : 1025 - 1030
  • [44] Association of thrombocytopenia and use of intra-aortic balloon pump
    Vonderheide, RH
    Kuter, DJ
    Thadhani, R
    CIRCULATION, 1996, 94 (08) : 4297 - 4297
  • [45] A rare complication of intra-aortic balloon pump placement
    Varadarajan, B
    Karski, J
    Vegas, A
    Heinrich, L
    JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2005, 19 (02) : 259 - 260
  • [46] Temporal trends in intra-aortic balloon pump use
    Cohen, B.
    Kalmanovich, E. K.
    Perl, L. P.
    Greenberg, G. G.
    Samara, A. S.
    Beigel, R. B.
    Ovdat, T. O.
    Kornowski, R. K.
    Orvin, K. O.
    EUROPEAN HEART JOURNAL, 2023, 44
  • [47] AXILLARY INTRA-AORTIC BALLOON PUMP: NEUROVASCULAR CONSIDERATIONS
    Duarte, Gustavo
    Lopez, Jose L.
    Sanchez, Alejandro
    Uppal, Dipan
    Alonso, Mileydis
    McDonald, Malcolm Lawson
    Snipelisky, David
    Estep, Jerry D.
    Baran, David A.
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2024, 83 (13) : 3559 - 3559
  • [48] Role of intra-aortic balloon pump in myocardial contusion
    Vu, T.
    Stahl, K. D.
    Asensio, J. A.
    ANNALS OF THE ROYAL COLLEGE OF SURGEONS OF ENGLAND, 2011, 93 (06) : 490 - 491
  • [49] CLINICAL-APPLICATIONS OF THE INTRA-AORTIC BALLOON PUMP
    STEINHOFF, NG
    LUNDELL, DC
    ARIZONA MEDICINE, 1981, 38 (01) : 19 - 20
  • [50] CHANGING PATTERNS OF INTRA-AORTIC BALLOON PUMP ASSISTANCE
    WEISEL, RD
    GOLDMAN, BS
    TEASDALE, S
    BAIGRIE, RS
    CANADIAN JOURNAL OF SURGERY, 1978, 21 (02) : 125 - &