Erroneous Pressure Reading from an Unpowered Intra-aortic Balloon Pump: Case Report

被引:0
|
作者
Belway, Dean [1 ]
Corrigan, Andrea [1 ]
Manchuk, Derek [2 ]
Bittira, Bindu [3 ,4 ]
机构
[1] Perfus Serv, Sudbury, ON P3E 5J1, Canada
[2] Northern Ontario Sch Med, Dept Anesthesia, Sudbury, ON P3E 5J1, Canada
[3] Northern Ontario Sch Med, Div Cardiac Surg, Sudbury, ON P3E 5J1, Canada
[4] Hlth Sci North, 41 Ramsey Lake Rd, Sudbury, ON P3E 5J1, Canada
关键词
case report; equipment malfunctions; intra-aortic balloon pump; CARDIAC-SURGERY; COUNTERPULSATION;
D O I
10.1097/MAT.0000000000001695
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Intra-aortic balloon pump (IABP) technology is an established treatment modality for patients with acute cardiac failure. Advances in IABP technology have simplified use and reliability such that electromechanical malfunctions are relatively rare. We present a case of an unanticipated pressure output signal from a powered off Cardiosave IABP console resulting in an erroneous mean arterial blood pressure waveform being displayed during cardiopulmonary bypass. The patient suffered no ill effects and made a full recovery. As a result of this incident, our policies have been revised to ensure that all patients with an IABP brought to the operating room for surgery have at least two arterial lines, one of which is not off the IABP, to prevent future occurrences.
引用
收藏
页码:E211 / E214
页数:4
相关论文
共 50 条
  • [31] Intra-aortic balloon pump: is the tide turning?
    Gelsomino, Sandro
    Johnson, Daniel M.
    Lorusso, Roberto
    CRITICAL CARE, 2018, 22
  • [32] Usefulness of Intra-aortic Balloon Pump Counterpulsation
    van Nunen, Lokien X.
    Noc, Marko
    Kapur, Navin K.
    Patel, Manesh R.
    Perera, Divaka
    Pijls, Nico H. J.
    AMERICAN JOURNAL OF CARDIOLOGY, 2016, 117 (03): : 469 - 476
  • [33] Intra-aortic balloon pump in heart surgery
    Engin, Mesut
    Onur, Tugba
    Onur, Anil
    GENERAL THORACIC AND CARDIOVASCULAR SURGERY, 2024, 72 (06) : 426 - 427
  • [34] Intra-aortic balloon pump entrapment without rupture: a case series
    Sogomonian, Robert
    Liu, Kevin
    Haftevani, Emma A.
    Gowda, Ramesh M.
    FUTURE CARDIOLOGY, 2020, 16 (05) : 425 - 432
  • [35] Intra-aortic balloon pump microperforation resulting in balloon entrapment
    Warrillow, S. J.
    Fealy, N. G.
    Houston, S. A.
    ANAESTHESIA AND INTENSIVE CARE, 2008, 36 (06) : 916 - 917
  • [36] Intra-Aortic Balloon Pump Simulating Thoracic Aortic Dissection
    Siegel, Yoel
    Munera, Felipe
    Durso, Anthony M.
    Roth, Patrick
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2018, 42 (04) : 630 - 631
  • [37] Undiagnosed aortic dissection in patient with intra-aortic balloon pump
    Raut, Monish S.
    Maheshwari, Arun
    Sharma, Manish
    ANNALS OF CARDIAC ANAESTHESIA, 2016, 19 (03) : 549 - +
  • [38] PARAPLEGIA FOLLOWING INTRA-AORTIC BALLOON SUPPORT - REPORT OF A CASE
    ORR, E
    MCKITTRICK, J
    DAGOSTINO, R
    FISHER, GU
    WRIGHT, R
    POPAT, K
    DJERGAIAN, R
    LOVE, J
    JOURNAL OF CARDIOVASCULAR SURGERY, 1989, 30 (06): : 1013 - 1014
  • [40] INTRA-AORTIC BALLOON-PUMP RUPTURE AND ENTRAPMENT
    BRODELL, GK
    TUZCU, EM
    WEISS, SJ
    SIMPFENDORFER, C
    CLEVELAND CLINIC JOURNAL OF MEDICINE, 1989, 56 (07) : 740 - 742